Index: AlarmDefs.h =================================================================== diff -u -r8cf94d51bbaedea33f8758fade8330cddbb5d34f -r7416b2a6c12f4d2c772a17c80a7f305bae909d8e --- AlarmDefs.h (.../AlarmDefs.h) (revision 8cf94d51bbaedea33f8758fade8330cddbb5d34f) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 7416b2a6c12f4d2c772a17c80a7f305bae909d8e) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) “rkallala” -* @date (last) 15-Jan-2026 +* @author (last) Varshini Nagabooshanam +* @date (last) 22-May-2026 * * @author (original) Sean Nash * @date (original) 01-Aug-2024 @@ -56,11 +56,11 @@ ALARM_ID_DD_VALVE_CONTROL_FAILURE = 25, ///< DD FPGA not accepting commanded valve states ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT = 26, ///< TD arterial pressure sensor freshness/error timeout fault ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT = 27, ///< TD venous pressure sensor freshness/error timeout fault - ALARM_ID_AVAILABLE_28 = 28, ///< Available alarms - ALARM_ID_AVAILABLE_29 = 29, ///< Available alarms - ALARM_ID_AVAILABLE_30 = 30, ///< Available alarms - ALARM_ID_AVAILABLE_31 = 31, ///< Available alarms - ALARM_ID_AVAILABLE_32 = 32, ///< Available alarms + ALARM_ID_TD_TMP_PRESSURE_LOW = 28, ///< TD TMP pressure low + ALARM_ID_TD_TMP_PRESSURE_HIGH = 29, ///< TD TMP pressure high + ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 30, ///< TD venous bubble self-test failure. + ALARM_ID_TD_BLOOD_PRIME_PAUSE_WARNING = 31, ///< TD blood prime pause warning + ALARM_ID_TD_BLOOD_PRIME_PAUSE_TIMEOUT = 32, ///< TD blood prime pause timeout ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE = 33, ///< TD voltage out of range fault ALARM_ID_TD_AC_POWER_LOST = 34, ///< TD AC power lost ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT = 35, ///< TD AC power lost during treatment @@ -86,7 +86,7 @@ ALARM_ID_DD_CONCENTRATE_PUMP_FAULT = 55, ///< DD concentrate pump fault ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE = 56, ///< DD temperature sensor out of range ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT = 57, ///< DD temperature sensors freshness timeout fault - ALARM_ID_AVAILABLE_58 = 58, ///< DD Available alarms + ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT = 58, ///< DD balancing chamber fill timeout fault ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC = 59, ///< DD barometric sensor coefficients bad CRC ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE = 60, ///< DD fresh dialysate pump RPM out of range ALARM_ID_DD_D12_PUMP_OFF_FAULT = 61, ///< DD fresh dialysate pump running when it is not commanded to run @@ -106,7 +106,7 @@ ALARM_ID_DD_SC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE = 75, ///< DD spent chamber state 1 fill pressure out of range ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE = 76, ///< DD spent chamber state 2 fill pressure out of range ALARM_ID_DD_SC_FILL_TIMEOUT_FAULT = 77, ///< DD spent chamber fill timeout fault - ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT = 78, ///< TD air trap fill timeout alarm during treatment + ALARM_ID_TD_AIR_TRAP_FILL_TIMEOUT = 78, ///< TD air trap fill timeout ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS = 79, ///< TD air trap illegal levels ALARM_ID_TD_PINCH_VALVE_FAULT = 80, ///< TD pinch valve fault ALARM_ID_TD_VALVE_HOMING_FAILED = 81, ///< TD pinch valve homing fault @@ -116,12 +116,12 @@ ALARM_ID_FP_COMM_TOO_MANY_BAD_CRCS = 85, ///< FP too many bad crcs ALARM_ID_FP_VALVE_CONTROL_FAILURE = 86, ///< FP valve control failure ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE = 87, ///< DD balance chamber state1 fill start pressure not in range - ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE = 88, ///< DD balance chamber state1 fill end pressure not in range + ALARM_ID_DD_AVAILABLE_1 = 88, ///< Available alarm ID 1 ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE = 89, ///< DD balance chamber state2 fill start pressure not in range - ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE = 90, ///< DD balance chamber state2 fill end pressure not in range + ALARM_ID_DD_AVAILABLE_2 = 90, ///< Available alarm ID 2 ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE = 91, ///< DD generate dialysate hydraulics chamber ngative pressure out of range ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT = 92, ///< DD spent chamber fill timeout fault - ALARM_ID_AVAILABLE_93 = 93, ///< Available alarms + ALARM_ID_DD_D79_RINSE_PUMP_SPEED_OUT_OF_RANGE = 93, ///< DD D79 rinse pump commanded vs measured speed out of tolerance ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE = 94, ///< DD balance chamber state1 fill initiated and pressure drop is not in range ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE = 95, ///< DD balance chamber state2 fill initiated and pressure drop is not in range ALARM_ID_TD_TREATMENT_STOPPED_BY_USER = 96, ///< TD treatment paused by user @@ -194,6 +194,37 @@ ALARM_ID_FP_PERMEATE_FLOW_OUT_HIGH_RANGE = 162, ///< FP permeate flow high fault ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE = 163, ///< FP permeate flow low fault ALARM_ID_FP_INLET_PRESSURE_LOW_RANGE = 164, ///< FP inlet pressure low fault + ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT = 165, ///< TD air trap lower timeout + ALARM_ID_TD_SYRINGE_PUMP_SELF_TEST_FAILURE = 166, ///< TD syringe pump self-test failure alarm + ALARM_ID_TD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR = 167, ///< TD