Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -r33ce323edd28935192fdb15010fd51eaa610aa58 -rbcafc1d20046aaf42e7f697f9c33376e0b6b6126 --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 33ce323edd28935192fdb15010fd51eaa610aa58) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision bcafc1d20046aaf42e7f697f9c33376e0b6b6126) @@ -1,775 +1,752 @@ [0] Title = Indication for when no alarms is a possible situation ListTitle = -Message = [1] Title = Service Required: Hemodialysis Device ListTitle = TD Software fault. Software found itself in an unexpected state -Message = TD Fault: SW Invalid State [2] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state -Message = DD Fault: Software Invalid State [3] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state -Message = DD Fault: Software Invalid State [4] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA POST failure -Message = TD POST: FPGA Compatibility [5] Title = Service Required: Hemodialysis Device ListTitle = TD Watchdog POST failure -Message = TD POST: WD Safestate [6] Title = Service Required: Hemodialysis Device ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up -Message = TD POST: Proc- UI Comm [7] Title = Service Required: Hemodialysis Device ListTitle = TD processor clock speed checks against FPGA clock failure -Message = TD Fault: Proc- FPGA Clock [8] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA communication down for too long -Message = TD Fault: FPGA Com [9] Title = Service Required: Hemodialysis Device ListTitle = User interface communication timeout -Message = TD Fault: UI Comm [10] Title = Communication Error ListTitle = Dialysate generator communication timeout -Message = Loss of DD Comm (Timeout) [11] Title = Service Required: Hemodialysis Device ListTitle = Too many bad CRCs detected on received system messages -Message = TD Fault: CAN CRC Com [12] Title = Service Required: Hemodialysis Device ListTitle = System message that required acknowledgment was not acknowledged -Message = TD Fault: CAN ACK UI Com [13] Title = Communication Error ListTitle = TD to DD comm error. -Message = Loss of DD Comm (DD ACK) [14] Title = Service Required: Hemodialysis Device ListTitle = TD critical data integrity check failed -Message = TD Fault: Critical Data Err [15] Title = Service Required: Dialysate Device ListTitle = DD critical data integrity check failed -Message = DD Fault: Critical Data Err [16] Title = Service Required: Hemodialysis Device ListTitle = TD watchdog expired -Message = TD Fault: WD Error [17] Title = Service Required: Dialysate Device ListTitle = DD watchdog expired -Message = DD Fault: WD Error [18] Title = Service Required: Dialysate Device ListTitle = Too many bad CRCs detected on received system messages -Message = DD Fault: CAN CRC Com [19] Title = Service Required: Dialysate Device ListTitle = DD to RO comm error -Message = Loss of RO Comm (RO ACK) [20] Title = Service Required: Dialysate Device ListTitle = DD to TD comm error -Message = Loss of TD Comm (TD ACK) [21] Title = Service Required: Dialysate Device ListTitle = DD FPGA POST failure -Message = DD POST: FPGA Compatibility [22] Title = Service Required: Dialysate Device ListTitle = FPGA clock speed failure -Message = DD Fault: Proc- FPGA Clock [23] Title = Service Required: Dialysate Device ListTitle = DD FPGA communication down for too long -Message = DD Fault: FPGA Com [24] Title = Service Required: Hemodialysis Device ListTitle = TD alarm audio failed POST -Message = TD POST: Primary Alarm Current [25] Title = Service Required: Dialysate Device ListTitle = DD FPGA not accepting commanded valve states -Message = DD Fault: Valve Control [26] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor reporting error or not being read successfully -Message = TD Arterial Pressure Sensor Error [27] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor reporting error or not being read successfully -Message = TD Venous Pressure Sensor Error [28] -Title = Service Required: Dialysate Device +Title = Low TMP Pressure ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully -Message = DD Hyd outlet Pressure Sensor Error [29] -Title = Service Required: Dialysate Device +Title = High TMP Pressure ListTitle = DD bibag pressure sensor reporting error or not being read successfully -Message = DD Bibag Pressure Sensor Error [30] -Title = Service Required: Dialysate Device -ListTitle = DD spent dialysate pressure sensor reporting error or not being read successfully -Message = DD Spent Dial Pressure Sensor Error +Title = Self Test: Venous Air Detector +ListTitle = TD venous air bubble detector self-test failure [31] -Title = Service Required: Dialysate Device -ListTitle = DD fresh dialysate pressure sensor reporting error or not being read successfully -Message = DD Fresh Dial Pressure Sensor Error +Title = MAX Blood Prime Pause Time +ListTitle = Blood Prime pause approaching timeout (> 4 min). [32] -Title = Service Required: Dialysate Device -ListTitle = DD transmembrane sensor reporting error or not being read successfully -Message = DD Transmemb Pressure Sensor Error +Title = MAX Blood Prime Pause Time Reached +ListTitle = Blood Prime pause timeout reached (> 5 min). [33] Title = Service Required: Hemodialysis Device ListTitle = TD monitored voltage is out of range -Message = TD Fault: Voltage Monitor [34] Title = Power Loss ListTitle = TD AC power lost alarm -Message = Loss of AC Power [35] Title = Power Loss ListTitle = TD AC power lost alarm -Message = Blood Stop: Loss of AC Power [36] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully -Message = DD Conductivity Sensor Error [37] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully -Message = DD Conductivity Sensor Error [38] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully -Message = DD Conductivity Sensor Error [39] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully -Message = DD Conductivity Sensor Error [40] Title = Venous Air Detected ListTitle = TD venous air bubble detected alarm -Message = Blood Stop: Venous Air Bubble [41] Title = Blood Pump Stopped: Door Open ListTitle = TD cartridge door opened alarm -Message = Blood Stop: Door Open [42] Title = Stuck Button ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up -Message = TD POST: Stuck Button [43] Title = Low Arterial Pressure ListTitle = Arterial pressure too low during treatment -Message = Blood Stop: Arterial Pres. Low [44] Title = High Arterial Pressure ListTitle = Arterial pressure too high during treatment -Message = Blood Stop: Arterial Pres. High [45] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor is reading out of range -Message = TD Fault: Arterial Press Range [46] Title = Self Test: Cartridge ListTitle = TD arterial pressure self-test failure alarm -Message = Process: Pre- Tx Remove Cartridge [47] Title = Low Venous Pressure ListTitle = Venous pressure too low during treatment -Message = Blood Stop: Venous Pres. Low [48] Title = High Venous Pressure ListTitle = Venous pressure too high during treatment -Message = Blood Stop: Venous Pres. High [49] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor is reading out of range -Message = TD Fault: Venous Press Range [50] Title = Self Test: Cartridge ListTitle = TD venous pressure self-test failure alarm -Message = Process: Pre- Tx Remove Cartridge [51] Title = Self Test: Cartridge Check ListTitle = TD pre-treatment mode dry self-tests failure -Message = Process: Pre- Tx Dry Self- Test [52] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D11_PUMP speed control error -Message = DD Fault: Pump Speed Fault [53] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D10_PUMP speed control error -Message = DD Fault: Pump Speed Fault [54] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump hall sensor out of range. -Message = DD Fault: Pump Speed Fault [55] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump fault -Message = DD Fault: Pump Fault [56] Title = Service Required: Dialysate Device ListTitle = DD temperature sensor out of range -Message = DD Fault: Temp Sensor Range [57] Title = Service Required: Dialysate Device ListTitle = DD temperature sensors not being read successfully -Message = DD Fault: Temperature RTD Comm [58] Title = Service Required: Dialysate Device -ListTitle = DD barometric sensor not being read successfully -Message = DD Fault: Baro temp Comm +ListTitle = DD Balancing chamber fill timeout error [59] Title = Service Required: Dialysate Device ListTitle = DD barometric pressure sensor bad coefficients CRC -Message = DD Fault: Baro temp Cal CRC [60] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump RPM out of range -Message = DD Fault: Fresh Dial Pump Speed Fault [61] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump off fault -Message = DD Fault: Fresh Dial Pump Speed Not Off [62] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump current out of range -Message = DD Fault: Fresh Dial Pump Current [63] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump invalid direction -Message = DD Fault: Fresh Dial Pump Dir Fault [64] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump direction FPGA fault -Message = DD Fault: Fresh Dial pump Dir Err Cnt [65] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump RPM out of range -Message = DD Fault: Spent Dial Pump Speed Fault [66] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump off fault -Message = DD Fault: Spent Dial Pump Speed Not Off [67] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump current out of range -Message = DD Fault: Spent Dial Pump Current [68] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump invalid direction -Message = DD Fault: Spent Dial Pump Dir Fault [69] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump direction FPGA fault -Message = DD Fault: Spent Dial pump Dir Err Cnt [70] Title = Service Required: Dialysate Device ListTitle = DD D5 heater is On with low fluid level -Message = DD Fault: Heater On With No Fluid [71] Title = Service Required: Dialysate Device ListTitle = DD D45 heater is On with low fluid level -Message = DD Fault: Heater On With No Fluid [72] Title = Service Required: Dialysate Device -ListTitle = DD water inlet input pressure sensor reporting error or not being read successfully -Message = DD Water Inlet Input Press Sensor Error +ListTitle = TD barometric pressure sensor bad coefficients CRC [73] Title = Service Required: Dialysate Device -ListTitle = DD water inlet output pressure sensor reporting error or not being read successfully -Message = DD Water Inlet Output Press Sensor Error +ListTitle = Barometric pressure is out of range [74] -Title = Incoming Water Pressure Input Low -ListTitle = DD Inlet water pressure input too low -Message = DD: Inlet Water Pres Input Low +Title = Service Required: Dialysate Device +ListTitle = DD UF pump D76_PUMP speed control error [75] -Title = Incoming Water Pressure Input High -ListTitle = DD inlet water pressure input is too high -Message = DD: Inlet Water Pres Input High +Title = Service Required: Dialysate Device +ListTitle = DD Spent chamber fill state 1 start pressure is not in range [76] -Title = Incoming Water Pressure Output Low -ListTitle = DD Inlet water pressure output too low -Message = DD: Inlet Water Pres Output Low +Title = Service Required: Dialysate Device +ListTitle = DD Spent chamber fill state 2 start pressure is not in range [77] -Title = Incoming Water Pressure Output High -ListTitle = DD inlet water pressure output is too high -Message = DD: Inlet Water Pres Output High +Title = Service Required: Dialysate Device +ListTitle = DD Spent chamber fill timeout error [78] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Air trap fill timeout during treatment -Message = Blood Stop: Air Trap Fill [79] Title = Service Required: Hemodialysis Device ListTitle = Air trap level sensors reporting illegal combination of air/fluid -Message = TD Fault: Air Trap Sensor [80] Title = Service Required: Hemodialysis Device ListTitle = Pinch valve reporting a fault -Message = TD Fault: Pinch Valve [81] Title = Service Required: Hemodialysis Device ListTitle = TD valve homing failed -Message = TD Fault: Valve Home [82] Title = Service Required: Hemodialysis Device ListTitle = TD valve transition time out -Message = TD Fault: Valve Transition Err [83] Title = Communication Error ListTitle = DD to RO comm error. -Message = RO Loss of DD Comm (DD ACK) [84] Title = Service Required: Dialysate Device ListTitle = RO FPGA communication down for too long -Message = RO Fault: FPGA Comm [85] Title = Service Required: Hemodialysis Device ListTitle = RO Too many bad CRCs detected on received system messages -Message = RO Fault: CAN CRC Comm [86] Title = Service Required: Filtration Device ListTitle = RO FPGA not accepting commanded valve states -Message = RO Fault: Valve Control [87] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill start pressure is not in range -Message = DD Fault: BC Pressure not in range [88] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill end pressure is not in range -Message = DD Fault: BC Pressure not in range [89] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill