Index: AlarmDefs.h =================================================================== diff -u -r2d9b194c68326df3f80f60a7b2c154409cd9597f -rbeacec5020e8d073014576928b5a7d4283bd183c --- AlarmDefs.h (.../AlarmDefs.h) (revision 2d9b194c68326df3f80f60a7b2c154409cd9597f) +++ AlarmDefs.h (.../AlarmDefs.h) (revision beacec5020e8d073014576928b5a7d4283bd183c) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) Nico Ramirez -* @date (last) 17-Apr-2026 +* @author (last) Varshini Nagabooshanam +* @date (last) 07-May-2026 * * @author (original) Sean Nash * @date (original) 01-Aug-2024 @@ -216,10 +216,23 @@ ALARM_ID_TD_SYRINGE_PUMP_PRIME_TIMEOUT = 184, ///< TD syringe pump prime timeout ALARM_ID_FP_P9_COND_SENSOR_FPGA_FAULT = 185, ///< FP P9 conductivity sensor fpga fault ALARM_ID_FP_P18_COND_SENSOR_FPGA_FAULT = 186, ///< FP P18 conductivity sensor fpga fault - ALARM_ID_TD_CPU_RAM_ERROR = 187, ///< TD CPU RAM error - ALARM_ID_DD_CPU_RAM_ERROR = 188, ///< DD CPU RAM error - ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED = 189, ///< TD integrity post test fault - ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED = 190, ///< TD integrity post test fault + ALARM_ID_TD_INVALID_TUBE_SET = 187, ///< TD invalid tube set + ALARM_ID_TD_INSTALL_NEW_BLOOD_TUBING_SET = 188, ///< TD install new blood tubing set + ALARM_ID_TD_TUBE_SET_NOT_DETECTED = 189, ///< TD no tubing set detected + ALARM_ID_TD_TUBE_SET_AUTHENTICATION_TIMEOUT = 190, ///< TD tubing set authentication timeout + ALARM_ID_TD_PRESSURE_TOO_HIGH_TO_ZERO = 191, ///< TD pressure too high to zero + ALARM_ID_TD_VENOUS_PRESSURIZATION_TIMEOUT = 192, ///< TD venous pressurization timeout + ALARM_ID_TD_VENOUS_PRESSURE_DECAY = 193, ///< TD venous pressure decay failure + ALARM_ID_TD_VENOUS_PRESSURE_LEAK = 194, ///< TD venous pressure lea + ALARM_ID_TD_VENOUS_PRESSURE_RELIEF = 195, ///< TD venous pressure relief failure + ALARM_ID_TD_ARTERIAL_PRESSURIZATION_TIMEOUT = 196, ///< TD arterial pressurization timeout + ALARM_ID_TD_ARTERIAL_PRESSURE_DECAY = 197, ///< TD arterial pressure decay failure + ALARM_ID_TD_ARTERIAL_PRESSURE_LEAK = 198, ///< TD arterial pressure leak + ALARM_ID_TD_ARTERIAL_PRESSURE_RELIEF = 199, ///< TD arterial pressure relief failure + ALARM_ID_TD_CPU_RAM_ERROR = 200, ///< TD CPU RAM error + ALARM_ID_DD_CPU_RAM_ERROR = 201, ///< DD CPU RAM error + ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED = 202, ///< TD integrity post test fault + ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED = 203, ///< TD integrity post test fault NUM_OF_ALARM_IDS ///< Number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -385,6 +398,7 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_FPGA_COMM_TIMEOUT }, { ALARM_PRIORITY_HIGH, 1, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_VALVE_CONTROL_FAILURE }, + // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, @@ -411,6 +425,19 @@ { ALARM_PRIORITY_HIGH, 201, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_VENOUS_PRESSURE_HIGH }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_LOW, 700, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, + { ALARM_PRIORITY_LOW, 717, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_INVALID_TUBE_SET }, + { ALARM_PRIORITY_LOW, 718, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_INSTALL_NEW_BLOOD_TUBING_SET }, + { ALARM_PRIORITY_LOW, 719, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_TUBE_SET_NOT_DETECTED }, + { ALARM_PRIORITY_LOW, 720, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_TUBE_SET_AUTHENTICATION_TIMEOUT }, + { ALARM_PRIORITY_LOW, 721, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_PRESSURE_TOO_HIGH_TO_ZERO }, + { ALARM_PRIORITY_LOW, 722, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURIZATION_TIMEOUT }, + { ALARM_PRIORITY_LOW, 723, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_DECAY }, + { ALARM_PRIORITY_LOW, 724, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_LEAK }, + { ALARM_PRIORITY_LOW, 725, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_PRESSURE_RELIEF }, + { ALARM_PRIORITY_LOW, 726, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURIZATION_TIMEOUT }, + { ALARM_PRIORITY_LOW, 727, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_DECAY }, + { ALARM_PRIORITY_LOW, 728, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_LEAK }, + { ALARM_PRIORITY_LOW, 729, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_PRESSURE_RELIEF }, // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID { ALARM_PRIORITY_LOW, 710, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR }, @@ -521,9 +548,9 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_FP, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FULL_TIMEOUT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 150, ALM_SRC_DD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , ALARM_ID_DD_BLOOD_LEAK_DETECTED }, - { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, - { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, - { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, + { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_INVALID_CAL_RECORD }, + { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE }, + { ALARM_PRIORITY_HIGH, 2, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE }, { ALARM_PRIORITY_MEDIUM, 698, ALM_SRC_DD, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT, }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_FP, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_GEN_PERMEATE_TANK_FILL_TIMEOUT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_FP, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_FP_INLET_PRESSURE_CRITICAL_OUT_RANGE }, @@ -733,6 +760,19 @@ { 711, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { 715, ALARM_ID_TD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM }, { 716, ALARM_ID_TD_SYRINGE_PUMP_OCCLUSION }, + { 717, ALARM_ID_TD_INVALID_TUBE_SET }, + { 718, ALARM_ID_TD_INSTALL_NEW_BLOOD_TUBING_SET }, + { 719, ALARM_ID_TD_TUBE_SET_NOT_DETECTED }, + { 720, ALARM_ID_TD_TUBE_SET_AUTHENTICATION_TIMEOUT }, + { 721, ALARM_ID_TD_PRESSURE_TOO_HIGH_TO_ZERO }, + { 722, ALARM_ID_TD_VENOUS_PRESSURIZATION_TIMEOUT }, + { 723, ALARM_ID_TD_VENOUS_PRESSURE_DECAY }, + { 724, ALARM_ID_TD_VENOUS_PRESSURE_LEAK }, + { 725, ALARM_ID_TD_VENOUS_PRESSURE_RELIEF }, + { 726, ALARM_ID_TD_ARTERIAL_PRESSURIZATION_TIMEOUT }, + { 727, ALARM_ID_TD_ARTERIAL_PRESSURE_DECAY }, + { 728, ALARM_ID_TD_ARTERIAL_PRESSURE_LEAK }, + { 729, ALARM_ID_TD_ARTERIAL_PRESSURE_RELIEF }, { 730, ALARM_ID_TD_SYRINGE_PUMP_SYRINGE_EMPTY }, { 745, ALARM_ID_TD_SYRINGE_DETECTED }, { 850, ALARM_ID_BATTERY_LOW }, @@ -970,10 +1010,23 @@ { 185 , ALARM_ID_FP_P9_COND_SENSOR_FPGA_FAULT , "FP | Service Required: FP Device | A problem was detected with the FP device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Conductivity Sensor Error | FP conductivity sensor (P9) located in inlet water path reporting error or not being read successfully | Conductvity Sensor (P9) is reporting an error or FPGA is failing to read sensor. | "}, { 186 , ALARM_ID_FP_P18_COND_SENSOR_FPGA_FAULT , "FP | Service Required: FP Device | A problem was detected with the FP