Index: AlarmDefs.h =================================================================== diff -u -r848df42ec75f88178fd55ff17eac495d1495534d -r8dc55f71fa867cc11804a71c81ed797e37481541 --- AlarmDefs.h (.../AlarmDefs.h) (revision 848df42ec75f88178fd55ff17eac495d1495534d) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 8dc55f71fa867cc11804a71c81ed797e37481541) @@ -1345,7 +1345,6 @@ { 200 , ALARM_ID_TD_TUBE_SET_MISMATCH , "TD | Wrong Tube Set Installed | The installed tube set does not match the selected treatment type. \n- Remove the current tube set.\n- Install the correct tube set for the selected treatment:\n HD or HD with Online Fluid requires an HD tube set.\n HDF requires an HDF tube set. | TD Tube Set Mismatch | TD installed tube set type does not match the selected treatment modality | If the scanned tube set type does not match the tube set type required for the selected treatment modality. | "}, { 201 , ALARM_ID_TD_PRE_TX_HEPARIN_LOAD_SYRINGE , "UI TD | Load Syringe | Fill Syringe with {140:2:mL:1} of heparin.\n- Connect heparin filled syringe to infusion port.\n- Insert syringe into the syringe pump.\n- Lock heparin syringe in place.\n- Unclamp heparin pinch clamp. | Process: Pre-Tx Load Syringe | User must load the heparin syringe before continuing heparin setup. | While in Pre-Treatment: Heparin Setup: Await Syringe Load Confirmation state. | "}, // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition - { 201 , ALARM_ID_FW_RESERVED_201 , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: RESERVED | RESERVED | RESERVED | "}, { 202 , ALARM_ID_FW_RESERVED_202 , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: RESERVED | RESERVED | RESERVED | "}, { 203 , ALARM_ID_FW_RESERVED_203 , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: RESERVED | RESERVED | RESERVED | "}, { 204 , ALARM_ID_FW_RESERVED_204 , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: RESERVED | RESERVED | RESERVED | "},