syringe pump is running while the BP is off + ALARM_ID_TD_SYRINGE_PUMP_FPGA_ADC_FAULT = 168, ///< TD syringe pump ADC FPGA fault + ALARM_ID_TD_SYRINGE_PUMP_FAULT = 169, ///< TD syringe pump fault reported by FPGA + ALARM_ID_TD_SYRINGE_PUMP_DAC_WRITE_ERROR = 170, ///< TD syringe pump DAC write failure + ALARM_ID_TD_SYRINGE_PUMP_SPEED_ERROR = 171, ///< TD syringe pump speed check error + ALARM_ID_TD_SYRINGE_PUMP_VOLUME_ERROR = 172, ///< TD syringe pump volume check error + ALARM_ID_TD_SYRINGE_PUMP_OVER_TRAVEL_ERROR = 173, ///< TD syringe pump over travel error + ALARM_ID_TD_SYRINGE_PUMP_NOT_STOPPED_ERROR = 174, ///< TD syringe pump not stopped in off state error + ALARM_ID_TD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR = 175, ///< TD syringe pump direction (from encoder) error + ALARM_ID_TD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR = 176, ///< TD syringe pump direction (from controller) error + ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_EMPTY = 177, ///< TD syringe empty alarm + ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_REMOVED = 178, ///< TD syringe pump syringe removed alarm + ALARM_ID_TD_SYRINGE_PUMP_STALL = 179, ///< TD syringe pump stall alarm + ALARM_ID_TD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD = 180, ///< TD heparin force sensor invalid calibration record + ALARM_ID_TD_SYRINGE_DETECTED = 181, ///< TD syringe detected alarm + ALARM_ID_TD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM = 182, ///< TD syringe pump not enough Heparin alarm + ALARM_ID_TD_SYRINGE_PUMP_OCCLUSION = 183, ///< TD syringe pump occlusion alarm + ALARM_ID_TD_SYRINGE_PUMP_PRIME_TIMEOUT = 184, ///< TD syringe pump prime timeout + ALARM_ID_FP_P9_COND_SENSOR_FPGA_FAULT = 185, ///< FP P9 conductivity sensor fpga fault + ALARM_ID_FP_P18_COND_SENSOR_FPGA_FAULT = 186, ///< FP P18 conductivity sensor fpga fault + ALARM_ID_TD_AUTO_LOAD_EJECT_BP_TIMEOUT = 187, ///< TD blood pump timeout during auto-load\eject operation + ALARM_ID_TD_AUTO_EJECT_EJECTOR_TIMEOUT = 188, ///< TD ejector failed to extend or retract during auto-eject operation + ALARM_ID_TD_EJECTOR_POST_TEST_FAILED = 189, ///< TD ejector POST homing failure + ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE = 190, ///< TD high blood pump torque during auto-load\eject operation + ALARM_ID_TD_AIR_PUMP_STALL = 191, ///< TD air pump stall detected + ALARM_ID_TD_CPU_RAM_ERROR = 192, ///< TD CPU RAM error + ALARM_ID_DD_CPU_RAM_ERROR = 193, ///< DD CPU RAM error + ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED = 194, ///< TD integrity post test fault + ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED = 195, ///< TD integrity post test fault NUM_OF_ALARM_IDS ///< Number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -256,6 +287,7 @@ ALARM_USER_ACTION_RINSEBACK = 1, ///< User selected rinseback from alarm recovery options ALARM_USER_ACTION_END_TREATMENT = 2, ///< User selected end treatment from alarm recovery options ALARM_USER_ACTION_ACK = 3, ///< User selected acknowledge (ok) from alarm recovery options + ALARM_USER_ACTION_TEMPOARY_BREAK = 4, ///< User selected temporary break NUMBER_OF_ALARM_USER_ACTIONS ///< Number of alarm user actions }; typedef enum Alarm_User_Actions ALARM_USER_ACTION_T; ///< Type for alarm user actions enumeration @@ -269,7 +301,7 @@ ALARM_STATE_FLAG_BIT_POS_NO_RESUME = 3, ///< The "resume" user recovery option is disabled ALARM_STATE_FLAG_BIT_POS_NO_RINSEBACK = 4, ///< The "rinseback" user recovery option is disabled ALARM_STATE_FLAG_BIT_POS_NO_END_TREATMENT = 5, ///< The "end treatment" user recovery option is disabled - ALARM_STATE_FLAG_BIT_POS____AVAILABLE = 6, ///< Not used - available for use + ALARM_STATE_FLAG_BIT_POS_NO_TEMPORARY_BREAK = 6, ///< The "treatment break" user recovery option is disabled ALARM_STATE_FLAG_BIT_POS_OK_BUTTON_ONLY = 7, ///< The "ok" user recovery option is enabled ALARM_STATE_FLAG_BIT_POS_ALARMS_TO_ESCALATE = 8, ///< One or more active alarms will escalate in time ALARM_STATE_FLAG_BIT_POS_ALARMS_SILENCED = 9, ///< Alarms have been temporarily silenced by user @@ -358,15 +390,16 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FPGA_COMM_TIMEOUT }, { ALARM_PRIORITY_HIGH, 1, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_VALVE_CONTROL_FAILURE }, + // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_28 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_29 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_30 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_31 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_32 }, - { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 204, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_TMP_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 205, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_TMP_PRESSURE_HIGH }, + { ALARM_PRIORITY_LOW, 711, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 292, ALM_SRC_TD, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_BLOOD_PRIME_PAUSE_WARNING }, + { ALARM_PRIORITY_HIGH, 291, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_BLOOD_PRIME_PAUSE_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE }, { ALARM_PRIORITY_LOW , 852, ALM_SRC_TD, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_AC_POWER_LOST }, { ALARM_PRIORITY_HIGH, 100, ALM_SRC_TD, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D17_COND_SENSOR_FPGA_FAULT }, @@ -392,7 +425,7 