device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Conductivity Sensor Error | FP conductivity sensor (P18) located in inlet water path reporting error or not being read successfully | Conductvity Sensor (P18) is reporting an error or FPGA is failing to read sensor. | "}, { 187 , ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT , "TD | Blood Pump Stopped: Venous Air Detector | A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Air Trap Lowering | Air trap lowering timeout during treatment | If the air trap lowering timeout during treatment. | "}, - { 188 , ALARM_ID_TD_CPU_RAM_ERROR , "TD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD CPU RAM error | TD CPU RAM check failure | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, - { 189 , ALARM_ID_DD_CPU_RAM_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD CPU RAM error | DD CPU RAM check failure | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, - { 190 , ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Firmware Image Integrity Failure | TD firmware image integrity POST test failed | If the calculated CRC of the TD image does not match the expected CRC. | "}, - { 191 , ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED , "DD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Firmware Image Integrity Failure | DD firmware image integrity POST test failed | If the calculated CRC of the DD image does not match the expected CRC. | "}, + { 187 , ALARM_ID_TD_INVALID_TUBE_SET , "TD | Invalid Blood Tubing Set | A problem has been detected during self-test.\n- The installed blood tubing set is not valid.\n- Install a valid blood tubing set intended for the selected treatment modality. | Process: Pre-Tx Dry Self-Test | TD detected invalid tubing set during authentication | If authentication response indicates the tubing set is invalid. | "}, + { 188 , ALARM_ID_TD_INSTALL_NEW_BLOOD_TUBING_SET , "TD | Install New Blood Tubing Set | A problem has been detected during self-test.\n- Fluid detected in the tubing set or no valid tubing set installed.\n- Install a new, unused blood tubing set and ensure proper installation. | Process: Pre-Tx Dry Self-Test | TD detected used, wet, or missing blood tubing set | If bubble is detected, air trap levels indicate fluid, or barcode scan fails. | "}, + { 189 , ALARM_ID_TD_TUBE_SET_NOT_DETECTED , "TD | No Blood Tubing Set Detected | A problem has been detected during self-test.\n- No blood tubing set detected in the system.\n- Install a blood tubing set and ensure proper placement. | Process: Pre-Tx Dry Self-Test | TD did not detect tubing set during verification | If no tubing set is detected during tube set loaded check. | "}, + { 190 , ALARM_ID_TD_TUBE_SET_AUTHENTICATION_TIMEOUT , "TD | Tubing Set Authentication Timeout | A problem has been detected during self-test.\n- Authentication response not received.\n- Retry or check system communication. | Process: Pre-Tx Dry Self-Test | TD did not receive authentication response from UI | If no authentication response is received within timeout period. | "}, + { 191 , ALARM_ID_TD_PRESSURE_TOO_HIGH_TO_ZERO , "TD | Pressure Too High to Zero | A problem has been detected during self-test.\n- Pressure levels are too high to perform sensor zeroing.\n- Ensure system is open to atmosphere and retry. | Process: Pre-Tx Dry Self-Test | TD detected high pressure during zeroing | If arterial or venous pressure ≥ 15 mmHg during pressure sensor normal setup. | "}, + { 192 , ALARM_ID_TD_VENOUS_PRESSURIZATION_TIMEOUT , "TD | Venous Pressurization Timeout | A problem has been detected during self-test.