Index: AlarmDefs.h =================================================================== diff -u -rea9c4de9a38bfee6644fd22c3a8d629ea476cfdf -r609f848305356e5b9735f6d56fa451a35a14982d --- AlarmDefs.h (.../AlarmDefs.h) (revision ea9c4de9a38bfee6644fd22c3a8d629ea476cfdf) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 609f848305356e5b9735f6d56fa451a35a14982d) @@ -1005,7 +1005,7 @@ { 195 , ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED , "DD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Firmware Image Integrity Failure | DD firmware image integrity POST test failed | If the calculated CRC of the DD image does not match the expected CRC. | "}, { 196 , ALARM_ID_TD_EMPTY_SALINE_BAG , "TD | Blood Pump Stopped: Saline Bag Empty | The saline bag appears empty.\n- Check the saline bag and replace if empty. | Blood Stop: Saline Bag Empty | TD Saline bag is empty | If the measured arterial pressure < -300.0 mmHg for more than 250 msec. | "}, { 197 , ALARM_ID_DD_D92_PUMP_RPM_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Subtitution Pump Fault | DD substitution pump RPM out of range | If the substitution pump RPM > 20.0 of the target RPM. | "}, - { 198, ALARM_ID_TD_PRIME_COMPLETED_MEDIUM , "TD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre-treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre-Tx Prime Complete | TD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes (escalation to alarm 197) | "}, + { 198, ALARM_ID_TD_PRIME_COMPLETED_MEDIUM , "TD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre-treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre-Tx Prime Complete | TD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes | "}, }; #endif // manual alarms