start pressure is not in range -Message = DD Fault: BC Pressure not in range [90] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill end pressure is not in range -Message = DD Fault: BC Pressure not in range [91] Title = Service Required: Dialysate Device ListTitle = DD hydraulics chamber degassing pressure is not in range -Message = DD Fault: Hyd Cham Neg Pres not in range [92] -Title = -ListTitle = -Message = +Title = Service Required: Dialysate Device +ListTitle = DD Bicarb chamber fill timeout error [93] -Title = -ListTitle = -Message = +Title = Service Required: Dialysate Device +ListTitle = DD D79 rinse pump commanded vs measured speed out of tolerance [94] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range -Message = DD Fault: BC Pressure not in range [95] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range -Message = DD Fault: BC Pressure not in range [96] Title = Blood Pump Stopped: Treatment Paused ListTitle = TD treatment stopped by user action - pressed stop button -Message = Blood Stopped: Treatment Pause [97] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully -Message = DD Conductivity Sensor Error [98] Title = Approaching MAX Blood Pump Stop Time ListTitle = Blood sitting too long warning (> 4 min) -Message = Blood Stopped: Warning [99] Title = Blood Pump Stop Time Exceeded ListTitle = Blood sitting too long after treatment stopped by user alarm (> 5 min) -Message = Blood Stop: Blood Clot Risk [100] Title = Service Required: Hemodialysis Device ListTitle = TD communication timeout -Message = TD Fault: UI- Proc Comm [101] Title = Service Required: Hemodialysis Device ListTitle = UI POST OS Version compatibility failure -Message = TD POST: OS Version [102] Title = Service Required: Hemodialysis Device ListTitle = UI POST Application Integrity (Sha256Sum) failure -Message = TD POST: UI Image [103] Title = Service Required: Hemodialysis Device ListTitle = UI POST CANBus failure -Message = TD POST: UI CAN Driver [104] Title = Service Required: Hemodialysis Device ListTitle = If the UI POST reports display failure. -Message = UI POST Display failure [105] Title = Service Required: Hemodialysis Device ListTitle = UI POST Touch failure -Message = TD POST: UI Touch Driver [106] Title = Service Required: Hemodialysis Device ListTitle = UI POST SD-Card failure -Message = TD POST: UI SD- Card [107] Title = Service Required: Hemodialysis Device ListTitle = UI POST RTC failure -Message = TD POST: UI RTC [108] Title = Service Required: Hemodialysis Device ListTitle = UI POST WiFi failure -Message = TD POST: UI WiFi [109] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bluetooth failure -Message = TD POST: UI Bluetooth [110] Title = Service Required: Hemodialysis Device ListTitle = UI POST Ethernet failure -Message = TD POST: UI Ethernet [111] Title = Service Required: Hemodialysis Device ListTitle = UI POST Sound failure -Message = TD POST: UI Sound [112] Title = Service Required: Hemodialysis Device ListTitle = UI POST cloud sync POST failure -Message = TD POST: UI CloudSync [113] Title = Service Required: Hemodialysis Device ListTitle = UI POST Invalid Year -Message = TD POST Alarm: UI Invalid Year [114] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bad Settings -Message = TD POST Alarm: UI Bad Settings [115] Title = Service Required: Hemodialysis Device ListTitle = SD card is non-operational -Message = TD Fault: UI- SD- Card Fault [116] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [117] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [118] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [119] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [120] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [121] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [122] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [123] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [124] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [125] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [126] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [127] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [128] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [129] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [130] Title = Service Required: Hemodialysis Device ListTitle = RESERVED -Message = TD Fault: UI- RESERVED [131] Title = Battery Capacity Low ListTitle = Insufficient battery charge to support 20 min without A/C power -Message = TD Alarm (Low): Battery Low [132] Title = High Dialysate Temperature ListTitle = Dialysate temperature is out of high safety range -Message = Dial. Gen: Dialysate Temp High [133] Title = High Dialysate Temperature ListTitle = Dialysate temperature is above the target temperature -Message = Dial. Gen: Dialysate Temp High [134] Title = Low Dialysate Temperature ListTitle = Dialysate temperature is below the target temperature -Message = Dial. Gen: Dialysate Temp Low [135] Title = Service Required: FP Device ListTitle = FP pressure relief is too low -Message = FP Fault: Pressure Relief Valve Low [136] Title = Service Required: FP Device ListTitle = FP pressure relief is too High -Message = FP Fault: Pressure Relief Valve High [137] Title = Service Required: FP Device ListTitle = FP device is using too much power -Message = FP Fault: Power usage has exceeded budget [138] Title = Service Required: FP Device ListTitle = FP RO pump flow out of range -Message = FP Fault: RO Pump Flow Low [139] Title = Low Inlet Water Temperature ListTitle = Inlet Water temperature is too low -Message = FP Water Gen: Inlet Water Temp Low [140] Title = High Inlet Water Temperature ListTitle = Inlet Water temperature is too high -Message = FP Water Gen: Inlet Water Temp High [141] Title = Low Inlet Water Pressure ListTitle = Inlet Water pressure is too low -Message = FP Water Gen: Inlet Pressure Low [142] Title = High Inlet Water Pressure ListTitle = Inlet Water pressure is too high -Message = FP Water Gen: Inlet Pressure High [143] Title = High Inlet Water Conductivity ListTitle = Inlet Water conductivity is too high -Message = FP Water Gen: Inlet Conductivity High [144] Title = Low Inlet Water Conductivity ListTitle = Inlet Water conductivity is too low -Message = FP Water Gen: Inlet Conductivity Low [145] Title = Service Required: FP Device ListTitle = FP RO pump pressure out of range -Message = FP Fault: RO Pump High Pressure [146] Title = Service Required: FP Device ListTitle = FP outlet primary conductivity sensor out of range -Message = FP Fault: RO Permeate Cond High [147] Title = Service Required: FP Device ListTitle = FP inlet primary conductivity sensor out of range -Message = FP Fault: RO Permeate Cond Low [148] Title = Service Required: FP Device ListTitle = FP RO rejection ratio out of range -Message = FP Fault: RO Rejection [149] Title = Service Required: FP Device ListTitle = FP pressure regulator is out of bounds -Message = FP Fault: Pressure regulator not in range [150] -Title = FP: Filter Health -ListTitle = FP pressure difference between the filters is too high -Message = FP Filters: Pressure difference too high +Title = Service Required: FP Device +ListTitle = FP inlet temperature too high during filters flush [151] Title = FP: Filter Health ListTitle = FP conductivity after flushing the system is too high -Message = FP Filters: Permeate conductivity high [152] Title = FP: Filter Health -ListTitle = FP volume difference between integrated flow and expected volume -Message = FP Filters: Volume difference too high +ListTitle = FP took more than 60 seconds to fill the permeate tank [153] Title = FP: Permeate Overfill ListTitle = FP tank level is too high -Message = FP Tank: Tank is too full [154] Title = Service Required: Dialysate Device ListTitle = DD dialysate temperature sensors drift timeout. -Message = DD Fault: D28 and D30 Delta +[155] +Title = Blood Leak +ListTitle = DD blood leak detected alarm + +[156] +Title = Service Required: Hemodialysis Device +ListTitle = DD blood leak sensor invalid calibration record + +[157] +Title = Service Required: Hemodialysis Device +ListTitle = DD blood leak sensor embedded mode communication failed + +[158] +Title = Service Required: Hemodialysis Device +ListTitle = DD blood leak sensor set point set failure + +[159] +Title = Blood Leak: Bypass +ListTitle = DD blood leak detected and it is in the process of attempting to remove the bubbles/foam + +[160] +Title = FP: Permeate Underfill +ListTitle = FP tank level is too low + +[161] +Title = High Inlet Water Pressure Fault +ListTitle = Inlet Water pressure is too high + +[162] +Title = High Permeate flow +ListTitle = Permeate flow is too high from set target flow rate + +[163] +Title = low Permeate flow +ListTitle = Permeate flow is too low from set target flow rate + +[164] +Title = low inlet pressure +ListTitle = Inlet pressure is too low + +[165] +Title = Blood Pump Stopped: Venous Air Detector +ListTitle = Air trap lowering timeout during treatment + +[166] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump self-test failure alarm + +[167] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump is running while the BP is off + +[168] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump ADC error + +[169] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump fault reported by FPGA + +[170] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump DAC write failure + +[171] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump speed check error + +[172] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump volume check error + +[173] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump over travel error + +[174] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump not stopped in off state error + +[175] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump direction (from encoder) error + +[176] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump direction (from controller) error + +[177] +Title = Heparin Syringe Empty +ListTitle = TD syringe empty alarm + +[178] +Title = Blood Stopped: Syringe Not Detected +ListTitle = TD syringe pump syringe removed alarm + +[179] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump stall alarm + +[180] +Title = Service Required: Hemodialysis Device +ListTitle = TD heparin force sensor invalid calibration record + +[181] +Title = Syringe Detected +ListTitle = TD syringe detected alarm + +[182] +Title = Not Enough Heparin +ListTitle = TD syringe pump not enough Heparin alarm + +[183] +Title = Heparin Pump Occlusion +ListTitle = TD syringe pump occlusion alarm + +[184] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump timeout alarm + +[185] +Title = Service Required: FP Device +ListTitle = FP conductivity sensor (P9) located in inlet water path reporting error or not being read successfully + +[186] +Title = Service Required: FP Device +ListTitle = FP conductivity sensor (P18) located in inlet water path reporting error or not being read successfully + +[187] +Title = Air Pump Stall Detected +ListTitle = Air pump stall detected during treatment + Index: configurations/Instructions/Concentrates_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Concentrates_2.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Connect_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Connect_2.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Connect_3.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Connect_4.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Connect_Advanced_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Install_Advanced_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Install_Advanced_2.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_BloodTube_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_BloodTube_2.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_BloodTube_3.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_BloodTube_4.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_Bloodlines_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_Bloodlines_2.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_Bloodlines_3.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_Bloodlines_4.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_Bloodlines_5.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_CPLD_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_CPLD_2.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_CPLD_3.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_CPLD_4.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_CPLD_5.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Installation_CPLD_6.