\n- Venous pressure did not reach required level.\n- Check tubing connections and system integrity. | Process: Pre-Tx Dry Self-Test | TD failed to reach venous pressure target | If venous pressure does not reach threshold within allowed time. | "}, + { 193 , ALARM_ID_TD_VENOUS_PRESSURE_DECAY , "TD | Venous Pressure Decay | A problem has been detected during self-test.\n- Venous pressure dropped more than expected.\n- Check for leaks or loose connections. | Process: Pre-Tx Dry Self-Test | TD detected excessive venous pressure decay | If venous pressure reduction exceeds allowed tolerance during stabilization. | "}, + { 194 , ALARM_ID_TD_VENOUS_PRESSURE_LEAK , "TD | Venous Pressure Leak | A problem has been detected during self-test.\n- Possible leak in venous line.\n- Inspect tubing and connections. | Process: Pre-Tx Dry Self-Test | TD detected venous pressure leak | If venous pressure difference exceeds tolerance during leak check. | "}, + { 195 , ALARM_ID_TD_VENOUS_PRESSURE_RELIEF , "TD | Venous Pressure Relief Failure | A problem has been detected during self-test.\n- Pressure did not return to baseline.\n- Ensure valves and tubing are functioning properly. | Process: Pre-Tx Dry Self-Test | TD failed venous pressure relief check | If venous pressure does not return within ±5 mmHg after relief. | "}, + { 196 , ALARM_ID_TD_ARTERIAL_PRESSURIZATION_TIMEOUT , "TD | Arterial Pressurization Timeout | A problem has been detected during self-test.\n- Arterial pressure did not reach required level.\n- Check tubing and pump operation. | Process: Pre-Tx Dry Self-Test | TD failed to reach arterial pressure target | If arterial pressure does not reach threshold within allowed time. | "}, + { 197 , ALARM_ID_TD_ARTERIAL_PRESSURE_DECAY , "TD | Arterial Pressure Decay | A problem has been detected during self-test.\n- Arterial pressure dropped more than expected.\n- Check for leaks or loose connections. | Process: Pre-Tx Dry Self-Test | TD detected excessive arterial pressure decay | If arterial pressure reduction exceeds allowed tolerance during stabilization. | "}, + { 198 , ALARM_ID_TD_ARTERIAL_PRESSURE_LEAK , "TD | Arterial Pressure Leak | A problem has been detected during self-test.\n- Possible leak in arterial line.\n- Inspect tubing and connections. | Process: Pre-Tx Dry Self-Test | TD detected arterial pressure leak | If arterial pressure difference exceeds tolerance during leak check. | "}, + { 199 , ALARM_ID_TD_ARTERIAL_PRESSURE_RELIEF , "TD | Arterial Pressure Relief Failure | A problem has been detected during self-test.\n- Pressure did not return to baseline.\n- Ensure valves and tubing are functioning properly. | Process: Pre-Tx Dry Self-Test | TD failed arterial pressure relief check | If arterial pressure does not return within ±5 mmHg after relief. | "}, + { 200 , ALARM_ID_TD_CPU_RAM_ERROR , "TD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD CPU RAM error | TD CPU RAM check failure | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, + { 201 , ALARM_ID_DD_CPU_RAM_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD CPU RAM error | DD CPU RAM check failure | If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. | "}, + { 202 , ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Firmware Image Integrity Failure | TD firmware image integrity POST test failed | If the calculated CRC of the TD image does not match the expected CRC. | "}, + { 203 , ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED , "DD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Firmware Image Integrity Failure | DD firmware image