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CONCENTRATE_PUMP_FAULT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_58 }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_OFF_FAULT }, @@ -413,7 +446,7 @@ // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_SC_FILL_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 13, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT }, + { ALARM_PRIORITY_HIGH, 13, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AIR_TRAP_FILL_TIMEOUT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_PINCH_VALVE_FAULT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VALVE_HOMING_FAILED }, @@ -423,12 +456,12 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_FP, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_COMM_TOO_MANY_BAD_CRCS }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_FP, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_VALVE_CONTROL_FAILURE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_AVAILABLE_1 }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_AVAILABLE_2 }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_93 }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D79_RINSE_PUMP_SPEED_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { ALARM_PRIORITY_MEDIUM, 300, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_TREATMENT_STOPPED_BY_USER }, @@ -437,31 +470,31 @@ { ALARM_PRIORITY_HIGH, 11, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, ALARM_ID_TD_BLOOD_SITTING_TOO_LONG }, { ALARM_PRIORITY_HIGH, 0, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_COMM_TIMEOUT }, // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_OS_VERSION }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SHASUM }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CANBUS }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_DISPLAY }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_TOUCH }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SDCARD }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_RTC }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_WIFI }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_BLUETOOTH }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_ETHERNET }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SOUND }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CLOUDSYNC }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_INVALID_YEAR }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SETTINGS_BAD }, - { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_EVNT_FAILURE_SDCARD }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_116 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_117 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_118 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_119 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_120 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_121 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_122 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_123 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_124 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_125 }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_OS_VERSION }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SHASUM }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CANBUS }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_DISPLAY }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_TOUCH }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SDCARD }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_RTC }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_WIFI }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_BLUETOOTH }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_ETHERNET }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SOUND }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CLOUDSYNC }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_INVALID_YEAR }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SETTINGS_BAD }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_UI, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_EVNT_FAILURE_SDCARD }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_116 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_117 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_118 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_119 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_120 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_121 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_122 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_123 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_124 }, + { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_125 }, // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_126 }, { ALARM_PRIORITY_LOW, 999, ALM_SRC_UI, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_127 }, @@ -494,15 +527,47 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_FP, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 150, ALM_SRC_DD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , ALARM_ID_DD_BLOOD_LEAK_DETECTED }, - { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, - { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, - { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, + { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, + { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, + { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, { ALARM_PRIORITY_MEDIUM, 