Index: AlarmDefs.h =================================================================== diff -u -r622c6745b1adaeff233e57a3779757b12d60b08c -rafa6b0daa2cb0321900975045a9fde181715c569 --- AlarmDefs.h (.../AlarmDefs.h) (revision 622c6745b1adaeff233e57a3779757b12d60b08c) +++ AlarmDefs.h (.../AlarmDefs.h) (revision afa6b0daa2cb0321900975045a9fde181715c569) @@ -121,7 +121,7 @@ ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE = 90, ///< DD balance chamber state2 fill end pressure not in range ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE = 91, ///< DD generate dialysate hydraulics chamber ngative pressure out of range ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT = 92, ///< DD spent chamber fill timeout fault - ALARM_ID_AVAILABLE_93 = 93, ///< Available alarms + ALARM_ID_DD_D79_RINSE_PUMP_SPEED_OUT_OF_RANGE = 93, ///< DD D79 rinse pump commanded vs measured speed out of tolerance ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE = 94, ///< DD balance chamber state1 fill initiated and pressure drop is not in range ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE = 95, ///< DD balance chamber state2 fill initiated and pressure drop is not in range ALARM_ID_TD_TREATMENT_STOPPED_BY_USER = 96, ///< TD treatment paused by user @@ -428,7 +428,7 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_93 }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D79_RINSE_PUMP_SPEED_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { ALARM_PRIORITY_MEDIUM, 300, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_TREATMENT_STOPPED_BY_USER }, @@ -656,7 +656,7 @@ { 999, ALARM_ID_AVAILABLE_31 }, { 999, ALARM_ID_AVAILABLE_32 }, { 999, ALARM_ID_AVAILABLE_58 }, - { 999, ALARM_ID_AVAILABLE_93 }, + { 110, ALARM_ID_DD_D79_RINSE_PUMP_SPEED_OUT_OF_RANGE }, { 999, ALARM_ID_UI_RESERVED_116 }, { 999, ALARM_ID_UI_RESERVED_117 }, { 999, ALARM_ID_UI_RESERVED_118 }, @@ -791,7 +791,7 @@ { 90 , ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill end pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill end. | "}, { 91 , ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Hyd Cham Neg Pres not in range | DD hydraulics chamber degassing pressure is not in range | If the hydraulics chamber pressure is not less than or equal to calculated target degas pressure psig. | "} { 92 , ALARM_ID_DD_BICARB_FILL_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Bicarb Chamber fill timeout Error | DD Bicarb chamber fill timeout error | If Bicarb chamber fill time > = 1 seconds | "}, - { 93 , ALARM_ID_AVAILABLE_93 , " | | | | | | "}, + { 93 , ALARM_ID_DD_D79_RINSE_PUMP_SPEED_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Rinse Pump Speed Fault | DD D79 rinse pump commanded vs measured speed out of tolerance | If rinse pump measured RPM deviates beyond 10%% of target RPM when pump is commanded on. Data1=measured RPM, Data2=target RPM. | "}, { 94 , ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state1 fill in progress. | "}, { 95 , ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state2 fill in progress. | "}, { 96 , ALARM_ID_TD_TREATMENT_STOPPED_BY_USER , "TD | Blood Pump Stopped: Treatment Paused | Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. | Blood Stopped: Treatment Pause | TD treatment stopped by user action - pressed stop button | If the user has pressed the stop button. | "},