integrity POST test failed | If the calculated CRC of the DD image does not match the expected CRC. | "}, }; #endif // manual alarms Index: MsgDefs.h =================================================================== diff -u -r73bd843c681e79ccdc34c967a9108eb3eadabab6 -rbeacec5020e8d073014576928b5a7d4283bd183c --- MsgDefs.h (.../MsgDefs.h) (revision 73bd843c681e79ccdc34c967a9108eb3eadabab6) +++ MsgDefs.h (.../MsgDefs.h) (revision beacec5020e8d073014576928b5a7d4283bd183c) @@ -8,8 +8,8 @@ * * @file MsgDefs.h * -* @author (last) Sameer Kalliadan Poyil -* @date (last) 29-Apr-2026 +* @author (last) Arpita Srivastava +* @date (last) 18-May-2026 * * @author (original) Sean Nash * @date (original) 01-Aug-2024 @@ -195,8 +195,21 @@ MSG_ID_UI_WATER_SAMPLE_RESULT_REQUEST = 0xA3, ///< UI water sample result request MSG_ID_UI_PRESSURE_LIMIT_WIDEN_REQUEST = 0xA4, ///< UI pressure limit widen request. MSG_ID_TD_PRESSURE_LIMIT_WIDEN_RESPONSE = 0xA5, ///< TD pressure limit widen response. - MSG_ID_FFU_SIGNAL_TD_UPDATE_AVAILABLE = 0xA6, ///< FFU signal TD update is available. - MSG_ID_FFU_SIGNAL_DD_UPDATE_AVAILABLE = 0xA7, ///< FFU signal DD update is available. + MSG_ID_UI_RECIRCULATE_REQUEST = 0xA6, ///< UI recirculate request + MSG_ID_TD_RECIRCULATE_RESPONSE = 0xA7, ///< TD recirculate response + MSG_ID_TD_RECIRCULATE_DATA = 0xA8, ///< TD recirculate data + MSG_ID_UI_ADJUST_TREATMENT_LOGS_REQUEST = 0xA9, ///< UI adjust treatment logs request. + MSG_ID_TD_ADJUST_TREATMENT_LOGS_RESPONSE = 0xAA, ///< TD adjust treatment logs response + MSG_ID_UI_ADJUST_PATIENT_DISCONNECT_CONFIRM_REQUEST = 0xAB, ///< UI adjust patient disconnect confirm request + MSG_ID_TD_ADJUST_PATIENT_DISCONNECT_CONFIRM_RESPONSE = 0xAC, ///< TD adjust patient disconnect confirm response + MSG_ID_TD_DRY_SELF_TEST_PROGRESS_DATA = 0xAD, ///< TD dry self test progress data + MSG_ID_TD_TUBING_SET_BARCODE_SCAN_REQUEST = 0xAE, ///< TD tube set authentication request + MSG_ID_TD_TUBE_SET_AUTHENTICATION_REQUEST = 0xAF, ///< UI tube set authentication acknowledge request + MSG_ID_TD_TUBE_SET_AUTHENTICATION_ACK_RESPONSE = 0xB0, ///< TD tube set authentication acknowledge response + MSG_ID_TD_SYRINGE_PUMP_DATA = 0xB1, ///< TD broadcast of syringe pump data + MSG_ID_TD_HEPARIN_PAUSE_RESUME_RESPONSE= 0xB2, ///< TD response to pause/resume Heparin delivery request + MSG_ID_FFU_SIGNAL_TD_UPDATE_AVAILABLE = 0xB3, ///< FFU signal TD update is available. + MSG_ID_FFU_SIGNAL_DD_UPDATE_AVAILABLE = 0xB4, ///< FFU signal DD update is available. // service/test CAN messages MSG_ID_FIRST_TD_TESTER_MESSAGE = 0x8000, ///< First TD test message ID @@ -272,6 +285,19 @@ MSG_ID_TD_BLOOD_PRIME_VOLUME_OVERRIDE = 0x8045, ///< TD blood prime override. MSG_ID_TD_BLOOD_PRIME_PUBLISH_INTERVAL_OVERRIDE_REQUEST = 0x8046, ///< TD blood prime publish interval broadcast override. MSG_ID_TD_ENABLE_VENOUS_BUBBLE_ALARM = 0x8047, ///< TD enable venous bubble alarm. + MSG_ID_TD_SYRINGE_PUMP_OPERATION_REQUEST = 0x8048, ///< TD syringe pump operation request + MSG_ID_HD_SYRINGE_PUMP_PUBLISH_INTERVAL_OVERRIDE = 0x8049, ///< TD syringe pump data publish interval override request + MSG_ID_TD_SYRINGE_PUMP_RATE_OVERRIDE_REQUEST = 0x8050, ///< TD syringe pump measured rate override request + MSG_ID_TD_SYRINGE_PUMP_FORCE_OVERRIDE_REQUEST = 0x8051, ///< TD syringe pump measured force override request + MSG_ID_TD_SYRINGE_PUMP_HOME_OVERRIDE_REQUEST = 0x8052, ///< TD syringe pump measured home override request + MSG_ID_TD_SYRINGE_PUMP_POSITION_OVERRIDE_REQUEST = 0x8053, ///< TD syringe pump measured position override request + MSG_ID_TD_SYRINGE_PUMP_VOLUME_OVERRIDE_REQUEST = 0x8054, ///< TD syringe pump measured volume override request + MSG_ID_TD_SYRINGE_PUMP_STATUS_OVERRIDE_REQUEST = 0x8055, ///< TD syringe pump status override request + MSG_ID_TD_SYRINGE_PUMP_ENCODER_STATUS_OVERRIDE_REQUEST = 0x8056, ///< TD syringe pump Encoder status override request + MSG_ID_TD_SYRINGE_PUMP_ADC_DAC_STATUS_OVERRIDE_REQUEST = 0x8057, ///< TD syringe pump ADC and DAC status override request + MSG_ID_TD_SYRINGE_PUMP_ADC_READ_COUNTER_OVERRIDE_REQUEST = 0x8058, ///< TD syringe pump ADC read counter override request + MSG_ID_TD_HEPARIN_BOLUS_TARGET_RATE_OVERRIDE_REQUEST = 0x8059, ///< TD heparin bolus target rate override request + MSG_ID_TD_SYRINGE_PUMP_FORCE_SENSOR_CALIBRATION_REQUEST = 0x8060, ///< TD set syringe pump force sensor DAC calibration request MSG_ID_FIRST_DD_TESTER_MESSAGE = 0xA000, ///< First DD test message ID MSG_ID_DD_TESTER_LOGIN_REQUEST = MSG_ID_FIRST_DD_TESTER_MESSAGE, ///< DD tester log-in @@ -534,6 +560,7 @@ REQUEST_REJECT_REASON_ACID_CONV_FACTOR_MISMATCH = 54, ///< Acid conversion factor from UI mismatch with FW REQUEST_REJECT_REASON_INVALID_PATIENT_WEIGHTS = 55, ///< Patient pre and estimated target weights are invalid REQUEST_REJECT_REASON_UF_VOL_AND_WEIGHTS_INCONSISTENT = 56, ///< UF volume and the patient weights are not consistent + REQUEST_REJECT_REASON_TD_AUTHENTICATION_INVALID = 57, ///< TD authentication invalid NUM_OF_REQUEST_REJECT_REASONS ///< Number of settings change reject codes }; typedef enum Request_Reject_Reasons REQUEST_REJECT_REASON_CODE_T; ///< Type for settings change rejection reason codes enumeration Index: TDDefs.h =================================================================== diff -u -r7f5c34c92560f954d2994019415bc7954f5ab562 -rbeacec5020e8d073014576928b5a7d4283bd183c --- TDDefs.h (.../TDDefs.h) (revision 7f5c34c92560f954d2994019415bc7954f5ab562) +++ TDDefs.h (.../TDDefs.h) (revision beacec5020e8d073014576928b5a7d4283bd183c) @@ -7,8 +7,8 @@ * * @file TDDefs.h * -* @author (last) Raghu Kallala -* @date (last) 28-Apr-2026 +* @author (last) Arpita Srivastava +* @date (last) 12-May-2026 * * @author (original) Sean Nash * @date (original) 01-Aug-2024 @@ -223,24 +223,26 @@ /// Enumeration of dry self-tests sub-mode states. enum TD_Pre_Treatment_Dry_Self_Tests_States { - DRY_SELF_TESTS_START_STATE = 0, ///< Dry self-tests starting state - DRY_SELF_TESTS_WAIT_FOR_DOOR_CLOSE_STATE, ///< Wait for door to close before executing self-tests - DRY_SELF_TESTS_USED_CARTRIDGE_CHECK_STATE, ///< Used cartridge check dry self-test state - DRY_SELF_TESTS_CARTRIDGE_LOADED_CHECK_STATE, ///< Cartridge loaded check dry self-test state - DRY_SELF_TESTS_SYRINGE_PUMP_SEEK_STATE, ///< Seek syringe pumps state - DRY_SELF_TESTS_PRESSURE_SENSORS_NORMAL_SETUP_STATE, ///< Pressure sensor setup state - DRY_SELF_TESTS_PRESSURE_SENSORS_VENOUS_SETUP_STATE, ///< Venous pressure sensor dry self-test setup valves and pump state - DRY_SELF_TESTS_PRESSURE_SENSORS_VENOUS, ///< Venous pressure sensor dry self-test - DRY_SELF_TESTS_PRESSURE_SENSORS_ARTERIAL_SETUP_STATE, ///< Arterial pressure sensor dry self-test setup valves and pump state - DRY_SELF_TESTS_PRESSURE_SENSORS_ARTERIAL, ///< Arterial pressure sensor dry self-test - DRY_SELF_TESTS_PRESSURE_SENSORS_DECAY_STATE, ///< Pressure sensors verify pressure loss state - DRY_SELF_TESTS_PRESSURE_SENSORS_STABILITY_STATE, ///< Pressure sensors verify pressure stability state - DRY_SELF_TESTS_PRESSURE_SENSORS_NORMAL_STATE, ///< Pressure sensors verify normal pressure readings state - DRY_SELF_TESTS_SYRINGE_PUMP_PRIME_STATE, ///< Prime