Index: AlarmDefs.h =================================================================== diff -u -r8e342610940555081af481b1b7e41b7aa09010d9 -r895fad7be90248f1287eac6648093c21c4a47927 --- AlarmDefs.h (.../AlarmDefs.h) (revision 8e342610940555081af481b1b7e41b7aa09010d9) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 895fad7be90248f1287eac6648093c21c4a47927) @@ -184,6 +184,11 @@ ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE = 152, ///< FP membrane flush integrated out of range ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT = 153, ///< FP tank full time out alarm ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE = 154, ///< DD dialysate temperature sensors out of range + ALARM_ID_DD_BLOOD_LEAK_DETECTED = 155, ///< DD blood leak detected + ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD = 156, ///< DD blood leak invalid calibration record + ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE = 157, ///< DD blood leak sensor embedded mode failure + ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE = 158, ///< DD blood leak sensor set point failure + ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT = 159, ///< DD blood leak sensor recovering please wait NUM_OF_ALARM_IDS ///< Number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -482,6 +487,11 @@ { ALARM_PRIORITY_HIGH, 110, TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, TRUE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, + { ALARM_PRIORITY_HIGH, 150, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , ALARM_ID_DD_BLOOD_LEAK_DETECTED }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, + { ALARM_PRIORITY_HIGH, 1, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, + { ALARM_PRIORITY_MEDIUM, 698, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT, }, }; // Priority Rank Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID /// Table of alarms and their rank. @@ -508,6 +518,9 @@ { 1, ALARM_ID_TD_UI_COMM_POST_FAILED }, { 1, ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE }, { 1, ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED }, + { 1, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, + { 1, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, + { 1, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, { 10, ALARM_ID_TD_SOFTWARE_FAULT }, { 10, ALARM_ID_TD_UI_COMM_TIMEOUT }, { 10, ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS }, @@ -599,6 +612,7 @@ { 111, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_FP }, { 111, ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_TD }, { 111, ALARM_ID_FP_CAN_MESSAGE_NOT_ACKED_BY_DD }, + { 150, ALARM_ID_DD_BLOOD_LEAK_DETECTED }, { 200, ALARM_ID_TD_VENOUS_PRESSURE_LOW }, { 201, ALARM_ID_TD_VENOUS_PRESSURE_HIGH }, { 202, ALARM_ID_TD_ARTERIAL_PRESSURE_LOW }, @@ -615,6 +629,7 @@ { 435, ALARM_ID_FP_INLET_PRESSURE_OUT_HIGH_RANGE }, { 436, ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_HIGH_RANGE }, { 437, ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_LOW_RANGE }, + { 698, ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT }, { 700, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, { 700, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, @@ -825,6 +840,11 @@ { 152 , ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Volume difference too high | FP volume difference between integrated flow and expected volume | If P7 is <90% of the target integrated volume ( 1L ) after 30 seconds ± 150ms of flushing. | "}, { 153 , ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT , "FP | FP: Permeate Overfill | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Tank: Tank is too full | FP tank level is too high | If the float sensor reads high for at least 60 seconds | "}, { 154 , ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: D28 and D30 Delta | DD dialysate temperature sensors drift timeout. | If D28 and D30 have drifted for more 1 C for a certain period of time. | "}, + { 155 , ALARM_ID_DD_BLOOD_LEAK_DETECTED , "DD | Blood Leak | A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. | Blood Stop: Dialyzer Blood Leak | DD blood leak detected alarm | If the blood leak is detected based on the blood leak status reported by the sensor. | "}, + { 156 , ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD POST: BLD Cal | DD blood leak sensor invalid calibration record | If the BLD data that was received from NVDataMgmt is NOT legitimate. | "}, + { 157 , ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD POST: BLD Config Emb Mode | DD blood leak sensor embedded mode communication failed | If the blood leak sensor embedded mode is unresponsive. | "}, + { 158 , ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE , "DD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD POST: BLD Config Set Point | DD blood leak sensor set point set failure | If the set point of the blood leak sensor does not match the set point in the calibration record. | "}, + { 159 , ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT , "DD | Blood Leak: Bypass | The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. | Blood Stop: Dialyzer Blood Leak Bypass | DD blood leak detected and it is in the process of attempting to remove the bubbles/foam | If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. | "}, }; #endif // manual alarms