Index: AlarmDefs.h =================================================================== diff -u -r45907574a0110204746643f627960946ee7e06b6 -r202074aac58e33ea1be4d91b9600e1ef6670b6e4 --- AlarmDefs.h (.../AlarmDefs.h) (revision 45907574a0110204746643f627960946ee7e06b6) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 202074aac58e33ea1be4d91b9600e1ef6670b6e4) @@ -229,6 +229,12 @@ ALARM_ID_DD_D92_PUMP_RPM_OUT_OF_RANGE = 197, ///< DD D92 pump rpm out of range ALARM_ID_TD_PRIME_COMPLETED_MEDIUM = 198, ///< TD prime completed medium priority alarm ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS = 199, ///< TD repeated air trap lower events + ALARM_ID_DD_DIALYSATE_CONDUCTIVITY_LOW_DETECTED = 200, ///< DD dialysate solution conductivity low + ALARM_ID_DD_DIALYSATE_CONDUCTIVITY_HIGH_DETECTED = 201, ///< DD dialysate solution conductivity high + ALARM_ID_DD_ACID_CONCENTRATE_CONDUCTIVITY_LOW = 202, ///< DD acid concentrate conductivity low + ALARM_ID_DD_ACID_CONCENTRATE_CONDUCTIVITY_HIGH = 203, ///< DD acid concentrate conductivity high + ALARM_ID_DD_BICARBONATE_CONDUCTIVITY_LOW = 204, ///< DD bicarbonate concentrate conductivity low + ALARM_ID_DD_BICARBONATE_CONDUCTIVITY_HIGH = 205, ///< DD bicarbonate concentrate conductivity high NUM_OF_ALARM_IDS ///< Number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -576,6 +582,12 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D92_PUMP_RPM_OUT_OF_RANGE }, { ALARM_PRIORITY_MEDIUM, 590, ALM_SRC_TD, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_TD_PRIME_COMPLETED_MEDIUM }, { ALARM_PRIORITY_HIGH, 17, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS }, + { ALARM_PRIORITY_MEDIUM, 403, ALM_SRC_DD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_DIALYSATE_CONDUCTIVITY_LOW_DETECTED }, + { ALARM_PRIORITY_MEDIUM, 403, ALM_SRC_DD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_DIALYSATE_CONDUCTIVITY_HIGH_DETECTED }, + { ALARM_PRIORITY_MEDIUM, 404, ALM_SRC_DD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_ACID_CONCENTRATE_CONDUCTIVITY_LOW }, + { ALARM_PRIORITY_MEDIUM, 404, ALM_SRC_DD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_ACID_CONCENTRATE_CONDUCTIVITY_HIGH }, + { ALARM_PRIORITY_MEDIUM, 405, ALM_SRC_DD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_BICARBONATE_CONDUCTIVITY_LOW }, + { ALARM_PRIORITY_MEDIUM, 405, ALM_SRC_DD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , ALARM_ID_DD_BICARBONATE_CONDUCTIVITY_HIGH }, }; // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID /// Table of alarms and their rank. @@ -745,6 +757,12 @@ { 401, ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_SAFETY_TEMP }, { 401, ALARM_ID_DD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP }, { 402, ALARM_ID_DD_DIALYSATE_TEMP_BELOW_TARGET_TEMP }, + { 403, ALARM_ID_DD_DIALYSATE_CONDUCTIVITY_LOW_DETECTED }, + { 403, ALARM_ID_DD_DIALYSATE_CONDUCTIVITY_HIGH_DETECTED }, + { 404, ALARM_ID_DD_ACID_CONCENTRATE_CONDUCTIVITY_LOW }, + { 404, ALARM_ID_DD_ACID_CONCENTRATE_CONDUCTIVITY_HIGH }, + { 405, ALARM_ID_DD_BICARBONATE_CONDUCTIVITY_LOW }, + { 405, ALARM_ID_DD_BICARBONATE_CONDUCTIVITY_HIGH }, { 432, ALARM_ID_FP_INLET_TEMPERATURE_OUT_LOW_RANGE }, { 433, ALARM_ID_FP_INLET_TEMPERATURE_OUT_HIGH_RANGE }, { 434, ALARM_ID_FP_INLET_PRESSURE_OUT_LOW_RANGE }, @@ -1010,6 +1028,12 @@ { 197 , ALARM_ID_DD_D92_PUMP_RPM_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Subtitution Pump Fault | DD substitution pump RPM out of range | If the substitution pump RPM > 20.0 of the target RPM. | "}, { 198 , ALARM_ID_TD_PRIME_COMPLETED_MEDIUM , "TD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre-treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre-Tx Prime Complete | TD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes | "}, { 199 , ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS , "TD | Blood Pump Stopped: Venous Air Detector | The air trap level required repeated lowering events. \n- Confirm blood tubing connections are secure.