Index: configurations/Parameters/DataList.conf =================================================================== diff -u -r33ce323edd28935192fdb15010fd51eaa610aa58 -rba4926fb8bcf11efa28d635ccc174cab80015193 --- configurations/Parameters/DataList.conf (.../DataList.conf) (revision 33ce323edd28935192fdb15010fd51eaa610aa58) +++ configurations/Parameters/DataList.conf (.../DataList.conf) (revision ba4926fb8bcf11efa28d635ccc174cab80015193) @@ -2,6 +2,14 @@ Fail Pass +[Treatment Modality] +HD +HDF + +[HDF Treatment Mode] +Pre-Dilution +Post-Dilution + [Blood Flow Rate Ranges] # U32 - mL/min Blood_Flow_Rate_Def = +50 # LEAHI-SRS-287 Blood_Flow_Rate_Min = +50 # LEAHI-SRS-287 @@ -23,7 +31,7 @@ [Heparin Dispensing Rate Ranges] # F32 - mL/hr (0 = OFF) Heparin_Dispensing_Rate_Def = +0.0 # LEAHI-SRS-296 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. Heparin_Dispensing_Rate_Min = +0.2 # LEAHI-SRS-296 -Heparin_Dispensing_Rate_Max = +1.0 # LEAHI-SRS-296 +Heparin_Dispensing_Rate_Max = +3.0 # LEAHI-PRS 215 && PRS 238 Heparin_Dispensing_Rate_Res = +0.1 # LEAHI-SRS-296 [Heparin Bolus Volume Ranges] # F32 - mL (0 = OFF) @@ -48,10 +56,32 @@ Dialysate_Temperature_Res = +0.5 # LEAHI-SRS-303 [Acid Concentrate Options] # U32 -1.0 K, 2.5 Ca, 1 Mg # LEAHI-SRS-288 -2.0 K, 2.5 Ca, 1 Mg # LEAHI-SRS-288 -3.0 K, 2.5 Ca, 1 Mg # LEAHI-SRS-288 +1.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 +2.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 +3.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 +[Acid Concentrate Ranges] # K +Acid_Concentrate_Potassium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Potassium_Min = +0.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Potassium_Max = +4.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Potassium_Res = +1.0 # LEAHI-PRS-660, LEAHI-PRS-241 + # Ca +Acid_Concentrate_Calcium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Calcium_Min = +0.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Calcium_Max = +3.5 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Calcium_Res = +0.25 # LEAHI-PRS-660, LEAHI-PRS-241 + +[Dialysate Composition Ranges] # mEq/L +Dialysate_Composition_Bicarbonate_Def = +137 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Bicarbonate_Min = +130 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Bicarbonate_Max = +155 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Bicarbonate_Res = +1 # LEAHI-PRS-215, LEAHI-PRS-238 + +Dialysate_Composition_Sodium_Def = +33 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Sodium_Min = +20 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Sodium_Max = +40 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Sodium_Res = +1 # LEAHI-PRS-215, LEAHI-PRS-238 + [Bicarbonate Concentrate Options] # U32 Sodium Bicarbonate # LEAHI-SRS-289 @@ -64,11 +94,28 @@ Optiflux F200NRe # LEAHI-SRS-290 Optiflux F250NRe # LEAHI-SRS-290 +[Dry Bicarb Cartridge Size] # U32 +650 g +720 g +760 g +900 g +1100 g +1250 g + [Heparin Type Options] # U32 -Unfractionated 1,000 IU/mL # LEAHI-SRS-291 +Unfractionated 1,000 # LEAHI-SRS-291 +[Vitals Measurement Interval Options] # min (0 = OFF) +0 # LEAHI-PRS-131 +5 # LEAHI-PRS-131 +10 # LEAHI-PRS-131 +15 # LEAHI-PRS-131 +20 # LEAHI-PRS-131 +30 # LEAHI-PRS-131 +60 # LEAHI-PRS-131 + [Trancememberance Pressure Ranges] -# Adjustment, Arterial Window +# Adjustment, Trancememberance Window Trancememberance_Adjust_Window_Def = +40 # LEAHI-PRS-215 ? Trancememberance_Adjust_Window_Min = +40 # LEAHI-PRS-255 ? Trancememberance_Adjust_Window_Max = +100 # LEAHI-PRS-255 ? @@ -105,21 +152,72 @@ Venous_Monitor_Min = -100 # LEAHI-PRS-253 LEAHI-SRS-220 Venous_Monitor_Max = +500 # LEAHI-PRS-253 LEAHI-SRS-220 -[Blood Pressure Measure Interval Ranges] # min (0 = OFF) -Blood_Pressure_Measure_Interval_Def = +0 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Min = +5 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Max = +60 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Res = +5 # LEAHI-SRS-304 - [Rinseback Volume Ranges] # mL/min -Rinseback_Volume_Def = 300 # LEAHI-PRS-559 -Rinseback_Volume_Min = 0 # LEAHI-? -Rinseback_Volume_Max = 0 # LEAHI-? -Rinseback_Volume_Res = 0 # LEAHI-? # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore. +Rinseback_Volume_Def = +300 # LEAHI-PRS-559 +Rinseback_Volume_Min = +200 # LEAHI-PRS-238 +Rinseback_Volume_Max = +400 # LEAHI-PRS-238 +Rinseback_Volume_Res = +10 # LEAHI-PRS-238 [Rinseback Flow Rate Ranges] # mL/min Rinseback_Flow_Rate_Def = +200 # LEAHI-PRS-353 Rinseback_Flow_Rate_Min = +100 # LEAHI-PRS-354 -Rinseback_Flow_Rate_Max = +300 # LEAHI-PRS-354 -Rinseback_Flow_Rate_Res = +10 # LEAHI-PRS-354 # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore. +Rinseback_Flow_Rate_Max = +500 # LEAHI-PRS-354 +Rinseback_Flow_Rate_Res = +10 # LEAHI-PRS-354 +[Ultrafiltration Volume Ranges] # L +Ultrafiltration_Volume_Def = +0.0 # LEAHI-PRS-127, LEAHI-PRS-238 +Ultrafiltration_Volume_Min = +0.0 # LEAHI-PRS-127, LEAHI-PRS-238 +Ultrafiltration_Volume_Max = +0.8 # LEAHI-PRS-127, LEAHI-PRS-238 +Ultrafiltration_Volume_Res = +0.1 # LEAHI-PRS-127 + +[Substitution Volume Ranges] # L +Substitution_Volume_Def = +1.0 # LEAHI-PRS-238 +Substitution_Volume_Min = +1.0 # LEAHI-PRS-238 +Substitution_Volume_Max = +30.0 # LEAHI-PRS-238 +Substitution_Volume_Res = +0.1 # LEAHI-PRS-238 + +[Minimum RO Rejection Ratio Alarm] +Min_RO_Rejection_Ratio_Alarm_Def = +90 # LEAHI-PRS-223 +Min_RO_Rejection_Ratio_Alarm_Min = +1 # LEAHI-PRS-223 +Min_RO_Rejection_Ratio_Alarm_Max = +99 # LEAHI-PRS-223 +Min_RO_Rejection_Ratio_Alarm_Res = +1 # LEAHI-PRS-223 + +[Disinfection Frequency] +Disinfection_Frequency_Def = +3 # LEAHI-PRS-223 +Disinfection_Frequency_Min = +1 # LEAHI-PRS-223 +Disinfection_Frequency_Max = +5 # LEAHI-PRS-223 +Disinfection_Frequency_Res = +1 # LEAHI-PRS-223 + +[Disinfection Parameters Cycle Time] +Disinfection_Parameters_Cycle_Time_Def = +30 # LEAHI-PRS-223 +Disinfection_Parameters_Cycle_Time_Min = +1 # LEAHI-PRS-223 +Disinfection_Parameters_Cycle_Time_Max = +60 # LEAHI-PRS-223 +Disinfection_Parameters_Cycle_Time_Res = +1 # LEAHI-PRS-223 + +[Maximum Blood Pump Stop Time] +Max_Blood_Pump_Stop_Time_Def = +5 # LEAHI-PRS-223 +Max_Blood_Pump_Stop_Time_Min = +5 # LEAHI-PRS-223 +Max_Blood_Pump_Stop_Time_Max = +30 # LEAHI-PRS-223 +Max_Blood_Pump_Stop_Time_Res = +1 # LEAHI-PRS-223 + +[Inlet Water Cond. Low Threshold] +Inlet_Water_Cond_Low_Threshold_Def = +200 # LEAHI-PRS-223 +Inlet_Water_Cond_Low_Threshold_Min = +20 # LEAHI-PRS-223 +Inlet_Water_Cond_Low_Threshold_Max = +200 # LEAHI-PRS-223 +Inlet_Water_Cond_Low_Threshold_Res = +10 # LEAHI-PRS-223 + +[Inlet Water Cond. High Threshold] +Inlet_Water_Cond_High_Threshold_Def = +50 # LEAHI-PRS-223 +Inlet_Water_Cond_High_Threshold_Min = +20 # LEAHI-PRS-223 +Inlet_Water_Cond_High_Threshold_Max = +200 # LEAHI-PRS-223 +Inlet_Water_Cond_High_Threshold_Res = +10 # LEAHI-PRS-223 + +[Chlorine Water Sample Check] +Chlorine_Water_Sample_Check_Def = +240 # LEAHI-PRS-223 +Chlorine_Water_Sample_Check_Min = +15 # LEAHI-PRS-223 +Chlorine_Water_Sample_Check_Max = +480 # LEAHI-PRS-223 +Chlorine_Water_Sample_Check_Res = +5 # LEAHI-PRS-223 + +[Water Sample Test Result Required] +Disabled +Enabled