Index: AlarmDefs.h =================================================================== diff -u -rc312551a9fd141133e3f0c7d2e4ddc2e5fa0359a -r90bc35dbb35f6ef05c3f2ea177fe0f75a9d9841f --- AlarmDefs.h (.../AlarmDefs.h) (revision c312551a9fd141133e3f0c7d2e4ddc2e5fa0359a) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 90bc35dbb35f6ef05c3f2ea177fe0f75a9d9841f) @@ -69,7 +69,18 @@ ALARM_ID_DD_CD3_SENSOR_FPGA_FAULT = 38, ///< DD CD3 conductivity sensor freshness/error timeout fault ALARM_ID_DD_CD4_SENSOR_FPGA_FAULT = 39, ///< DD CD4 conductivity sensor freshness/error timeout fault ALARM_ID_TD_VENOUS_BUBBLE_DETECTED = 40, ///< TD venous bubble detected - NUM_OF_ALARM_IDS ///< Total number of alarms + ALARM_ID_TD_CARTRIDGE_DOOR_OPENED = 41, ///< TD front door open + ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED = 42, ///< TD stuck button self test failure + ALARM_ID_TD_ARTERIAL_PRESSURE_LOW = 43, ///< TD arterial pressure low + ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH = 44, ///< TD arterial pressure high + ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE = 45, ///< TD arterial pressure out of range fault + ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE = 46, ///< TD arterial pressure self-test failure + ALARM_ID_TD_VENOUS_PRESSURE_LOW = 47, ///< TD venous pressure low + ALARM_ID_TD_VENOUS_PRESSURE_HIGH = 48, ///< TD venous pressure high + ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE = 49, ///< TD venous pressure out of range fault + ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE = 50, ///< TD venous pressure self-test failure + ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE = 51, ///< TD set pressure offset is out or range fault + NUM_OF_ALARM_IDS ///< Number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -243,6 +254,18 @@ { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CD3_SENSOR_FPGA_FAULT }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CD4_SENSOR_FPGA_FAULT }, { ALARM_PRIORITY_HIGH, 12, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED, }, + { ALARM_PRIORITY_HIGH, 210, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_CARTRIDGE_DOOR_OPENED }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 202, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_ARTERIAL_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 203, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 700, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 200, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_VENOUS_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 201, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_VENOUS_PRESSURE_HIGH }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 700, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, + // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID + { ALARM_PRIORITY_LOW, 710, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, }; // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID /// Table of alarms and their rank. @@ -252,6 +275,7 @@ { 1, ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED }, { 1, ALARM_ID_TD_UI_COMM_POST_FAILED }, { 1, ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE }, + { 1, ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED }, { 10, ALARM_ID_TD_SOFTWARE_FAULT }, { 10, ALARM_ID_TD_UI_COMM_TIMEOUT }, { 10, ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS }, @@ -264,6 +288,8 @@ { 10, ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT }, { 10, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, { 10, ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE }, + { 10, ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE }, + { 10, ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE }, { 12, ALARM_ID_TD_VENOUS_BUBBLE_DETECTED }, { 100, ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT }, { 110, ALARM_ID_DD_CRITICAL_DATA_ERROR }, @@ -287,6 +313,14 @@ { 111, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD }, { 111, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_RO }, { 111, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_TD }, + { 200, ALARM_ID_TD_VENOUS_PRESSURE_LOW }, + { 201, ALARM_ID_TD_VENOUS_PRESSURE_HIGH }, + { 202, ALARM_ID_TD_ARTERIAL_PRESSURE_LOW }, + { 203, ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH }, + { 210, ALARM_ID_TD_CARTRIDGE_DOOR_OPENED }, + { 700, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, + { 700, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, + { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, { 852, ALARM_ID_TD_AC_POWER_LOST }, { 999, ALARM_ID_NO_ALARM }, }; @@ -345,15 +379,27 @@ { 30 , ALARM_ID_DD_SPENT_DIALYSATE_PRES_TIMEOUT_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Spent Dial Pressure Sensor Error | DD spent dialysate pressure sensor reporting error or not being read successfully | Spent dialysate pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 31 , ALARM_ID_DD_FRESH_DIALYSATE_PRES_TIMEOUT_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fresh Dial Pressure Sensor Error | DD fresh dialysate pressure sensor reporting error or not being read successfully | Fresh dialysate pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 32 , ALARM_ID_DD_TRANSMEMB_PRES_TIMEOUT_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Transmemb Pressure Sensor Error | DD transmembrane sensor reporting error or not being read successfully | Transmembrane pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, - { 33 , ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | HD Fault: Voltage Monitor | HD monitored voltage is out of range | If the HD monitored