Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -r6f6fc4adb8a488e2ee5ba8279ac8a175f8727fe2 -r5d6be308e1cfe0bd6491c6c8c73c6ce761d9cfcb --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 6f6fc4adb8a488e2ee5ba8279ac8a175f8727fe2) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 5d6be308e1cfe0bd6491c6c8c73c6ce761d9cfcb) @@ -189,6 +189,10 @@ [47] Title = Low Venous Pressure ListTitle = Venous pressure too low during treatment +Ensure venous access lines are secured. +Check needle placement and fix access issues, if needed. +Ensure transducers are secured and no moisture is present. Replace if needed. +Inspect dialyzer for clots. Follow facility policy to end treatment if clots are present. [48] Title = High Venous Pressure