Index: AlarmDefs.h
===================================================================
diff -u -r86c541dc9f6ed51cde9ca9f1697989d58eff7c31 -r09805a6b655f1268ae1677ad317b04fc3217a901
--- AlarmDefs.h (.../AlarmDefs.h) (revision 86c541dc9f6ed51cde9ca9f1697989d58eff7c31)
+++ AlarmDefs.h (.../AlarmDefs.h) (revision 09805a6b655f1268ae1677ad317b04fc3217a901)
@@ -1042,7 +1042,7 @@
{ 198 , ALARM_ID_TD_PRIME_COMPLETED_MEDIUM , "TD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre-treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre-Tx Prime Complete | TD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes | "},
{ 199 , ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS , "TD | Blood Pump Stopped: Venous Air Detector | The air trap level required repeated lowering events. \n- Confirm blood tubing connections are secure.\n- Inspect the air trap for abnormal fluid levels. \n- Resume treatment if the condition has been corrected. | Blood Stop: Repeated Air Trap Lower | Repeated air trap lower events detected during treatment | If 5 air trap lower events occur within 5 minutes during treatment. | "},
{ 200 , ALARM_ID_TD_TUBE_SET_MISMATCH , "TD | Wrong Tube Set Installed | The installed tube set does not match the selected treatment type. \n- Remove the current tube set.\n- Install the correct tube set for the selected treatment:\n HD or HD with Online Fluid requires an HD tube set.\n HDF requires an HDF tube set. | TD Tube Set Mismatch | TD installed tube set type does not match the selected treatment modality | If the scanned tube set type does not match the tube set type required for the selected treatment modality. | "},
- { 201 , ALARM_ID_TD_PRE_TX_HEPARIN_LOAD_SYRINGE , "UI TD | Load Syringe | Fill syringe with 4mL of heparin.\n- Connect heparin filled syringe to infusion port.\n- Insert syringe into the syringe pump.\n- Lock heparin syringe in place.\n- Unclamp heparin pinch clamp. | Process: Pre-Tx Load Syringe | User must load the heparin syringe before continuing heparin setup. | While in Pre-Treatment: Heparin Setup: Await Syringe Load Confirmation state. | "},
+ { 201 , ALARM_ID_TD_PRE_TX_HEPARIN_LOAD_SYRINGE , "UI TD | Load Syringe | Fill Syringe with {140:2:mL:1} of heparin.\n- Connect heparin filled syringe to infusion port.\n- Insert syringe into the syringe pump.\n- Lock heparin syringe in place.\n- Unclamp heparin pinch clamp. | Process: Pre-Tx Load Syringe | User must load the heparin syringe before continuing heparin setup. | While in Pre-Treatment: Heparin Setup: Await Syringe Load Confirmation state. | "},
};
#endif // manual alarms