Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -r1d4e0030bba189d1a92418d73e333aa0a36ccc5b -r30f5cdf4ba3babf95f708d46b28c445fa4b0bebf --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 1d4e0030bba189d1a92418d73e333aa0a36ccc5b) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 30f5cdf4ba3babf95f708d46b28c445fa4b0bebf) @@ -783,22 +783,22 @@ ListTitle = DD firmware image integrity POST test failed [196] -Title = Arterial Pressurization Timeout -ListTitle = TD failed to reach arterial pressure target +Title = Blood Pump Stopped: Saline Bag Empty +ListTitle = TD Saline bag is empty [197] -Title = Arterial Pressure Decay -ListTitle = TD detected excessive arterial pressure decay +Title = Service Required: Dialysate Device +ListTitle = DD substitution pump RPM out of range [198] -Title = Arterial Pressure Leak -ListTitle = TD detected arterial pressure leak +Title = Prime Complete - 30 MIN WARNING +ListTitle = TD prime completed medium priority alarm [199] -Title = Arterial Pressure Relief Failure -ListTitle = TD failed arterial pressure relief check +Title = Blood Pump Stopped: Venous Air Detector +ListTitle = Repeated air trap lower events detected during treatment [200] -Title = Air Pump Stall Detected -ListTitle = Air pump stall detected during treatment +Title = Wrong Tube Set Installed +ListTitle = TD installed tube set type does not match the selected treatment modality