syringe pump state - DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_DETECTION_STATE, ///< Occlusion detection state - DRY_SELF_TESTS_STOPPED_STATE, ///< Dry self-test stopped state - DRY_SELF_TESTS_COMPLETE_STATE, ///< Dry self-test complete state - NUM_OF_DRY_SELF_TESTS_STATES ///< Number of dry self-tests states + DRY_SELF_TESTS_START_STATE = 0, ///< Dry self-tests start state + DRY_SELF_TESTS_WAIT_FOR_DOOR_CLOSE_STATE = 0, ///< Wait for door close state + DRY_SELF_TESTS_USED_TUBE_SET_CHECK_STATE, ///< Used tube set check state + DRY_SELF_TESTS_TUBE_SET_LOADED_CHECK_STATE, ///< Tube set loaded check state + DRY_SELF_TESTS_TUBE_SET_AUTHENTICATION_STATE, ///< Tube set authentication state + DRY_SELF_TESTS_SYRINGE_PUMP_SEEK_STATE, ///< Syringe pump seek state + DRY_SELF_TESTS_PRESSURE_SENSOR_NORMAL_SETUP_STATE, ///< Pressure sensor normal setup state + DRY_SELF_TESTS_PRESSURE_VENOUS_SETUP_STATE, ///< Pressure venous setup state + DRY_SELF_TESTS_VENOUS_PRESSURE_STABILIZATION_STATE, ///< Venous pressure stabilization state + DRY_SELF_TESTS_VENOUS_PRESSURE_LEAK_CHECK_STATE, ///< Venous pressure leak check state + DRY_SELF_TESTS_VENOUS_PRESSURE_RELIEF_STATE, ///< Venous pressure relief state + DRY_SELF_TESTS_PRESSURE_ARTERIAL_SETUP_STATE, ///< Pressure arterial setup state + DRY_SELF_TESTS_ARTERIAL_PRESSURE_STABILIZATION_STATE, ///< Arterial pressure stabilization state + DRY_SELF_TESTS_ARTERIAL_PRESSURE_LEAK_CHECK_STATE, ///< Arterial pressure leak check state + DRY_SELF_TESTS_ARTERIAL_PRESSURE_RELIEF_STATE, ///< Arterial pressure relief state + DRY_SELF_TESTS_SYRINGE_PUMP_PRIME_STATE, ///< Syringe prime state + DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_CHECK_STATE, ///< Syringe pump occlusion check state + DRY_SELF_TESTS_COMPLETE_STATE, ///< Complete state + DRY_SELF_TESTS_STOPPED_STATE, ///< Stopped state + NUM_OF_DRY_SELF_TESTS_STATES ///< Number of dry self-tests states }; /// Type for dry self-tests enumeration. typedef enum TD_Pre_Treatment_Dry_Self_Tests_States DRY_SELF_TESTS_STATE_T; @@ -781,7 +783,9 @@ TD_EVENT_STARTUP = 0, ///< TD startup event TD_EVENT_OP_MODE_CHANGE, ///< TD Op mode change event TD_EVENT_SUB_MODE_CHANGE, ///< TD Op sub-mode change event - TD_EVENT_DRY_SELF_TEST_CARTRIDGE_RESULT, ///< TD dry self test cartridge result + TD_EVENT_DRY_SELF_TEST_TUBESET_RESULT, ///< TD dry self test cartridge result + TD_EVENT_TUBING_SET_BARCODE_SCAN_REQUEST, ///< TD tubing set barcode scan request + TD_EVENT_TUBING_SET_BARCODE_SCAN_AUTHENTICATED, ///< TD tubing set barcode scan authenticated TD_EVENT_DRY_SELF_TEST_PRESSURE_RESULT, ///< TD dry self test pressure result TD_EVENT_WET_SELF_TEST_DISPLACEMENT_RESULT, ///< TD wet self test displacement result TD_EVENT_CPU_RAM_ERROR_STATUS, ///< TD CPU RAM error status @@ -798,10 +802,10 @@ TD_EVENT_AIR_TRAP_FILL, ///< TD started/stopped an air trap fill operation. TD_EVENT_AIR_TRAP_LOWER, ///< TD started/stopped an air trap lower level operation. TD_EVENT_AIR_PUMP_ON_OFF, ///< TD turned air pump on/off. - TD_EVENT_DRY_SELF_TEST_PRESSURE_DECAY_WAIT_PERIOD, ///< TD dry self test pressure decay waiting period. TD_EVENT_INSTIT_RECORD_UPDATE, ///< TD new institutional record has been updated. TD_EVENT_VALVE_POS_CHANGE, ///< TD pinch valve position change TD_EVENT_VALVE_HOMED_POS_SETTING, ///< TD pinch valve homed encoder positions for A/B/C. + TD_EVENT_SYRINGE_PUMP_STATE, ///< TD syringe pump state change event NUM_OF_TD_EVENT_IDS ///< Total number of TD events }; typedef enum TD_Event_List TD_EVENT_ID_T; ///< Type for TD event list enumeration