Index: AlarmDefs.h =================================================================== diff -u -ra8805fe07b822d849b495d7fb0242a45d83d9595 -r1cc7e202f5bc6994887727d510b5b2716f1a04e0 --- AlarmDefs.h (.../AlarmDefs.h) (revision a8805fe07b822d849b495d7fb0242a45d83d9595) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 1cc7e202f5bc6994887727d510b5b2716f1a04e0) @@ -8,7 +8,7 @@ * @file AlarmDefs.h * * @author (last) Varshini Nagabooshanam -* @date (last) 29-Jan-2026 +* @date (last) 17-Mar-2026 * * @author (original) Sean Nash * @date (original) 01-Aug-2024 @@ -58,9 +58,9 @@ ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT = 27, ///< TD venous pressure sensor freshness/error timeout fault ALARM_ID_TD_TMP_PRESSURE_LOW = 28, ///< TD TMP pressure low ALARM_ID_TD_TMP_PRESSURE_HIGH = 29, ///< TD TMP pressure high - ALARM_ID_AVAILABLE_30 = 30, ///< Available alarms - ALARM_ID_AVAILABLE_31 = 31, ///< Available alarms - ALARM_ID_AVAILABLE_32 = 32, ///< Available alarms + ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 30, ///< TD venous bubble self-test failure. + ALARM_ID_TD_BLOOD_PRIME_PAUSE_WARNING = 31, ///< TD blood prime pause warning + ALARM_ID_TD_BLOOD_PRIME_PAUSE_TIMEOUT = 32, ///< TD blood prime pause timeout ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE = 33, ///< TD voltage out of range fault ALARM_ID_TD_AC_POWER_LOST = 34, ///< TD AC power lost ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT = 35, ///< TD AC power lost during treatment @@ -86,7 +86,7 @@ ALARM_ID_DD_CONCENTRATE_PUMP_FAULT = 55, ///< DD concentrate pump fault ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE = 56, ///< DD temperature sensor out of range ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT = 57, ///< DD temperature sensors freshness timeout fault - ALARM_ID_AVAILABLE_58 = 58, ///< DD Available alarms + ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT = 58, ///< DD balancing chamber fill timeout fault ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC = 59, ///< DD barometric sensor coefficients bad CRC ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE = 60, ///< DD fresh dialysate pump RPM out of range ALARM_ID_DD_D12_PUMP_OFF_FAULT = 61, ///< DD fresh dialysate pump running when it is not commanded to run @@ -195,6 +195,7 @@ ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE = 163, ///< FP permeate flow low fault ALARM_ID_FP_INLET_PRESSURE_LOW_RANGE = 164, ///< FP inlet pressure low fault ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT = 165, ///< TD air trap lower timeout + ALARM_ID_TD_AIR_PUMP_STALL = 166, ///< TD air pump stall detected NUM_OF_ALARM_IDS ///< Number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -362,11 +363,11 @@ // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 204, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, ALARM_ID_TD_TMP_PRESSURE_LOW }, - { ALARM_PRIORITY_HIGH, 205, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, ALARM_ID_TD_TMP_PRESSURE_HIGH }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_30 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_31 }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_32 }, + { ALARM_PRIORITY_HIGH, 204, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_TMP_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 205, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_TMP_PRESSURE_HIGH }, + { ALARM_PRIORITY_LOW, 711, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 292, ALM_SRC_TD, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_BLOOD_PRIME_PAUSE_WARNING }, + { ALARM_PRIORITY_HIGH, 291, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_BLOOD_PRIME_PAUSE_TIMEOUT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, { ALARM_PRIORITY_LOW , 852, ALM_SRC_TD, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_AC_POWER_LOST }, { ALARM_PRIORITY_HIGH, 100, ALM_SRC_TD, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT }, @@ -393,7 +394,7 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CONCENTRATE_PUMP_FAULT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_58 }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_OFF_FAULT }, @@ -505,6 +506,7 @@ { ALARM_PRIORITY_HIGH, 10, ALM_SRC_FP, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_FP, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_INLET_PRESSURE_LOW_RANGE }, { ALARM_PRIORITY_HIGH, 13, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 206, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AIR_PUMP_STALL }, }; // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID /// Table of alarms and their rank. @@ -574,6 +576,7 @@ { 110, ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT }, { 110, ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT }, { 110, ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT }, + { 110, ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT }, { 110, ALARM_ID_FP_FPGA_COMM_TIMEOUT }, { 110, ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR }, { 110, ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR }, @@ -638,8 +641,11 @@ { 203, ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH }, { 204, ALARM_ID_TD_TMP_PRESSURE_LOW }, { 205, ALARM_ID_TD_TMP_PRESSURE_HIGH }, + { 206, ALARM_ID_TD_AIR_PUMP_STALL }, { 210, ALARM_ID_TD_CARTRIDGE_DOOR_OPENED }, { 290, ALARM_ID_TD_BLOOD_SITTING_WARNING }, + { 291, ALARM_ID_TD_BLOOD_PRIME_PAUSE_TIMEOUT }, + { 292, ALARM_ID_TD_BLOOD_PRIME_PAUSE_WARNING }, { 300, ALARM_ID_TD_TREATMENT_STOPPED_BY_USER }, { 401, ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP }, { 401, ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP }, @@ -654,12 +660,9 @@ { 700, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, { 700, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, + { 711, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { 850, ALARM_ID_BATTERY_LOW }, { 852, ALARM_ID_TD_AC_POWER_LOST }, - { 999, ALARM_ID_AVAILABLE_30 }, - { 999, ALARM_ID_AVAILABLE_31 }, - { 999, ALARM_ID_AVAILABLE_32 }, - { 999, ALARM_ID_AVAILABLE_58 }, { 999, ALARM_ID_UI_RESERVED_116 }, { 999, ALARM_ID_UI_RESERVED_117 }, { 999, ALARM_ID_UI_RESERVED_118 }, @@ -729,9 +732,9 @@ { 27 , ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Venous Pressure Sensor Error | TD venous pressure sensor reporting error or not being read successfully | Venous pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 28 , ALARM_ID_TD_TMP_PRESSURE_LOW , "TD | Low TMP Pressure | TMP pressure is low | | | | "}, { 29 , ALARM_ID_TD_TMP_PRESSURE_HIGH , "TD | High TMP Pressure | TMP pressure is high | | | | "}, - { 30 , ALARM_ID_AVAILABLE_30 , " | | | | | | "}, - { 31 , ALARM_ID_AVAILABLE_31 , " | | | | | | "}, - { 32 , ALARM_ID_AVAILABLE_32 , " | | | | | | "}, + { 30 , ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE , "TD | Self Test: Venous Air Detector | A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. | Process: Pre- Tx Wet Self- Test | TD venous air bubble detector self-test failure | If the venous air bubble self test did not detect bubble for more than 500 msec | "}, + { 31 , ALARM_ID_TD_BLOOD_PRIME_PAUSE_WARNING , "TD | MAX Blood Prime Pause Time | The Blood Prime pause time has nearly reached the maximum limit.\n- You must resume Blood Prime before the pause timer expires.\n- If Blood Prime is not resumed before the timer expires,\nBlood Prime will be terminated and treatment will transition to a safe state. | Blood Prime Paused: Warning | Blood Prime pause approaching timeout (> 4 min). | During Blood Prime paused state, when the pause duration exceeds 4 minutes. | "}, + { 32 , ALARM_ID_TD_BLOOD_PRIME_PAUSE_TIMEOUT , "TD | MAX Blood Prime Pause Time Reached | The maximum Blood Prime pause time has been reached.\n- Blood Prime has been paused for more than the allowed duration.\n- Treatment must be restarted or handled according to system safety procedures. | Blood Prime Paused: Alarm | Blood Prime pause timeout reached (> 5 min). | During Blood Prime paused state, when the pause duration exceeds 5 minutes | "}, { 33 , ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Voltage Monitor | TD monitored voltage is out of range | If the TD monitored voltages are out of range for a certain period of time. | "}, { 34 , ALARM_ID_TD_AC_POWER_LOST , "TD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | TD AC power lost alarm | The TD Software shall trigger an alarm if an AC power loss is detected while the TD system is in standby, service, or treatment parameters mode. | "}, { 35 , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT , "TD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | TD AC power lost alarm | The TD Software shall trigger an alarm if an AC power loss is detected while the TD system is in pre-treatment, treatment, or post-treatment mode. | "}, @@ -758,7 +761,7 @@ { 55 , ALARM_ID_DD_CONCENTRATE_PUMP_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Pump Fault | DD concentrate pump fault | If the concentrate pumps fault is reported by the FPGA. | "}, { 56 , ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Temp Sensor Range | DD temperature sensor out of range | If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. | "}, { 57 , ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Temperature RTD Comm | DD temperature sensors not being read successfully | If the FPGA does not report fresh data for a certain period of time | "}, - { 58 , ALARM_ID_AVAILABLE_58 , " | | | | | | "}, + { 58 , ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD BC fill timeout Error | DD Balancing chamber fill timeout error | If BC fill time exceeds the allowed timeout based on the balancing chamber switching period. | "}, { 59 , ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Baro temp Cal CRC | DD barometric pressure sensor bad coefficients CRC | If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. | "}, { 60 , ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Fresh Dial Pump Speed Fault | DD fresh dialysate pump RPM out of range | If the fresh dialysate pump RPM > 300.0 of the target RPM. | "}, { 61 , ALARM_ID_DD_D12_PUMP_OFF_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Fresh Dial Pump Speed Not Off | DD fresh dialysate pump off fault | If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. | "}, @@ -869,7 +872,8 @@ { 162 , ALARM_ID_FP_PERMEATE_FLOW_OUT_HIGH_RANGE , "FP | High Permeate flow | Permeate flow is too high. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Permeate flow High | Permeate flow is too high from set target flow rate | If the permeate flow is above 25% of set target flow rate for a certain period of time. | "}, { 163 , ALARM_ID_FP_PERMEATE_FLOW_OUT_LOW_RANGE , "FP | low Permeate flow | Permeate flow is too low. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Permeate flow Low | Permeate flow is too low from set target flow rate | If the permeate flow is above 25% of set target flow rate for a certain period of time. | "}, { 164 , ALARM_ID_FP_INLET_PRESSURE_LOW_RANGE , "FP | low inlet pressure | Inlet pressure is too low. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Inlet pressure Low | Inlet pressure is too low | If P8 is < 1 PSI | "}, - { 165 , ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT , "TD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Lowering | Air trap lowering timeout during treatment | If the air trap lowering timeout during treatment. | "}, + { 165 , ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT , "TD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Lowering | Air trap lowering timeout during treatment | If the air trap lowering timeout during treatment. | "}, + { 166 , ALARM_ID_TD_AIR_PUMP_STALL , "TD | Air Pump Stall Detected | The air pump is commanded to run but no rotation is detected.\n- Check tubing for blockage or occlusion.\n- Verify pump operation and connections.\n- If issue persists, terminate treatment and inspect system. | Blood Pump: Air Pump Stall | Air pump stall detected during treatment | If the air pump is commanded on and the measured RPM is zero for multiple controller cycles. | "}, }; #endif // manual alarms