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Instructions.conf =================================================================== diff -u -r65465f788ce5d6cc7cdfdaf270650a22ed8c39f6 -rbcafc1d20046aaf42e7f697f9c33376e0b6b6126 --- configurations/Instructions/Instructions.conf (.../Instructions.conf) (revision 65465f788ce5d6cc7cdfdaf270650a22ed8c39f6) +++ configurations/Instructions/Instructions.conf (.../Instructions.conf) (revision bcafc1d20046aaf42e7f697f9c33376e0b6b6126) @@ -8,127 +8,139 @@ # In case extra string or precision is not being used leave it empty eg. {77:1::} # the format shall always have three ':' and message id and parameter number shall not be ignored. # otherwise will NOT be identified as a value reserved block and will be used as part of the string. +# Entry format: +# [^] +# "Title" = +# <Instruction step> = <image.png> +# - OR - +# [<Prefix '$' for HDF mode><Step Tite>] +# "Title" = <Title string> +# <Instruction step> = <image.png> +# Rich Text is supported: +# ex: +# <b></b> - bold +# <i></i> - italic +# <u> - underlined text +# <$HEX>text</$> - color text // Ex. <$2268FF>TEST</$> - Do not add '0x' to HEX value # ###################################################################################################### -[Sample^Water Sample] -Place cup in the water sampling pocket. Press and hold the "Water Sample" button. = 9_water sampling.png +[Concentrates] +Title = Connect Acid and Bicarbonate Concentrates +Remove acid connector and connect to acid concentrate. = Concentrates_1.png +Connect dry bicarbonate. = Concentrates_2.png -[Sample^Water Sample Result] -Test the water sample per your clinic's instructions and enter the result. = 42_water test pass fail.png +[Installation^Blood Tubing Set] +Title = Install Blood Tubing Set +Open the blood pump door. = Installation_BloodTube_1.png +Insert blood tubing set and ensure it is secure. = Installation_BloodTube_2.png +Push the blood tube into the blood pump. = Installation_BloodTube_3.png +Close the blood pump door. = Installation_BloodTube_4.png -[Sample^Water Sample Result Failed] -Water sample result failed. = 40_water test fail.png +[Installation^Bloodlines] +Title = Install Bloodlines +Insert saline line into the first slot. = Installation_Bloodlines_1.png +Insert <$B53333>arterial</$> line into the second slot. = Installation_Bloodlines_2.png +Ensure the <$2268FF>venous</$> line is placed in the bubble detector. = Installation_Bloodlines_3.png +Insert the <$2268FF>venous</$> line into the third slot. = Installation_Bloodlines_4.png +Connect the <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines using the recirculation connector. = Installation_Bloodlines_5.png -[Consumables^Consumables Installation] -Remove the concentrate and dialysate caps. = 47_remove DG caps.png -Connect the acid concentrate to the acid port. = 10_install acid line.png -Connect the bicarbonate concentrate to the bicarbonate port. = 11_install bicarb line.png +[Installation^Pressure Lines and Dialyzer] +Title = Install Pressure Lines and Dialyzer +Connect the <$B53333>arterial</$> pressure line to the <$B53333>arterial</$> pressure transducer port. = Installation_CPLD_1.png +Connect the <$2268FF>venous</$> pressure line to the <$2268FF>venous</$> pressure transducer port. = Installation_CPLD_2.png +Insert dialyzer into the holder and connect the dialysate lines. = Installation_CPLD_3.png +Connect the bloodlines to the dialyzer. = Installation_CPLD_4.png +Clamp the <$2268FF>venous</$> chamber line. = Installation_CPLD_5.png +Clamp the heparin line. = Installation_CPLD_6.png -[Disposables^Cartridge Installation] -Open the front panel door. = 1_HD door open-01.png -Unlock the pump track by moving the lever to its full upward position. = 69_pump track open without cart.png -Install the cartridge using the alignment pins. = 3_insert casette-01.png -Insert tubing into blood and dialysate pumps. = 4_loop casette tubes-01.png -Lock the pump track by moving the lever to its downward position. = 6_Pump track lock-01.png +[Installation^Saline] +Title = Connect Saline Bag +Hang saline bag on the hook. = Saline_1.png +Clamp saline line. = Saline_2.png +Use the saline line to spike saline bag. = Saline_3.png -[Disposables^Cartridge Connection] -Connect arterial and venous bloodlines using the recirculation connector. = 51_shunt patient lines.png -Place the arterial pressure line behind the arterial and venous bloodlines\nand connect to the arterial pressure transducer port. = 70_PBa connection Rev02.png -Connect the venous pressure line to the venous pressure transducer port. = 23_PBo connection.png -Connect the bloodlines to the dialyzer. = 20_connect blood lines to dialyzer.png -Connect the dialysate lines to the dialyzer. = 21_connect dialysate lines to dialyzer.png -Insert dialyzer into the dialyzer holder. = 19_insert dialyzer.png -Clamp the heparin line. = 50_clamp_heparin.png -Clamp venous chamber line. = 49_clamp venous chamber line.png -Connect the dialysate lines to the dialysate fluid ports. = 16_connect dialysate lines to DG.png +[Installation^Substitution] +Title = Connect Substitution Line +Connect substitution line to the appropriate Pre or Post dilution port. = Saline_1.png -[Disposables^Heparin Syringe] -Fill a syringe with {77:2:mL:1} heparin and connect to the heparin tubing line. = 25_fill syringe with heparin.png -Insert heparin filled syringe into the syringe pump. Unclamp heparin line. = 26_load syringe.png +[@Installation] +Title = Install Concentrates and Blood Tubing Set +Install acid and bicarbonate concentrates. = Install_Advanced_1.png +Install blood tubing set, dialyzer, and saline bag. = Install_Advanced_2.png -[Disposables^Saline Bag] -Clamp saline line and y-connector. = 53_pinch saline line and y connector.png -Hang the saline bag on the hook on the left side of the system.\nUse the saline line to spike a saline bag. = 29_spike and hang saline bag.png -Close the front panel door. Ensure lines are not kinked. = 66_hd door close with cart.png -Unclamp the saline line. Leave the y-connector clamped. = 54_unpinch saline and clamped y connector.png +[Connect] +Title = Connect Bloodlines to the Patient +Clamp <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines (2 clamps). = Connect_1.png +Remove recirculation connector from the bloodlines. = Connect_2.png +Connect bloodlines to patient access lines. = Connect_3.png +Unclamp <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines (2 clamps). = Connect_4.png -[Connection^Patient Connection] -Open the front panel door. = 67_hd door open with cart.png -Remove dialyzer from dialyzer holder and invert dialyzer. Reinstall dialyzer into dialyzer holder. = 52_invert_dialyzer.png -Close the front panel door. Ensure lines are not kinked. = 66_hd door close with cart.png -Draw a dialysate sample from the luer lock connection port.