Index: AlarmDefs.h =================================================================== diff -u -re3885d5c8a6bc1f819c024742a36a3e7f7abe232 -rbf2726e707c57ae7acd604d1c108a598be5d5c37 --- AlarmDefs.h (.../AlarmDefs.h) (revision e3885d5c8a6bc1f819c024742a36a3e7f7abe232) +++ AlarmDefs.h (.../AlarmDefs.h) (revision bf2726e707c57ae7acd604d1c108a598be5d5c37) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) Sameer Kalliadan Poyil -* @date (last) 02-Mar-2026 +* @author (last) Varshini Nagabooshanam +* @date (last) 13-Mar-2026 * * @author (original) Sean Nash * @date (original) 01-Aug-2024 @@ -58,7 +58,7 @@ ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT = 27, ///< TD venous pressure sensor freshness/error timeout fault ALARM_ID_TD_TMP_PRESSURE_LOW = 28, ///< TD TMP pressure low ALARM_ID_TD_TMP_PRESSURE_HIGH = 29, ///< TD TMP pressure high - ALARM_ID_AVAILABLE_30 = 30, ///< Available alarms + ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 30, ///< TD venous bubble self-test failure. ALARM_ID_AVAILABLE_31 = 31, ///< Available alarms ALARM_ID_AVAILABLE_32 = 32, ///< Available alarms ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE = 33, ///< TD voltage out of range fault @@ -362,9 +362,9 @@ // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 204, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, ALARM_ID_TD_TMP_PRESSURE_LOW }, - { ALARM_PRIORITY_HIGH, 205, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, ALARM_ID_TD_TMP_PRESSURE_HIGH }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_30 }, + { ALARM_PRIORITY_HIGH, 204, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_TMP_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 205, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_TMP_PRESSURE_HIGH }, + { ALARM_PRIORITY_LOW, 711, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_31 }, { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_32 }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, @@ -655,9 +655,9 @@ { 700, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, { 700, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, + { 711, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { 850, ALARM_ID_BATTERY_LOW }, { 852, ALARM_ID_TD_AC_POWER_LOST }, - { 999, ALARM_ID_AVAILABLE_30 }, { 999, ALARM_ID_AVAILABLE_31 }, { 999, ALARM_ID_AVAILABLE_32 }, { 999, ALARM_ID_UI_RESERVED_116 }, @@ -729,7 +729,7 @@ { 27 , ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Venous Pressure Sensor Error | TD venous pressure sensor reporting error or not being read successfully | Venous pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 28 , ALARM_ID_TD_TMP_PRESSURE_LOW , "TD | Low TMP Pressure | TMP pressure is low | | | | "}, { 29 , ALARM_ID_TD_TMP_PRESSURE_HIGH , "TD | High TMP Pressure | TMP pressure is high | | | | "}, - { 30 , ALARM_ID_AVAILABLE_30 , " | | | | | | "}, + { 30 , ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE , "TD | Self Test: Venous Air Detector | A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. | Process: Pre- Tx Wet Self- Test | TD venous air bubble detector self-test failure | If the venous air bubble self test did not detect bubble for more than 500 msec | "}, { 31 , ALARM_ID_AVAILABLE_31 , " | | | | | | "}, { 32 , ALARM_ID_AVAILABLE_32 , " | | | | | | "}, { 33 , ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Voltage Monitor | TD monitored voltage is out of range | If the TD monitored voltages are out of range for a certain period of time. | "},