698, ALM_SRC_DD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT, }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_FP, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_FP, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_INLET_PRESSURE_CRITICAL_OUT_RANGE }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_FP, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_PERMEATE_FLOW_OUT_HIGH_RANGE }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_FP, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_FP, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_INLET_PRESSURE_LOW_RANGE }, + { ALARM_PRIORITY_HIGH, 13, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_FPGA_ADC_FAULT }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_FAULT }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_DAC_WRITE_ERROR }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_SPEED_ERROR }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_VOLUME_ERROR }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_OVER_TRAVEL_ERROR }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_NOT_STOPPED_ERROR }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR }, + { ALARM_PRIORITY_LOW, 730, ALM_SRC_TD, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_EMPTY }, + // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID + { ALARM_PRIORITY_HIGH, 207, ALM_SRC_TD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_REMOVED }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_STALL }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD }, + { ALARM_PRIORITY_LOW, 745, ALM_SRC_TD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_DETECTED }, + { ALARM_PRIORITY_LOW, 715, ALM_SRC_TD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM }, + { ALARM_PRIORITY_LOW, 716, ALM_SRC_TD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_OCCLUSION }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SYRINGE_PUMP_PRIME_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_FP, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_P9_COND_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_FP, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_P18_COND_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_AUTO_LOAD_EJECT_BP_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_AUTO_EJECT_EJECTOR_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_EJECTOR_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_TD, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE }, + { ALARM_PRIORITY_HIGH, 206, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AIR_PUMP_STALL }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_CPU_RAM_ERROR }, + { ALARM_PRIORITY_HIGH, 10, ALM_SRC_DD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CPU_RAM_ERROR }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, ALM_SRC_DD, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED }, }; // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID /// Table of alarms and their rank. @@ -532,6 +597,10 @@ { 1, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, { 1, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, { 1, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, + { 1, ALARM_ID_TD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD }, + { 1, ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED }, + { 1, ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED }, + { 1, ALARM_ID_TD_EJECTOR_POST_TEST_FAILED }, { 10, ALARM_ID_TD_SOFTWARE_FAULT }, { 10, ALARM_ID_TD_UI_COMM_TIMEOUT }, { 10, ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS }, @@ -554,9 +623,27 @@ { 10, ALARM_ID_FP_PERMEATE_FLOW_OUT_HIGH_RANGE }, { 10, ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE }, { 10, ALARM_ID_FP_INLET_PRESSURE_LOW_RANGE }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_FPGA_ADC_FAULT }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_FAULT }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_DAC_WRITE_ERROR }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_SPEED_ERROR }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_VOLUME_ERROR }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_OVER_TRAVEL_ERROR }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_NOT_STOPPED_ERROR }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_STALL }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_PRIME_TIMEOUT }, + { 10, ALARM_ID_TD_SYRINGE_PUMP_SELF_TEST_FAILURE }, + { 10, ALARM_ID_TD_AUTO_LOAD_EJECT_BP_TIMEOUT }, + { 10, ALARM_ID_TD_AUTO_EJECT_EJECTOR_TIMEOUT }, + { 10, ALARM_ID_TD_CPU_RAM_ERROR }, + { 10, ALARM_ID_DD_CPU_RAM_ERROR }, { 11, ALARM_ID_TD_BLOOD_SITTING_TOO_LONG }, { 12, ALARM_ID_TD_VENOUS_BUBBLE_DETECTED }, - { 13, ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT }, + { 13, ALARM_ID_TD_AIR_TRAP_FILL_TIMEOUT }, + { 13, ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT }, { 100, ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT }, { 110, ALARM_ID_DD_CRITICAL_DATA_ERROR }, { 110, ALARM_ID_DD_WATCHDOG_EXPIRED }, @@ -571,6 +658,9 @@ { 110, ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT }, { 110, ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT }, { 110, ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT }, + { 110, ALARM_ID_FP_P9_COND_SENSOR_FPGA_FAULT }, + { 110, ALARM_ID_FP_P18_COND_SENSOR_FPGA_FAULT }, + { 110, ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT }, { 110, ALARM_ID_FP_FPGA_COMM_TIMEOUT }, { 110, ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR }, { 110, ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR }, @@ -593,9 +683,7 @@ { 110, ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON }, { 110, ALARM_ID_FP_COMM_TOO_MANY_BAD_CRCS }, { 110, ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE }, - { 110, ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, - { 110, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, { 110, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, @@ -615,13 +703,15 @@ { 110, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, { 110, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, { 110, ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR }, + { 110, ALARM_ID_DD_D79_RINSE_PUMP_SPEED_OUT_OF_RANGE }, { 110, ALARM_ID_DD_SC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE }, { 110, ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, { 110, ALARM_ID_DD_SC_FILL_TIMEOUT_FAULT }, { 110, ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT }, { 110, ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT }, { 110, ALARM_ID_TD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC }, { 110, ALARM_ID_TD_BARO_PRESSURE_OUT_OF_RANGE }, + { 110, ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE }, { 111, ALARM_ID_TD_DD_COMM_TIMEOUT }, { 111, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD }, { 111, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_FP }, @@ -632,8 +722,14 @@ { 201, ALARM_ID_TD_VENOUS_PRESSURE_HIGH }, { 202, ALARM_ID_TD_ARTERIAL_PRESSURE_LOW }, { 203, ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH }, + { 204, ALARM_ID_TD_TMP_PRESSURE_LOW }, + { 205, ALARM_ID_TD_TMP_PRESSURE_HIGH }, + { 206, ALARM_ID_TD_AIR_PUMP_STALL }, + { 207, ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_REMOVED }, { 210, ALARM_ID_TD_CARTRIDGE_DOOR_OPENED }, { 290, ALARM_ID_TD_BLOOD_SITTING_WARNING }, + { 291, ALARM_ID_TD_BLOOD_PRIME_PAUSE_TIMEOUT }, + { 292, ALARM_ID_TD_BLOOD_PRIME_PAUSE_WARNING }, { 300, ALARM_ID_TD_TREATMENT_STOPPED_BY_USER }, { 401, ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP }, { 401, ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP }, @@ -648,15 +744,15 @@ { 700, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, { 700, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, + { 711, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, + { 715, ALARM_ID_TD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM }, + { 716, ALARM_ID_TD_SYRINGE_PUMP_OCCLUSION }, + { 730, ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_EMPTY }, + { 745, ALARM_ID_TD_SYRINGE_DETECTED }, { 850, ALARM_ID_BATTERY_LOW }, { 852, ALARM_ID_TD_AC_POWER_LOST }, - { 999, ALARM_ID_AVAILABLE_28 }, - { 999, ALARM_ID_AVAILABLE_29 }, - { 999, ALARM_ID_AVAILABLE_30 }, - { 999, ALARM_ID_AVAILABLE_31 }, - { 999, ALARM_ID_AVAILABLE_32 }, - { 999, ALARM_ID_AVAILABLE_58 }, - { 999, ALARM_ID_AVAILABLE_93 }, + { 999, ALARM_ID_DD_AVAILABLE_1 }, + { 999, ALARM_ID_DD_AVAILABLE_2 }, { 999, ALARM_ID_UI_RESERVED_116 }, { 999, ALARM_ID_UI_RESERVED_117 }, { 999, ALARM_ID_UI_RESERVED_118 }, @@ -724,11 +820,11 @@ // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition { 26 , ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Arterial Pressure Sensor Error | TD arterial pressure sensor reporting error or not being read successfully | Arterial pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 27 , ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Venous Pressure Sensor Error | TD venous pressure sensor reporting error or not being read successfully | Venous pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, - { 28 , ALARM_ID_AVAILABLE_28 , " | | | | | | "}, - { 29 , ALARM_ID_AVAILABLE_29 , " | | | | | | "}, - { 30 , ALARM_ID_AVAILABLE_30 , " | | | | | | "}, - { 31 , ALARM_ID_AVAILABLE_31 , " | | | | | | "}, - { 32 , ALARM_ID_AVAILABLE_32 , " | | | | | | "}, + { 28 , ALARM_ID_TD_TMP_PRESSURE_LOW , "TD | Low TMP Pressure | TMP pressure is low | | | | "}, + { 29 , ALARM_ID_TD_TMP_PRESSURE_HIGH , "TD | High TMP Pressure | TMP pressure is high | | | | "}, + { 30 , ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE , "TD | Self Test: Venous Air Detector | A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. | Process: Pre- Tx Wet Self- Test | TD venous air bubble detector self-test failure | If the venous air bubble self test did not detect bubble for more than 500 msec | "}, + { 31 , ALARM_ID_TD_BLOOD_PRIME_PAUSE_WARNING , "TD | MAX Blood Prime Pause Time | The Blood Prime pause time has nearly reached the maximum limit.\n- You must resume Blood Prime before the pause timer expires.\n- If Blood Prime is not resumed before the timer expires,\nBlood Prime will be terminated and treatment will transition to a safe state. | Blood Prime Paused: Warning | Blood Prime pause approaching timeout (> 4 min). | During Blood Prime paused state, when the pause duration exceeds 4 minutes. | "}, + { 32 , ALARM_ID_TD_BLOOD_PRIME_PAUSE_TIMEOUT , "TD | MAX Blood Prime Pause Time Reached | The maximum Blood Prime pause time has been reached.\n- Blood Prime has been paused for more than the allowed duration.