\nTest the dialysate sample per your facility's policy = 17_dialysate sample.png -Wash your hands and establish vascular access per your facility's policy. = 43_wash hands.png -Clamp arterial and venous bloodlines. = 56_clamp shunted patient lines.png -Remove arterial and venous line recirculation connector. = 34_remove recirc connector.png -Connect arterial and venous bloodlines to your patient access lines. = 31_connect arterial and venous lines to patient.png +[@Connect] +Title = Connect Bloodlines to the Patient +Connect bloodlines to patient access lines. = Connect_Advanced_1.png -[Connection^Start Treatment] -Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png +[Disconnection] +Title = Remove Concentrates and Blood Tubing Set +Remove blood tubing set and dialyzer. = Install_Advanced_2.png +Remove acid and bicarbonate concentrates. = Install_Advanced_1.png -[^Rinseback Setup] -Clamp arterial bloodlines and disconnect. = 33_disconnect arterial lines.png -Connect arterial bloodline to the saline y-connector. = 35_connect saline y.png -Unclamp arterial bloodline and saline y-connector. = 38_unclamp saline lines.png +[Water Sample] +Title = Test Primary Carbon Filter +Test the primary carbon filter and select the result below. = Water_Sample_1.png -[^Rinseback Complete] -If rinseback is complete, clamp arterial bloodline and saline y-connector.\nFollow your facility's policy prior to disconnecting patient. = 64_rinseback clamp arterial and saline lines.png +[Blood Return Setup] +Title = Blood Return Connection +Clamp <$B53333>arterial</$> bloodlines and disconnect. = Return_Blood_1.png +Connect <$B53333>arterial</$> bloodline to the saline y-connector. = Return_Blood_2.png +Unclamp <$B53333>arterial</$> bloodline and saline y-connector. = Return_Blood_3.png -[^Rinseback Reconnect] -Clamp arterial bloodline and saline y-connector. Disconnect the arterial bloodline. = 48_disconnect saline y.png -Connect arterial bloodline to your patient access line. = 71_rinseback arterial re-connect.png -Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png +[$Blood Return Setup] +Title = Blood Return Connection +Clamp <$B53333>arterial</$> bloodlines and disconnect. +Connect <$B53333>arterial</$> bloodline to the substitution line using a sterile connector. +Unclamp <$B53333>arterial</$> bloodline and substitution line. -[Disconnection^Patient Disconnection] -Clamp venous bloodlines and disconnect. = 63_disconnect venous lines.png -Connect the venous bloodline to the luer lock connection port on the venous line. = 58_end treatment connect venous to sampling port.png +[Recirculate Set Up] +Title = Connect Recirculation Lines +Clamp <$2268FF>venous</$> bloodline and disconnect. +Clamp <$B53333>arterial</$> bloodlines and disconnect from saline y-connector. +Connect <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines using the recirculation connector. +Unclamp <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines. -[Disposables^Disposables Removal] -Open the front panel door. = 67_hd door open with cart.png -Disconnect the arterial and venous pressure lines. = 24_PBo and PBA removal.png -If heparin was used during treatment, remove syringe. = 28_remove syringe.png -Clamp dialysate lines and disconnect from the dialysate ports. = 18_disconect dialysate lines from DG.png -Replace dialysate cap. = 12_replace dialysate line cap.png -Disconnect acid and bicarbonate concentrates lines from the concentrate ports.= 13_remove acid and bicarb connect.png -Replace concentrate cap. = 59_install concentrate caps.png -Unlock the pump track by moving the lever to its upward position. = 8_pump track unlock.png -Remove cartridge, dialyzer, and saline bag from the HD device. = 5_remove casette-01.png -Lock the pump track by moving the lever to its downward position and close front panel door. = 2_HD door closed-01.png +[$Recirculate Set Up] +Title = Connect Recirculation Lines +Clamp <$2268FF>venous</$> bloodline and disconnect. +Clamp <$B53333>arterial</$> and disconnect from the substitution line. +Connect substitution line to the appropriate dilution port. +Connect <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines using the recirculation connector. +Unclamp <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines. +[Reconnect To Patient] +Title = Connect Patient Bloodlines +Clamp <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines. +Remove <$B53333>arterial</$> and <$2268FF>venous</$> line recirculation connector. +Connect <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines to patient access lines. +Unclamp <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines. -[^Recirculate Disconnect] -Clamp arterial bloodline and saline y-connector. Disconnect the arterial bloodline. = 48_disconnect saline y.png -Clamp venous bloodlines and disconnect. = 63_disconnect venous lines.png -Connect arterial and venous bloodlines using the recirculation connector. = 60_recirc shunt patient lines.png -Unclamp arterial and venous bloodlines. = 62_recirc unclamp recirc lines.png +[$Reconnect To Patient] +Title = Connect Patient Bloodlines +Clamp <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines. +Remove <$B53333>arterial</$> and <$2268FF>venous</$> line recirculation connector. +Connect <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines to patient access lines. +Unclamp <$B53333>arterial</$> and <$2268FF>venous</$> bloodlines. -[^Recirculate] -Treatment is paused. Blood and dialysate pumps are recirculating = 61_recirc recirculating pumps.png -[^Recirculate Stopped] -Recirculation is stopped. Blood pump is stopped. = 65_recirc stop TEMP.png +[Recirculate Set Up] +Title = Connect Recirculation Lines +Clamp venous bloodlines and disconnect. +Clamp arterial bloodlines and disconnect from saline y-connector. +Connect arterial and venous bloodlines using the recirculation connector. +Unclamp arterial and venous bloodlines. -[^Recirculate Reconnect] -Clamp arterial and venous bloodlines. = 56_clamp shunted patient lines.png -Remove arterial and venous line recirculation connector. = 34_remove recirc connector.png -Connect arterial and venous bloodlines to your patient access lines. = 31_connect arterial and venous lines to patient.png -Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png +[Reconnect To Patient] +Title = Connect Bloodlines To Patient +Clamp arterial and venous bloodlines. +Remove arterial and veous line recirculation connector. +Connect arterial and venous bloodlines to patient access lines. +Unclamp arterial and venous bloodlines. -[Water Flush^Water Flush] -Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png - -[Heat Disinfect^Heat Disinfect] -Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png - -[Heat Disinfect with Active Cool^Heat Disinfect with Active Cool] -Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png - -[Chemical Disinfect^Chemical Disinfect] -Remove the concentrate cap and connect the chemical disinfectant = 14_connect PAA.png - -[Chemical Flush^Chemical Flush] -Disconnect the chemical disinfectant and replace concentrate cap. = 15_disconnect PAA.png -Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png - -[RO Permeate Sample^RO Permeate Sample] -Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png - Index: configurations/Instructions/Return_Blood_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Return_Blood_2.