Index: AlarmDefs.h =================================================================== diff -u -raae848c41d514914dd56c619dd6491b07e5b63ad -r8e318768900b7ac6ec1d214566e4029252b39d54 --- AlarmDefs.h (.../AlarmDefs.h) (revision aae848c41d514914dd56c619dd6491b07e5b63ad) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 8e318768900b7ac6ec1d214566e4029252b39d54) @@ -489,9 +489,9 @@ { ALARM_PRIORITY_HIGH, 110, TRUE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 150, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , ALARM_ID_DD_BLOOD_LEAK_DETECTED }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, + { ALARM_PRIORITY_HIGH, 2, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, + { ALARM_PRIORITY_HIGH, 2, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, + { ALARM_PRIORITY_HIGH, 2, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, { ALARM_PRIORITY_MEDIUM, 698, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT, }, { ALARM_PRIORITY_HIGH, 110, TRUE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT }, }; // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID @@ -849,6 +849,7 @@ { 158 , ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD POST: BLD Config Set Point | DD blood leak sensor set point set failure | If the set point of the blood leak sensor does not match the set point in the calibration record. | "}, { 159 , ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "DD | Blood Leak: Bypass | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. | Blood Stop: Dialyzer Blood Leak Bypass | DD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "}, { 160 , ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT , "FP | FP: Permeate Underfill | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Tank: Tank is not filling | FP tank level is too low | If the float sensor reads low for at least 5 seconds | "}, + }; #endif // manual alarms