\n- Treatment must be restarted or handled according to system safety procedures. | Blood Prime Paused: Alarm | Blood Prime pause timeout reached (> 5 min). | During Blood Prime paused state, when the pause duration exceeds 5 minutes | "}, { 33 , ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Voltage Monitor | TD monitored voltage is out of range | If the TD monitored voltages are out of range for a certain period of time. | "}, { 34 , ALARM_ID_TD_AC_POWER_LOST , "TD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | TD AC power lost alarm | The TD Software shall trigger an alarm if an AC power loss is detected while the TD system is in standby, service, or treatment parameters mode. | "}, { 35 , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT , "TD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | TD AC power lost alarm | The TD Software shall trigger an alarm if an AC power loss is detected while the TD system is in pre-treatment, treatment, or post-treatment mode. | "}, @@ -755,7 +851,7 @@ { 55 , ALARM_ID_DD_CONCENTRATE_PUMP_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Pump Fault | DD concentrate pump fault | If the concentrate pumps fault is reported by the FPGA. | "}, { 56 , ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Temp Sensor Range | DD temperature sensor out of range | If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. | "}, { 57 , ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Temperature RTD Comm | DD temperature sensors not being read successfully | If the FPGA does not report fresh data for a certain period of time | "}, - { 58 , ALARM_ID_AVAILABLE_58 , " | | | | | | "}, + { 58 , ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD BC fill timeout Error | DD Balancing chamber fill timeout error | If BC fill time exceeds the allowed timeout based on the balancing chamber switching period. | "}, { 59 , ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Baro temp Cal CRC | DD barometric pressure sensor bad coefficients CRC | If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. | "}, { 60 , ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Fresh Dial Pump Speed Fault | DD fresh dialysate pump RPM out of range | If the fresh dialysate pump RPM > 300.0 of the target RPM. | "}, { 61 , ALARM_ID_DD_D12_PUMP_OFF_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Fresh Dial Pump Speed Not Off | DD fresh dialysate pump off fault | If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. | "}, @@ -776,7 +872,7 @@ // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition { 76 , ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD Spent chamber fill state 2 start pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 24 psig or > 28 psig) not in range for spent chamber fill state2 start. | "}, { 77 , ALARM_ID_DD_SC_FILL_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Spent Chamber fill timeout Error | DD Spent chamber fill timeout error | If the number of spent chamber fill > = 7 counts | "}, - { 78 , ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT , "TD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Fill | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "}, + { 78 , ALARM_ID_TD_AIR_TRAP_FILL_TIMEOUT , "TD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Fill | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "}, { 79 , ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Air Trap Sensor | Air trap level sensors reporting illegal combination of air/fluid | If the air trap level readings are not valid for more than 2 seconds. | "}, { 80 , ALARM_ID_TD_PINCH_VALVE_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Pinch Valve | Pinch valve reporting a fault | If the valve detects a driver, direction or encoder fault condition. | "}, { 81 , ALARM_ID_TD_VALVE_HOMING_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Valve Home | TD valve homing failed | If the number of failed homing > = 3 counts. | "}, @@ -786,12 +882,12 @@ { 85 , ALARM_ID_RO_COMM_TOO_MANY_BAD_CRCS , "RO | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | RO Fault: CAN CRC Comm | RO Too many bad CRCs detected on received system messages | If too many bad message CRCs are received within a set period of time | "}, { 86 , ALARM_ID_RO_VALVE_CONTROL_FAILURE , "RO | Service Required: Filtration Device | A problem was detected with the filtration device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | RO Fault: Valve Control | RO FPGA not accepting commanded valve states | If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. | "}, { 87 , ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill start pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state1 fill start. | "}, - { 88 , ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill end pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state1 fill end. | "}, + { 88 , ALARM_ID_DD_AVAILABLE_1 , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Alarm availabe | AVAILABLE | AVAILABLE. | "}, { 89 , ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill start pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill start. | "}, - { 90 , ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill end pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill end. | "}, + { 90 , ALARM_ID_DD_AVAILABLE_2 , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Alarm availabe | AVAILABLE | AVAILABLE. | "}, { 91 , ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Hyd Cham Neg Pres not in range | DD hydraulics chamber degassing pressure is not in range | If the hydraulics chamber pressure is not less than or equal to calculated target degas pressure psig. | "} { 92 , ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Bicarb Chamber fill timeout Error | DD Bicarb chamber fill timeout error | If Bicarb chamber fill time > = 1 seconds | "}, - { 93 , ALARM_ID_AVAILABLE_93 , " | | | | | | "}, + { 93 , ALARM_ID_DD_D79_RINSE_PUMP_SPEED_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Rinse Pump Speed Fault | DD D79 rinse pump commanded vs measured speed out of tolerance | If rinse pump measured RPM deviates beyond 10%% of target RPM when pump is commanded on. Data1=measured RPM, Data2=target RPM. | "}, { 94 , ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state1 fill in progress. | "}, { 95 , ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state2 fill in progress. | "}, { 96 , ALARM_ID_TD_TREATMENT_STOPPED_BY_USER , "TD | Blood Pump Stopped: Treatment Paused | Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. | Blood Stopped: Treatment Pause | TD treatment stopped by user action - pressed stop button | If the user has pressed the stop button. | "}, @@ -866,6 +962,38 @@ { 162 , ALARM_ID_FP_PERMEATE_FLOW_OUT_HIGH_RANGE , "FP | High Permeate flow | Permeate flow is too high. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Permeate flow High | Permeate flow is too high from set target flow rate | If the permeate flow is above 25% of set target flow rate for a certain period of time. | "}, { 163 , ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE , "FP | low Permeate flow | Permeate flow is too low. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Permeate flow Low | Permeate flow is too low from set target flow rate | If the permeate flow is above 25% of set target flow rate for a certain period of time. | "}, { 164 , ALARM_ID_FP_INLET_PRESSURE_LOW_RANGE , "FP | low inlet pressure | Inlet pressure is too low. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Inlet pressure Low | Inlet pressure is too low | If P8 is < 1 PSI | "}, + { 165 , ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT , "TD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Lowering | Air trap lowering timeout during treatment | If the air trap lowering timeout during treatment. | "}, + { 166 , ALARM_ID_TD_SYRINGE_PUMP_SELF_TEST_FAILURE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe DAC | TD syringe pump self-test failure alarm | If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. | "}, + { 167 , ALARM_ID_TD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump On Err | TD syringe pump is running while the BP is off | If the syringe pump is on while the blood pump is off | "}, + { 168 , ALARM_ID_TD_SYRINGE_PUMP_FPGA_ADC_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump ADC Com | TD syringe pump ADC error | If the ADC read is not fresh after POST init and the ADC read counter has not changed. | "}, + { 169 , ALARM_ID_TD_SYRINGE_PUMP_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump Driver | TD syringe pump fault reported by FPGA | If the FPGA reported the syringe pump fault | "}, + { 170 , ALARM_ID_TD_SYRINGE_PUMP_DAC_WRITE_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump DAC Err | TD syringe pump DAC write failure | If the syringe pump DAC write error is reported by the FPGA | "}, + { 171 , ALARM_ID_TD_SYRINGE_PUMP_SPEED_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump Speed | TD syringe pump speed check error | If the syringe measured rate is off by more than 5% or 0.1mL. | "}, + { 172 , ALARM_ID_TD_SYRINGE_PUMP_VOLUME_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump Vol | TD syringe pump volume check error | If the volume delivered is off by more than 5% or 0.1mL. | "}, + { 173 , ALARM_ID_TD_SYRINGE_PUMP_OVER_TRAVEL_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump OT | TD syringe pump over travel error | If the syringe pump position is deemed to have overtraveled | "}, + { 174 , ALARM_ID_TD_SYRINGE_PUMP_NOT_STOPPED_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump Speed | TD syringe pump not stopped in off state error | If the syringe pump position is not equal to its last position. | "}, + { 175 , ALARM_ID_TD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump Enc Dir | TD syringe pump direction (from encoder) error | If the measured encoder direction of the syringe pump does not agree with the expected direction | "}, + { 176 , ALARM_ID_TD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump Drv Dir | TD syringe pump direction (from controller) error | If the measured direction of the syringe pump motor controller does not agree with the expected direction | "}, + // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition + { 177 , ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_EMPTY , "TD | Heparin Syringe Empty | Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. | Heparin Syringe Empty | TD syringe empty alarm | If the syringe pump position is near empty position. | "}, + { 178 , ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_REMOVED , "TD | Blood Stopped: Syringe Not Detected | There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. | Blood Stop: Syringe Not Detected | TD syringe pump syringe removed alarm | If the syringe detection switch < 2.0 volts. | "}, + { 179 , ALARM_ID_TD_SYRINGE_PUMP_STALL , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Syringe Pump Stall | TD syringe pump stall alarm | If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. | "}, + { 180 , ALARM_ID_TD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD POST: SP DAC Cal | TD heparin force sensor invalid calibration record | If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. | "}, + { 181 , ALARM_ID_TD_SYRINGE_DETECTED , "TD | Syringe Detected | Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. | Process: Remove Syringe | TD syringe detected alarm | If the syringe detection switch > 2.0 volts. | "}, + { 182 , ALARM_ID_TD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM , "TD | Not Enough Heparin | There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. | Process: Pre- Tx Not Enough Hep. | TD syringe pump not enough Heparin alarm | If the estimated syringe volume needed for treatment is not sufficient | "}, + { 183 , ALARM_ID_TD_SYRINGE_PUMP_OCCLUSION , "TD | Heparin Pump Occlusion | A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. | Process: Pre- Tx Hep. Pump Occ. | TD syringe pump occlusion alarm | If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. | "}, + { 184 , ALARM_ID_TD_SYRINGE_PUMP_PRIME_TIMEOUT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Pre- Tx Hep. Pump Timeout. | TD syringe pump timeout alarm | If the syringe pump priming > 5 seconds | "}, + { 185 , ALARM_ID_FP_P9_COND_SENSOR_FPGA_FAULT , "FP | Service Required: FP Device | A problem was detected with the FP device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Conductivity Sensor Error | FP conductivity sensor (P9) located in inlet water path reporting error or not being read successfully | Conductvity Sensor (P9) is reporting an error or FPGA is failing to read sensor. | "}, + { 186 , ALARM_ID_FP_P18_COND_SENSOR_FPGA_FAULT , "FP | Service Required: FP Device | A problem was detected with the FP device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Conductivity Sensor Error | FP conductivity sensor (P18) located in inlet water path reporting error or not being read successfully | Conductvity Sensor (P18) is reporting an error or FPGA is failing to read sensor. | "}, + { 187 , ALARM_ID_TD_AUTO_LOAD_EJECT_BP_TIMEOUT , "TD | Service Required: Hemodialysis Device | The blood pump did not complete the required operation within the allowed time.\n- The blood pump rotor may be obstructed, or the Hall sensor may be faulty.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call Service to report the issue and schedule a repair. | TD Fault: BP Operation Timeout | Blood pump failed to complete homing or one full revolution during auto-load or auto-eject | If the blood pump fails to reach the home position or complete one full revolution within the allowed time during an auto-loading, or auto-eject homing, or auto-eject ejection operation. | "}, + { 188 , ALARM_ID_TD_AUTO_EJECT_EJECTOR_TIMEOUT , "TD | Service Required: Hemodialysis Device | The ejector did not reach the expected position within the allowed time.\n- The ejector motor may be faulty or optical sensors may be obstructed or faulty.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call Service to report the issue and schedule a repair. | TD Fault: Ejector Timeout | Ejector failed to reach extended or retracted position during auto-eject | If the ejector does not reach the fully extended or fully retracted position within the allowed time during the auto-eject sequence. | "}, + { 189 , ALARM_ID_TD_EJECTOR_POST_TEST_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call Service to report the issue and schedule a repair. | TD POST: Ejector Homing | TD ejector did not reach the fully retracted position within the allowed time during power-on self-test | If the ejector does not reach the fully retracted position within allowed time during the power-on self-test. | "}, + { 190 , ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE , "TD | Blood Pump Stopped: Tubing Set Jam | A problem was detected during tube set load/eject.\n- The blood pump is reversing to the home position.\n- Once the pump has returned to home, re-seat the blood tubing set and press Resume to retry loading, or wait for the pump to return to home and retry ejecting.\n- If the problem persists, press End Treatment. | Blood Stop: High BP Torque | Blood pump torque exceeded limit during Auto-Load/Eject one-revolution and tubing set jam suspected | If the measured blood pump torque exceeds TBD mNm during the Auto-Load or Auto-Eject one-revolution forward rotation. | "}, + { 191 , ALARM_ID_TD_AIR_PUMP_STALL , "TD | Air Pump Stall Detected | The air pump is commanded to run but no rotation is detected.\n- Check tubing for blockage or occlusion.\n- Verify pump operation and connections.\n- If issue persists, terminate treatment and inspect system. | Blood Pump: Air Pump Stall | Air pump stall detected during treatment | If the air pump is commanded on and the measured RPM is zero for multiple controller cycles. | "}, + { 192 , ALARM_ID_TD_CPU_RAM_ERROR , "TD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD CPU RAM error | TD CPU RAM check failure | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, + { 193 , ALARM_ID_DD_CPU_RAM_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD CPU RAM error | DD CPU RAM check failure | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, + { 194 , ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Firmware Image Integrity Failure | TD firmware image integrity POST test failed | If the calculated CRC of the TD image does not match the expected CRC. | "}, + { 195 , ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED , "DD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Firmware Image Integrity Failure | DD firmware image integrity POST test failed | If the calculated CRC of the DD image does not match the expected CRC. | "}, }; #endif // manual alarms