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Return_Blood_3.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Saline_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Saline_2.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Saline_3.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Saline_4.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Water_Sample_1.png =================================================================== diff -u Binary files differ Index: configurations/Parameters/DataList.conf =================================================================== diff -u -r33ce323edd28935192fdb15010fd51eaa610aa58 -rbcafc1d20046aaf42e7f697f9c33376e0b6b6126 --- configurations/Parameters/DataList.conf (.../DataList.conf) (revision 33ce323edd28935192fdb15010fd51eaa610aa58) +++ configurations/Parameters/DataList.conf (.../DataList.conf) (revision bcafc1d20046aaf42e7f697f9c33376e0b6b6126) @@ -2,6 +2,19 @@ Fail Pass +[Treatment Modality] +HD +HDF + +[HDF Treatment Mode] +Pre-Dilution +Post-Dilution + +[Hepatitis Status] +Unknown +Positive +Negative + [Blood Flow Rate Ranges] # U32 - mL/min Blood_Flow_Rate_Def = +50 # LEAHI-SRS-287 Blood_Flow_Rate_Min = +50 # LEAHI-SRS-287 @@ -23,7 +36,7 @@ [Heparin Dispensing Rate Ranges] # F32 - mL/hr (0 = OFF) Heparin_Dispensing_Rate_Def = +0.0 # LEAHI-SRS-296 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. Heparin_Dispensing_Rate_Min = +0.2 # LEAHI-SRS-296 -Heparin_Dispensing_Rate_Max = +1.0 # LEAHI-SRS-296 +Heparin_Dispensing_Rate_Max = +3.0 # LEAHI-PRS 215 && PRS 238 Heparin_Dispensing_Rate_Res = +0.1 # LEAHI-SRS-296 [Heparin Bolus Volume Ranges] # F32 - mL (0 = OFF) @@ -48,10 +61,32 @@ Dialysate_Temperature_Res = +0.5 # LEAHI-SRS-303 [Acid Concentrate Options] # U32 -1.0 K, 2.5 Ca, 1 Mg # LEAHI-SRS-288 -2.0 K, 2.5 Ca, 1 Mg # LEAHI-SRS-288 -3.0 K, 2.5 Ca, 1 Mg # LEAHI-SRS-288 +1.0 K, 2.50 Ca = +0.11192 # LEAHI-SRS-288 +2.0 K, 2.50 Ca = +0.11313 # LEAHI-SRS-288 +3.0 K, 2.50 Ca = +0.11435 # LEAHI-SRS-288 +[Acid Concentrate Ranges] # K +Acid_Concentrate_Potassium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Potassium_Min = +0.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Potassium_Max = +4.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Potassium_Res = +1.0 # LEAHI-PRS-660, LEAHI-PRS-241 + # Ca +Acid_Concentrate_Calcium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Calcium_Min = +0.0 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Calcium_Max = +3.5 # LEAHI-PRS-660, LEAHI-PRS-241 +Acid_Concentrate_Calcium_Res = +0.25 # LEAHI-PRS-660, LEAHI-PRS-241 + +[Dialysate Composition Ranges] # mEq/L +Dialysate_Composition_Bicarbonate_Def = +33 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Bicarbonate_Min = +20 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Bicarbonate_Max = +40 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Bicarbonate_Res = +1 # LEAHI-PRS-215, LEAHI-PRS-238 + +Dialysate_Composition_Sodium_Def = +137 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Sodium_Min = +130 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Sodium_Max = +155 # LEAHI-PRS-215, LEAHI-PRS-238 +Dialysate_Composition_Sodium_Res = +1 # LEAHI-PRS-215, LEAHI-PRS-238 + [Bicarbonate Concentrate Options] # U32 Sodium Bicarbonate # LEAHI-SRS-289 @@ -64,19 +99,36 @@ Optiflux F200NRe # LEAHI-SRS-290 Optiflux F250NRe # LEAHI-SRS-290 +[Dry Bicarb Cartridge Size] # U32 +650 g +720 g +760 g +900 g +1100 g +1250 g + [Heparin Type Options] # U32 -Unfractionated 1,000 IU/mL # LEAHI-SRS-291 +Unfractionated 1,000 # LEAHI-SRS-291 -[Trancememberance Pressure Ranges] +[Vitals Measurement Interval Options] # min (0 = OFF) +0 # LEAHI-PRS-131 +5 # LEAHI-PRS-131 +10 # LEAHI-PRS-131 +15 # LEAHI-PRS-131 +20 # LEAHI-PRS-131 +30 # LEAHI-PRS-131 +60 # LEAHI-PRS-131 + +[Transmembrane Pressure Ranges] # Adjustment, Arterial Window -Trancememberance_Adjust_Window_Def = +40 # LEAHI-PRS-215 ? -Trancememberance_Adjust_Window_Min = +40 # LEAHI-PRS-255 ? -Trancememberance_Adjust_Window_Max = +100 # LEAHI-PRS-255 ? -Trancememberance_Adjust_Window_Res = +20 # LEAHI-PRS-255 ? +Transmembrane_Adjust_Window_Def = +40 # LEAHI-PRS-215 ? +Transmembrane_Adjust_Window_Min = +40 # LEAHI-PRS-255 ? +Transmembrane_Adjust_Window_Max = +100 # LEAHI-PRS-255 ? +Transmembrane_Adjust_Window_Res = +20 # LEAHI-PRS-255 ? #Monitoring -Trancememberance_Monitor_Res = 1 # LEAHI-PRS-269 LEAHI-SRS-222 -Trancememberance_Monitor_Min = -650 # LEAHI-PRS-269 LEAHI-SRS-222 -Trancememberance_Monitor_Max = +250 # LEAHI-PRS-269 LEAHI-SRS-222 +Transmembrane_Monitor_Res = 1 # LEAHI-PRS-269 LEAHI-SRS-222 +Transmembrane_Monitor_Min = -650 # LEAHI-PRS-269 LEAHI-SRS-222 +Transmembrane_Monitor_Max = +250 # LEAHI-PRS-269 LEAHI-SRS-222 [Arterial Blood Pressure Ranges] # Adjustment, Arterial Window @@ -105,21 +157,91 @@ Venous_Monitor_Min = -100 # LEAHI-PRS-253 LEAHI-SRS-220 Venous_Monitor_Max = +500 # LEAHI-PRS-253 LEAHI-SRS-220 -[Blood Pressure Measure Interval Ranges] # min (0 = OFF) -Blood_Pressure_Measure_Interval_Def = +0 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Min = +5 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Max = +60 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Res = +5 # LEAHI-SRS-304 - [Rinseback Volume Ranges] # mL/min -Rinseback_Volume_Def = 300 # LEAHI-PRS-559 -Rinseback_Volume_Min = 0 # LEAHI-? -Rinseback_Volume_Max = 0 # LEAHI-? -Rinseback_Volume_Res = 0 # LEAHI-? # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore. +Rinseback_Volume_Def = +300 # LEAHI-PRS-559 +Rinseback_Volume_Min = +300 # LEAHI-PRS-238 +Rinseback_Volume_Max = +500 # LEAHI-PRS-238 +Rinseback_Volume_Res = +10 # LEAHI-PRS-238 [Rinseback Flow Rate Ranges] # mL/min Rinseback_Flow_Rate_Def = +200 # LEAHI-PRS-353 Rinseback_Flow_Rate_Min = +100 # LEAHI-PRS-354 -Rinseback_Flow_Rate_Max = +300 # LEAHI-PRS-354 -Rinseback_Flow_Rate_Res = +10 # LEAHI-PRS-354 # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore. +Rinseback_Flow_Rate_Max = +500 # LEAHI-PRS-354 +Rinseback_Flow_Rate_Res = +10 # LEAHI-PRS-354 +[Ultrafiltration Volume Ranges] # L +Ultrafiltration_Volume_Def = +0.0 # LEAHI-PRS-127, LEAHI-PRS-238 +Ultrafiltration_Volume_Min = +0.0 # LEAHI-PRS-127, LEAHI-PRS-238 +Ultrafiltration_Volume_Max = +8.0 # LEAHI-PRS-127, LEAHI-PRS-238 +Ultrafiltration_Volume_Res = +0.1 # LEAHI-PRS-127 + +[Substitution Volume Ranges] # L +Substitution_Volume_Def = +1.0 # LEAHI-PRS-238 +Substitution_Volume_Min = +1.0 # LEAHI-PRS-238 +Substitution_Volume_Max = +30.0 # LEAHI-PRS-238 +Substitution_Volume_Res = +0.1 # LEAHI-PRS-238 + +[Minimum RO Rejection Ratio Alarm] +Min_RO_Rejection_Ratio_Alarm_Def = +90 # LEAHI-PRS-223 +Min_RO_Rejection_Ratio_Alarm_Min = +1 # LEAHI-PRS-223 +Min_RO_Rejection_Ratio_Alarm_Max = +99 # LEAHI-PRS-223 +Min_RO_Rejection_Ratio_Alarm_Res = +1 # LEAHI-PRS-223 + +[Disinfection Frequency] +Disinfection_Frequency_Def = +3 # LEAHI-PRS-223 +Disinfection_Frequency_Min = +1 # LEAHI-PRS-223 +Disinfection_Frequency_Max = +5 # LEAHI-PRS-223 +Disinfection_Frequency_Res = +1 # LEAHI-PRS-223 + +[Disinfection Parameters Cycle Time] +Disinfection_Parameters_Cycle_Time_Def = +30 # LEAHI-PRS-223 +Disinfection_Parameters_Cycle_Time_Min = +1 # LEAHI-PRS-223 +Disinfection_Parameters_Cycle_Time_Max = +60 # LEAHI-PRS-223 +Disinfection_Parameters_Cycle_Time_Res = +1 # LEAHI-PRS-223 + +[Maximum Blood Pump Stop Time] +Max_Blood_Pump_Stop_Time_Def = +5 # LEAHI-PRS-223 +Max_Blood_Pump_Stop_Time_Min = +5 # LEAHI-PRS-223 +Max_Blood_Pump_Stop_Time_Max = +30 # LEAHI-PRS-223 +Max_Blood_Pump_Stop_Time_Res = +1 # LEAHI-PRS-223 + +[Inlet Water Cond. Low Threshold] +Inlet_Water_Cond_Low_Threshold_Def = +200 # LEAHI-PRS-223 +Inlet_Water_Cond_Low_Threshold_Min = +20 # LEAHI-PRS-223 +Inlet_Water_Cond_Low_Threshold_Max = +200 # LEAHI-PRS-223 +Inlet_Water_Cond_Low_Threshold_Res = +10 # LEAHI-PRS-223 + +[Inlet Water Cond. High Threshold] +Inlet_Water_Cond_High_Threshold_Def = +50 # LEAHI-PRS-223 +Inlet_Water_Cond_High_Threshold_Min = +20 # LEAHI-PRS-223 +Inlet_Water_Cond_High_Threshold_Max = +200 # LEAHI-PRS-223 +Inlet_Water_Cond_High_Threshold_Res = +10 # LEAHI-PRS-223 + +[Chlorine Water Sample Check] +Chlorine_Water_Sample_Check_Def = +240 # LEAHI-PRS-223 +Chlorine_Water_Sample_Check_Min = +15 # LEAHI-PRS-223 +Chlorine_Water_Sample_Check_Max = +480 # LEAHI-PRS-223 +Chlorine_Water_Sample_Check_Res = +5 # LEAHI-PRS-223 + +[Water Sample Test Result Required] +Disabled +Enabled + +[Isolated UF Duration Ranges] +Isolated_UF_Duration_Def = +30 # LEAHI-PRS-125 +Isolated_UF_Duration_Min = +30 # LEAHI-PRS-125 +Isolated_UF_Duration_Max = +480 # LEAHI-PRS-125 +Isolated_UF_Duration_Res = +15 # LEAHI-PRS-125 + +[Weight Ranges] +Weight_Def = +0.0 # LEAHI-PRS-215 +Weight_Min = +0.0 # LEAHI-PRS-215 +Weight_Max = +200.0 # LEAHI-PRS-215 +Weight_Res = +0.1 # LEAHI-PRS-215 + +[Prime Discard Volume Ranges] +Prime_Discard_Volume_Def = +300 # LEAHI-PRS-215 +Prime_Discard_Volume_Min = +300 # LEAHI-PRS-215 +Prime_Discard_Volume_Max = +1000 # LEAHI-PRS-215 +Prime_Discard_Volume_Res = +100 # LEAHI-PRS-215 + Index: configurations/Settings/System.conf =================================================================== diff -u -r65465f788ce5d6cc7cdfdaf270650a22ed8c39f6 -rbcafc1d20046aaf42e7f697f9c33376e0b6b6126 --- configurations/Settings/System.conf (.../System.conf) (revision 65465f788ce5d6cc7cdfdaf270650a22ed8c39f6) +++ configurations/Settings/System.conf (.../System.conf) (revision bcafc1d20046aaf42e7f697f9c33376e0b6b6126) @@ -1,10 +1,18 @@ +[AdvancedMode] +AdvancedMode = 0 + [Alarm] Volume = 5 [Date] Format = MM/dd/yyyy +[Feature Configurations] +IntegratedBPCuff = 0 +HeparinSyringePump = 1 +ChemicalDisinfection = 0 + [RoWaterMode] RoWaterMode = 0 Index: configurations/Settings/System.dflt =================================================================== diff -u -r65465f788ce5d6cc7cdfdaf270650a22ed8c39f6 -rbcafc1d20046aaf42e7f697f9c33376e0b6b6126 --- configurations/Settings/System.dflt (.../System.dflt) (revision 65465f788ce5d6cc7cdfdaf270650a22ed8c39f6) +++ configurations/Settings/System.dflt (.../System.dflt) (revision bcafc1d20046aaf42e7f697f9c33376e0b6b6126) @@ -5,6 +5,11 @@ [Date] Format = MM/dd/yyyy +[Feature Configurations] +IntegratedBPCuff = 0 +HeparinSyringePump = 1 +ChemicalDisinfection = 0 + [RoWaterMode] RoWaterMode = 0 @@ -13,3 +18,6 @@ [Time] Format = HH:mm:ss + +[AdvancedMode] +AdvancedMode = 0 Index: leahi =================================================================== diff -u -r33ce323edd28935192fdb15010fd51eaa610aa58 -rbcafc1d20046aaf42e7f697f9c33376e0b6b6126 Binary files differ Index: scripts/_errors_ =================================================================== diff -u -r33ce323edd28935192fdb15010fd51eaa610aa58 -rbcafc1d20046aaf42e7f697f9c33376e0b6b6126 --- scripts/_errors_ (.../_errors_) (revision 33ce323edd28935192fdb15010fd51eaa610aa58) +++ scripts/_errors_ (.../_errors_) (revision bcafc1d20046aaf42e7f697f9c33376e0b6b6126) @@ -56,6 +56,7 @@ ERR_CMDFAIL_WIFI_INFO_MASK=26 ERR_CMDFAIL_WIFI_INFO_GATEWAY=27 ERR_CMDFAIL_WIFI_INFO_DNS=28 +ERR_CMDFAIL_WIFI_DELETE_CONNECTION=29 # 3x: Brightness Errors 3x ERR_CMDFAIL_BRIGHTNESS_GET=30 @@ -66,3 +67,25 @@ ERR_CMDFAIL_USB_DEVICE_MOUNT=41 ERR_CMDFAIL_USB_FOLDER_REMOVE=42 ERR_CMDFAIL_USB_DEVICE_UMOUNT=43 + +# 5x: Date Time Errors +ERR_CMDFAIL_TIMEDATECTL=50 +ERR_CMDFAIL_DATE=51 +ERR_CMDFAIL_HW_CLOCK=52 + +# 6x: Factory Reset Errors +ERR_CMDFAIL_SETTINGS_CONF_PASSNOTFOUND=60 +ERR_CMDFAIL_SETTINGS_CONF_OVERWRITE=61 +ERR_CMDFAIL_SETTINGS_DFLT_RESTORE=62 +ERR_CMDFAIL_SETTINGS_REMOVE_LOG_FOLDER=63 +ERR_CMDFAIL_SETTINGS_REMOVE_SERVICE_LOG=64 +ERR_CMDFAIL_SETTINGS_COUNT_SERVICE_LOG=65 +ERR_CMDFAIL_SETTINGS_REMOVE_TX_LOG=66 +ERR_CMDFAIL_SETTINGS_COUNT_TX_LOG=67 + +# 7x: SSH Access Errors +ERR_CMDFAIL_SSH_SERVICE=70 +ERR_CMDFAIL_SSH_STATUS=71 +ERR_CMDFAIL_ROOT_LOGIN_ENABLED=72 +ERR_CMDFAIL_ROOT_LOGIN_DISABLED=73 +ERR_CMDFAIL_ROOT_LOGIN_STATUS=74 Index: scripts/wifi_scan.sh =================================================================== diff -u -r33ce323edd28935192fdb15010fd51eaa610aa58 -rbcafc1d20046aaf42e7f697f9c33376e0b6b6126 --- scripts/wifi_scan.sh (.../wifi_scan.sh) (revision 33ce323edd28935192fdb15010fd51eaa610aa58) +++ scripts/wifi_scan.sh (.../wifi_scan.sh) (revision bcafc1d20046aaf42e7f697f9c33376e0b6b6126) @@ -19,7 +19,7 @@ # Lists all the available Access Points using nmcli command of NetworkManager module. # Parameters # No Parameters -# Retruns +# Returns # Sorted list of the APs in defined column as specified in the $COLUMNS. # Considerations # For the Application to work with the List of APs