Index: AlarmDefs.h =================================================================== diff -u -rcc2feba726778c1bb89b70956a11832e7158af1c -rc2783b8068560b89b6bc91c14e46265974e0ca73 --- AlarmDefs.h (.../AlarmDefs.h) (revision cc2feba726778c1bb89b70956a11832e7158af1c) +++ AlarmDefs.h (.../AlarmDefs.h) (revision c2783b8068560b89b6bc91c14e46265974e0ca73) @@ -102,10 +102,10 @@ ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON = 71, ///< DD trimmer heater is On when no fluid is present ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT = 72, ///< DD water inlet input pressure sensor freshness/error timeout fault ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT = 73, ///< DD water inlet output pressure sensor freshness/error timeout fault - ALARM_ID_DD_INLET_WATER_PRESSURE_IN_LOW_RANGE = 74, ///< DD inlet water pressure input low range - ALARM_ID_DD_INLET_WATER_PRESSURE_IN_HIGH_RANGE = 75, ///< DD inlet water pressure input high range - ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_LOW_RANGE = 76, ///< DD inlet water pressure output low range - ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_HIGH_RANGE = 77, ///< DD inlet water pressure output high range + ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR = 74, ///< DD UF pump speed control error + ALARM_ID_AVAILABLE_75 = 75, ///< Available alarms + ALARM_ID_AVAILABLE_76 = 76, ///< Available alarms + ALARM_ID_AVAILABLE_77 = 77, ///< Available alarms ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT = 78, ///< TD air trap fill timeout alarm during treatment ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS = 79, ///< TD air trap illegal levels ALARM_ID_TD_PINCH_VALVE_FAULT = 80, ///< TD pinch valve fault @@ -174,7 +174,7 @@ ALARM_ID_FP_INLET_PRESSURE_OUT_HIGH_RANGE = 142, ///< FP inlet pressure high ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_HIGH_RANGE = 143, ///< FP inlet conductivity high ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_LOW_RANGE = 144, ///< FP inlet conductivty low - ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE = 145, ///< RO pump inlet pressure high + ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE = 145, ///< FP pump inlet pressure high ALARM_ID_FP_RO_OUTLET_CONDUCTIVITY_HIGH_RANGE = 146, ///< FP pump outlet conductivity high ALARM_ID_FP_RO_INLET_CONDUCTIVITY_HIGH_RANGE = 147, ///< FP pump inlet conductivity high ALARM_ID_FP_REJECTION_RATIO_LOW_RANGE = 148, ///< FP rejection ratio low @@ -300,6 +300,7 @@ BOOL alarmBlockEndTx; ///< Alarm should not trigger while in post-treatment mode BOOL alarmNoBloodRecirc; ///< Alarm prevents blood re-circulation BOOL alarmNoDialysateRecirc; ///< Alarm prevents dialysate re-circulation + BOOL alarmAutoResume; ///< Alarm should resume automatically once the condition is cleared BOOL alarmClearOnly; ///< Clear only this alarm on user acknowledgment BOOL alarmTreatmentLog; ///< Log alarm to treatment log if active during treatment ALARM_ID_T alarmID; ///< Alarm ID @@ -319,169 +320,169 @@ #ifdef __ALARM_MGMT_C__ /// Table of alarms and their static properties. const ALARM_T ALARM_TABLE[] = -{ // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID - { ALARM_PRIORITY_NONE, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NO_ALARM }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_SOFTWARE_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_SOFTWARE_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_FP_SOFTWARE_FAULT }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_FPGA_POST_TEST_FAILED }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_UI_COMM_POST_FAILED }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_FPGA_CLOCK_SPEED_CHECK_FAILURE }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_FPGA_COMM_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_UI_COMM_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 111, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_TD_DD_COMM_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_UI }, - { ALARM_PRIORITY_HIGH, 111, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_CRITICAL_DATA_ERROR }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CRITICAL_DATA_ERROR }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_WATCHDOG_EXPIRED }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_WATCHDOG_EXPIRED }, - { ALARM_PRIORITY_HIGH, 10, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_DD_COMM_TOO_MANY_BAD_CRCS }, - { ALARM_PRIORITY_HIGH, 111, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_FP }, - { ALARM_PRIORITY_HIGH, 111, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_TD }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FPGA_POST_TEST_FAILED }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FPGA_CLOCK_SPEED_CHECK_FAILURE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FPGA_COMM_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_VALVE_CONTROL_FAILURE }, - // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_HYD_OUTLET_PRES_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BIBAG_PRES_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_SPENT_DIALYSATE_PRES_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FRESH_DIALYSATE_PRES_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_TRANSMEMB_PRES_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, - { ALARM_PRIORITY_LOW , 852, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_AC_POWER_LOST }, - { ALARM_PRIORITY_HIGH, 100, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D17_COND_SENSOR_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 12, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED }, - { ALARM_PRIORITY_HIGH, 210, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_CARTRIDGE_DOOR_OPENED }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED }, - { ALARM_PRIORITY_HIGH, 202, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_ARTERIAL_PRESSURE_LOW }, - { ALARM_PRIORITY_HIGH, 203, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_LOW, 700, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, - { ALARM_PRIORITY_HIGH, 200, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_VENOUS_PRESSURE_LOW }, - { ALARM_PRIORITY_HIGH, 201, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_VENOUS_PRESSURE_HIGH }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_LOW, 700, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, - // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID - { ALARM_PRIORITY_LOW, 710, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_CONCENTRATE_PUMP_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BARO_SENSOR_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D12_PUMP_OFF_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D12_PUMP_CURRENT_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D12_PUMP_DIRECTION_INVALID }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D12_PUMP_DIRECTION_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D48_PUMP_RPM_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D48_PUMP_OFF_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D48_PUMP_CURRENT_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D48_PUMP_DIRECTION_INVALID }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D48_PUMP_DIRECTION_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT }, - { ALARM_PRIORITY_MEDIUM, 431, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_DD_INLET_WATER_PRESSURE_IN_LOW_RANGE }, - { ALARM_PRIORITY_MEDIUM, 430, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_DD_INLET_WATER_PRESSURE_IN_HIGH_RANGE }, - // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID - { ALARM_PRIORITY_MEDIUM, 431, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_LOW_RANGE }, - { ALARM_PRIORITY_MEDIUM, 430, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_HIGH_RANGE }, - { ALARM_PRIORITY_HIGH, 13, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_PINCH_VALVE_FAULT }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_VALVE_HOMING_FAILED }, - { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_VALVE_TRANSITION_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 111, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_FP_CAN_MESSAGE_NOT_ACKED_BY_DD }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_FP_FPGA_COMM_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_FP_COMM_TOO_MANY_BAD_CRCS }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_FP_VALVE_CONTROL_FAILURE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_92 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_93 }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, - { ALARM_PRIORITY_MEDIUM, 300, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_TREATMENT_STOPPED_BY_USER }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT }, - { ALARM_PRIORITY_HIGH, 290, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_BLOOD_SITTING_WARNING }, - { ALARM_PRIORITY_HIGH, 11, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_BLOOD_SITTING_TOO_LONG }, - { ALARM_PRIORITY_HIGH, 0, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_TD_COMM_TIMEOUT }, - // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_OS_VERSION }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SHASUM }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CANBUS }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_DISPLAY }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_TOUCH }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SDCARD }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_RTC }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_WIFI }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_BLUETOOTH }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_ETHERNET }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SOUND }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CLOUDSYNC }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_INVALID_YEAR }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SETTINGS_BAD }, - { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_EVNT_FAILURE_SDCARD }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_116 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_117 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_118 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_119 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_120 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_121 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_122 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_123 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_124 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_125 }, - // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_126 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_127 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_128 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_129 }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_UI_RESERVED_130 }, - { ALARM_PRIORITY_LOW, 850, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_BATTERY_LOW }, - { ALARM_PRIORITY_MEDIUM, 401, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP }, - { ALARM_PRIORITY_MEDIUM, 401, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP }, - { ALARM_PRIORITY_MEDIUM, 402, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_DD_DIALYSATE_TEMP_BELOW_TARGET_TEMP }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_PRESSURE_RELIEF_OUT_LOW_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_PRESSURE_RELIEF_OUT_HIGH_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_POWER_BUDGET_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_FLOW_RATE_LOW_RANGE }, - { ALARM_PRIORITY_MEDIUM, 432, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_FP_INLET_TEMPERATURE_OUT_LOW_RANGE }, - { ALARM_PRIORITY_MEDIUM, 433, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_FP_INLET_TEMPERATURE_OUT_HIGH_RANGE }, - { ALARM_PRIORITY_MEDIUM, 434, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_FP_INLET_PRESSURE_OUT_LOW_RANGE }, - { ALARM_PRIORITY_MEDIUM, 435, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_FP_INLET_PRESSURE_OUT_HIGH_RANGE }, - { ALARM_PRIORITY_MEDIUM, 436, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_HIGH_RANGE }, - { ALARM_PRIORITY_MEDIUM, 437, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_LOW_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_OUTLET_CONDUCTIVITY_HIGH_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_INLET_CONDUCTIVITY_HIGH_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_REJECTION_RATIO_LOW_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_INLET_PRESSURE_REGULATOR_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_FP_FILTER_HEALTH_PRESSURE_OUT_OF_RANGE }, - // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID - { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_FP_PERMEATE_FLUSH_CONDUCTIVITY_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, TRUE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, -}; // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr ClrOnly TxLog ID +{ // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID + { ALARM_PRIORITY_NONE, 999, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_NO_ALARM }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_SOFTWARE_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_SOFTWARE_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_SOFTWARE_FAULT }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_FPGA_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_UI_COMM_POST_FAILED }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_FPGA_CLOCK_SPEED_CHECK_FAILURE }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_FPGA_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_UI_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 111, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_DD_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_UI }, + { ALARM_PRIORITY_HIGH, 111, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_CRITICAL_DATA_ERROR }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CRITICAL_DATA_ERROR }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_WATCHDOG_EXPIRED }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_WATCHDOG_EXPIRED }, + { ALARM_PRIORITY_HIGH, 10, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_COMM_TOO_MANY_BAD_CRCS }, + { ALARM_PRIORITY_HIGH, 111, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_FP }, + { ALARM_PRIORITY_HIGH, 111, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_TD }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FPGA_POST_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FPGA_CLOCK_SPEED_CHECK_FAILURE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FPGA_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_VALVE_CONTROL_FAILURE }, + // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_HYD_OUTLET_PRES_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BIBAG_PRES_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_SPENT_DIALYSATE_PRES_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FRESH_DIALYSATE_PRES_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_TRANSMEMB_PRES_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, + { ALARM_PRIORITY_LOW , 852, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_AC_POWER_LOST }, + { ALARM_PRIORITY_HIGH, 100, FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D17_COND_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 12, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED }, + { ALARM_PRIORITY_HIGH, 210, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_CARTRIDGE_DOOR_OPENED }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED }, + { ALARM_PRIORITY_HIGH, 202, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_ARTERIAL_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 203, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 700, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 200, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_VENOUS_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 201, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_VENOUS_PRESSURE_HIGH }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 700, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, + // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID + { ALARM_PRIORITY_LOW, 710, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CONCENTRATE_PUMP_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BARO_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_OFF_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_CURRENT_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_DIRECTION_INVALID }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_DIRECTION_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D48_PUMP_RPM_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D48_PUMP_OFF_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D48_PUMP_CURRENT_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D48_PUMP_DIRECTION_INVALID }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D48_PUMP_DIRECTION_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_75 }, + // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_76 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_77 }, + { ALARM_PRIORITY_HIGH, 13, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_PINCH_VALVE_FAULT }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VALVE_HOMING_FAILED }, + { ALARM_PRIORITY_HIGH, 10, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VALVE_TRANSITION_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 111, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_CAN_MESSAGE_NOT_ACKED_BY_DD }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_FPGA_COMM_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_COMM_TOO_MANY_BAD_CRCS }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_VALVE_CONTROL_FAILURE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_92 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_93 }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, + { ALARM_PRIORITY_MEDIUM, 300, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_TREATMENT_STOPPED_BY_USER }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT }, + { ALARM_PRIORITY_HIGH, 290, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_BLOOD_SITTING_WARNING }, + { ALARM_PRIORITY_HIGH, 11, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, ALARM_ID_TD_BLOOD_SITTING_TOO_LONG }, + { ALARM_PRIORITY_HIGH, 0, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_COMM_TIMEOUT }, + // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_OS_VERSION }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SHASUM }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CANBUS }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_DISPLAY }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_TOUCH }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SDCARD }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_RTC }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_WIFI }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_BLUETOOTH }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_ETHERNET }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SOUND }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_CLOUDSYNC }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_INVALID_YEAR }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_POST_FAILURE_SETTINGS_BAD }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_EVNT_FAILURE_SDCARD }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_116 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_117 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_118 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_119 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_120 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_121 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_122 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_123 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_124 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_125 }, + // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_126 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_127 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_128 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_129 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_UI_RESERVED_130 }, + { ALARM_PRIORITY_LOW, 850, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_BATTERY_LOW }, + { ALARM_PRIORITY_MEDIUM, 401, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP }, + { ALARM_PRIORITY_MEDIUM, 401, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP }, + { ALARM_PRIORITY_MEDIUM, 402, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_DIALYSATE_TEMP_BELOW_TARGET_TEMP }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_PRESSURE_RELIEF_OUT_LOW_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_PRESSURE_RELIEF_OUT_HIGH_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_POWER_BUDGET_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_FLOW_RATE_LOW_RANGE }, + { ALARM_PRIORITY_MEDIUM, 432, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_TEMPERATURE_OUT_LOW_RANGE }, + { ALARM_PRIORITY_MEDIUM, 433, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_TEMPERATURE_OUT_HIGH_RANGE }, + { ALARM_PRIORITY_MEDIUM, 434, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_PRESSURE_OUT_LOW_RANGE }, + { ALARM_PRIORITY_MEDIUM, 435, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_PRESSURE_OUT_HIGH_RANGE }, + { ALARM_PRIORITY_MEDIUM, 436, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_HIGH_RANGE }, + { ALARM_PRIORITY_MEDIUM, 437, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_LOW_RANGE }, + { ALARM_PRIORITY_HIGH, 110, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_INLET_PRESSURE_OUT_HIGH_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_OUTLET_CONDUCTIVITY_HIGH_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_RO_INLET_CONDUCTIVITY_HIGH_RANGE }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_REJECTION_RATIO_LOW_RANGE }, + { ALARM_PRIORITY_HIGH, 110, TRUE , TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_FP_INLET_PRESSURE_REGULATOR_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_FILTER_HEALTH_PRESSURE_OUT_OF_RANGE }, + // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID + { ALARM_PRIORITY_HIGH, 110, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_PERMEATE_FLUSH_CONDUCTIVITY_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 110, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 110, TRUE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, +}; // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID /// Table of alarms and their rank. const ALARM_RANK_T ALARM_RANK_TABLE[] = @@ -593,6 +594,7 @@ { 110, ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE }, { 110, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, { 110, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, + { 110, ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR }, { 111, ALARM_ID_TD_DD_COMM_TIMEOUT }, { 111, ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD }, { 111, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_FP }, @@ -608,10 +610,6 @@ { 401, ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP }, { 401, ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP }, { 402, ALARM_ID_DD_DIALYSATE_TEMP_BELOW_TARGET_TEMP }, - { 430, ALARM_ID_DD_INLET_WATER_PRESSURE_IN_HIGH_RANGE }, - { 431, ALARM_ID_DD_INLET_WATER_PRESSURE_IN_LOW_RANGE }, - { 430, ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_HIGH_RANGE }, - { 431, ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_LOW_RANGE }, { 432, ALARM_ID_FP_INLET_TEMPERATURE_OUT_LOW_RANGE }, { 433, ALARM_ID_FP_INLET_TEMPERATURE_OUT_HIGH_RANGE }, { 434, ALARM_ID_FP_INLET_PRESSURE_OUT_LOW_RANGE }, @@ -623,6 +621,9 @@ { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, { 850, ALARM_ID_BATTERY_LOW }, { 852, ALARM_ID_TD_AC_POWER_LOST }, + { 999, ALARM_ID_AVAILABLE_75 }, + { 999, ALARM_ID_AVAILABLE_76 }, + { 999, ALARM_ID_AVAILABLE_77 }, { 999, ALARM_ID_AVAILABLE_92 }, { 999, ALARM_ID_AVAILABLE_93 }, { 999, ALARM_ID_UI_RESERVED_116 }, @@ -739,11 +740,11 @@ { 71 , ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Heater On With No Fluid | DD D45 heater is On with low fluid level | If the Balance air separation chamber fluid level is low for a certain period of time. | "}, { 72 , ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Water Inlet Input Press Sensor Error | DD water inlet input pressure sensor reporting error or not being read successfully | Water Inlet Input pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 73 , ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Water Inlet Output Press Sensor Error | DD water inlet output pressure sensor reporting error or not being read successfully | Water Inlet Output pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, - { 74 , ALARM_ID_DD_INLET_WATER_PRESSURE_IN_LOW_RANGE , "DD | Incoming Water Pressure Input Low | Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | DD: Inlet Water Pres Input Low | DD Inlet water pressure input too low | Inlet water pressure input < 25 PSI. | "}, - { 75 , ALARM_ID_DD_INLET_WATER_PRESSURE_IN_HIGH_RANGE , "DD | Incoming Water Pressure Input High | Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | DD: Inlet Water Pres Input High | DD inlet water pressure input is too high | If the inlet water pressure input is > 80 psig. | "}, + { 74 , ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Pump Speed Fault | DD UF pump D76_PUMP speed control error | If the Ultrafiltration pump D76_PUMP speed > 0.02 tolerance for a certain period of time. | "}, + { 75 , ALARM_ID_AVAILABLE_75 , " | | | | | | "}, // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition - { 76 , ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_LOW_RANGE , "DD | Incoming Water Pressure Output Low | Water inlet pressure after regulator is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | DD: Inlet Water Pres Output Low | DD Inlet water pressure output too low | Inlet water pressure output < 11 PSI. | "}, - { 77 , ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_HIGH_RANGE , "DD | Incoming Water Pressure Output High | Water inlet pressure after regulator is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | DD: Inlet Water Pres Output High | DD inlet water pressure output is too high | If the inlet water pressure output is > 20 psig. | "}, + { 76 , ALARM_ID_AVAILABLE_76 , " | | | | | | "}, + { 77 , ALARM_ID_AVAILABLE_77 , " | | | | | | "}, { 78 , ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT , "TD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Fill | Air trap fill timeout during treatment | If the air trap fill timeout during treatment. | "}, { 79 , ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Air Trap Sensor | Air trap level sensors reporting illegal combination of air/fluid | If the air trap level readings are not valid for more than 2 seconds. | "}, { 80 , ALARM_ID_TD_PINCH_VALVE_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Pinch Valve | Pinch valve reporting a fault | If the valve detects a driver, direction or encoder fault condition. | "}, @@ -820,8 +821,8 @@ { 149 , ALARM_ID_FP_POWER_BUDGET_OUT_OF_RANGE , "FP | Service Required: FP Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Power usage has exceeded budget| FP device is using too much power | If the device current is greater that its allowed percentage | "}, // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition { 150 , ALARM_ID_FP_FILTER_HEALTH_PRESSURE_OUT_OF_RANGE , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Pressure difference too high | FP pressure difference between the filters is too high | If the difference between M3 and P8 is ≥ 10 PSI after 2 minutes | "}, - { 151 , ALARM_ID_FP_MEMBRANE_FLUSH_CONDUCTIVITY_TIMEOUT , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Permeate conductivity high | FP conductivity after flushing the system is too high | If P18 conductivity > 200 uS/cm after 120 seconds ± 150ms of flushing. | "}, - { 152 , ALARM_ID_FP_MEMBRANE_FLUSH_VOLUME_OUT_OF_RANGE , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Volume difference too high | FP volume difference between integrated flow and expected volume | If P7 is <90% of the target integrated volume ( 1L ) after 30 seconds ± 150ms of flushing. | "}, + { 151 , ALARM_ID_FP_PERMEATE_FLUSH_CONDUCTIVITY_TIMEOUT , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Permeate conductivity high | FP conductivity after flushing the system is too high | If P18 conductivity > 200 uS/cm after 120 seconds ± 150ms of flushing. | "}, + { 152 , ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Volume difference too high | FP volume difference between integrated flow and expected volume | If P7 is <90% of the target integrated volume ( 1L ) after 30 seconds ± 150ms of flushing. | "}, { 153 , ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT , "FP | FP: Permeate Overfill | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Tank: Tank is too full | FP tank level is too high | If the float sensor reads high for at least 60 seconds | "}, { 154 , ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: D28 and D30 Delta | DD dialysate temperature sensors drift timeout. | If D28 and D30 have drifted for more 1 C for a certain period of time. | "}, };