voltages are out of range for a certain period of time. | "}, - { 34 , ALARM_ID_TD_AC_POWER_LOST , "TD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. | "}, - { 35 , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT , "TD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | HD AC power lost alarm | The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. | "}, + { 33 , ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Voltage Monitor | TD monitored voltage is out of range | If the TD monitored voltages are out of range for a certain period of time. | "}, + { 34 , ALARM_ID_TD_AC_POWER_LOST , "TD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | TD AC power lost alarm | The TD Software shall trigger an alarm if an AC power loss is detected while the TD system is in standby, service, or treatment parameters mode. | "}, + { 35 , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT , "TD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | TD AC power lost alarm | The TD Software shall trigger an alarm if an AC power loss is detected while the TD system is in pre-treatment, treatment, or post-treatment mode. | "}, { 36 , ALARM_ID_DD_CD1_SENSOR_FPGA_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (CD1) Error | DD conductivity sensor located before balancing chamber reporting error or not being read successfully | Conductvity Sensor (CD1) is reporting an error or FPGA is failing to read sensor. | "}, { 37 , ALARM_ID_DD_CD2_SENSOR_FPGA_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (CD2) Error | DD conductivity sensor located after balancing chamber reporting error or not being read successfully | Conductvity Sensor (CD2) is reporting an error or FPGA is failing to read sensor. | "}, { 38 , ALARM_ID_DD_CD3_SENSOR_FPGA_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (CD3) Error | DD conductivity sensor located after balancing chamber reporting error or not being read successfully | Conductvity Sensor (CD3) is reporting an error or FPGA is failing to read sensor. | "}, { 39 , ALARM_ID_DD_CD4_SENSOR_FPGA_FAULT , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (CD4) Error | DD conductivity sensor located in spent dialysate path reporting error or not being read successfully | Conductvity Sensor (CD4) is reporting an error or FPGA is failing to read sensor. | "}, - { 40 , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED , "TD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | HD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, -}; + { 40 , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED , "TD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | TD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, + { 41 , ALARM_ID_TD_CARTRIDGE_DOOR_OPENED , "TD | Blood Pump Stopped: Door Open | The device door is open.\n Close the device door. \n- Press Resume/OK to continue. | Blood Stop: Door Open | TD cartridge door opened alarm | If the cartridge door is opened as reported by the FPGA. | "}, + { 42 , ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED , "TD | Stuck Button | A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. | TD POST: Stuck Button | TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up | Stop or Off button are pressed for more than 1 sec during POST. | "}, + { 43 , ALARM_ID_TD_ARTERIAL_PRESSURE_LOW , "TD | Low Arterial Pressure | Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. Low | Arterial pressure too low during treatment | If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 44 , ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH , "TD | High Arterial Pressure | Arterial pressure is high.\n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. | Blood Stop: Arterial Pres. High | Arterial pressure too high during treatment | If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 45 , ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Arterial Press Range | TD arterial pressure sensor is reading out of range | If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, + { 46 , ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE , "TD | Self Test: Cartridge | A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | TD arterial pressure self-test failure alarm | If arterial pressure < -300 or > 100 mmHg | "}, + { 47 , ALARM_ID_TD_VENOUS_PRESSURE_LOW , "TD | Low Venous Pressure | Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access issues, \nif applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. Low | Venous pressure too low during treatment | If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 48 , ALARM_ID_TD_VENOUS_PRESSURE_HIGH , "TD | High Venous Pressure | Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. | Blood Stop: Venous Pres. High | Venous pressure too high during treatment | If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. | "}, + { 49 , ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Venous Press Range | TD venous pressure sensor is reading out of range | If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. | "}, + { 50 , ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE , "TD | Self Test: Cartridge | A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | TD venous pressure self-test failure alarm | If the venous pressure < -100 or > 600 mmHg | "}, + // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition + { 51 , ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "TD | Self Test: Cartridge Check | A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. | Process: Pre- Tx Dry Self- Test | TD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, + }; #endif // manual alarms