\n- Inspect the air trap for abnormal fluid levels. \n- Resume treatment if the condition has been corrected. | Blood Stop: Repeated Air Trap Lower | Repeated air trap lower events detected during treatment | If 5 air trap lower events occur within 5 minutes during treatment. | "}, + { 200 , ALARM_ID_DD_DIALYSATE_CONDUCTIVITY_LOW_DETECTED , "DD | Low Dialysate Conductivity | Dialysate conductivity is low. \n- Treatment is currently paused.\n- When dialysate conductivity is within target range,\n treatment will resume automatically. | Dial. Gen: Dialysate Cond Low | Dialysate solution conductivity is below the theoretical dialysate conductivity | If measured dialysate conductivity is below theoretical dialysate conductivity minus 5% for a certain period of time. | "}, + { 201 , ALARM_ID_DD_DIALYSATE_CONDUCTIVITY_HIGH_DETECTED , "DD | High Dialysate Conductivity | Dialysate conductivity is high. \n- Treatment is currently paused.\n- When dialysate conductivity is within target range,\n treatment will resume automatically. | Dial. Gen: Dialysate Cond High | Dialysate solution conductivity is above the theoretical dialysate conductivity | If measured dialysate conductivity is above theoretical dialysate conductivity plus 5% for a certain period of time. | "}, + { 202 , ALARM_ID_DD_ACID_CONCENTRATE_CONDUCTIVITY_LOW , "DD | Low Acid Concentrate Conductivity | Acid concentrate conductivity is low. \n- Treatment is currently paused.\n- When acid concentrate conductivity is within target range,\n treatment will resume automatically. | Dial. Gen: Acid Cond Low | Acid concentrate conductivity is below the theoretical acid concentrate conductivity | If measured acid concentrate conductivity is below theoretical acid concentrate conductivity minus 5% for a certain period of time. | "}, + { 203 , ALARM_ID_DD_ACID_CONCENTRATE_CONDUCTIVITY_HIGH , "DD | High Acid Concentrate Conductivity | Acid concentrate conductivity is high. \n- Treatment is currently paused.\n- When acid concentrate conductivity is within target range,\n treatment will resume automatically. | Dial. Gen: Acid Cond High | Acid concentrate conductivity is above the theoretical acid concentrate conductivity | If measured acid concentrate conductivity is above theoretical acid concentrate conductivity plus 5% for a certain period of time. | "}, + { 204 , ALARM_ID_DD_BICARBONATE_CONDUCTIVITY_LOW , "DD | Low Bicarbonate Concentrate Conductivity| Bicarbonate concentrate conductivity is low. \n- Treatment is currently paused.\n- When bicarbonate concentrate conductivity is within target range,\n treatment will resume automatically. | Dial. Gen: Bicarb Cond Low | Bicarbonate concentrate conductivity is below the theoretical bicarbonate concentrate conductivity | If measured bicarbonate concentrate conductivity is below theoretical bicarbonate concentrate conductivity minus 5% for a certain period of time. | "}, + { 205 , ALARM_ID_DD_BICARBONATE_CONDUCTIVITY_HIGH , "DD | High Bicarbonate Concentrate Conductivity| Bicarbonate concentrate conductivity is high. \n- Treatment is currently paused.\n- When bicarbonate concentrate conductivity is within target range,\n treatment will resume automatically. | Dial. Gen: Bicarb Cond High | Bicarbonate concentrate conductivity is above the theoretical bicarbonate concentrate conductivity | If measured bicarbonate concentrate conductivity is above theoretical bicarbonate concentrate conductivity plus 5% for a certain period of time. | "}, }; #endif // manual alarms