Index: configurations/Parameters/DataList.conf =================================================================== diff -u -rc855a83e140ebebbb1aa4c45b3cf3cc0b7e675ea -r128331b80874f8d828dbf57a15415ce5520eb704 --- configurations/Parameters/DataList.conf (.../DataList.conf) (revision c855a83e140ebebbb1aa4c45b3cf3cc0b7e675ea) +++ configurations/Parameters/DataList.conf (.../DataList.conf) (revision 128331b80874f8d828dbf57a15415ce5520eb704) @@ -56,9 +56,9 @@ Dialysate_Temperature_Res = +0.5 # LEAHI-SRS-303 [Acid Concentrate Options] # U32 -1.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 -2.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 -3.0 K, 2.50 Ca, 1 Mg # LEAHI-SRS-288 +1.0 K, 2.50 Ca # LEAHI-SRS-288 +2.0 K, 2.50 Ca # LEAHI-SRS-288 +3.0 K, 2.50 Ca # LEAHI-SRS-288 [Acid Concentrate Ranges] # K Acid_Concentrate_Potassium_Def = +2.0 # LEAHI-PRS-660, LEAHI-PRS-241 @@ -114,16 +114,16 @@ 30 # LEAHI-PRS-131 60 # LEAHI-PRS-131 -[Trancememberance Pressure Ranges] -# Adjustment, Trancememberance Window -Trancememberance_Adjust_Window_Def = +40 # LEAHI-PRS-215 ? -Trancememberance_Adjust_Window_Min = +40 # LEAHI-PRS-255 ? -Trancememberance_Adjust_Window_Max = +100 # LEAHI-PRS-255 ? -Trancememberance_Adjust_Window_Res = +20 # LEAHI-PRS-255 ? +[Transmembrane Pressure Ranges] +# Adjustment, Arterial Window +Transmembrane_Adjust_Window_Def = +40 # LEAHI-PRS-215 ? +Transmembrane_Adjust_Window_Min = +40 # LEAHI-PRS-255 ? +Transmembrane_Adjust_Window_Max = +100 # LEAHI-PRS-255 ? +Transmembrane_Adjust_Window_Res = +20 # LEAHI-PRS-255 ? #Monitoring -Trancememberance_Monitor_Res = 1 # LEAHI-PRS-269 LEAHI-SRS-222 -Trancememberance_Monitor_Min = -650 # LEAHI-PRS-269 LEAHI-SRS-222 -Trancememberance_Monitor_Max = +250 # LEAHI-PRS-269 LEAHI-SRS-222 +Transmembrane_Monitor_Res = 1 # LEAHI-PRS-269 LEAHI-SRS-222 +Transmembrane_Monitor_Min = -650 # LEAHI-PRS-269 LEAHI-SRS-222 +Transmembrane_Monitor_Max = +250 # LEAHI-PRS-269 LEAHI-SRS-222 [Arterial Blood Pressure Ranges] # Adjustment, Arterial Window @@ -167,7 +167,7 @@ [Ultrafiltration Volume Ranges] # L Ultrafiltration_Volume_Def = +0.0 # LEAHI-PRS-127, LEAHI-PRS-238 Ultrafiltration_Volume_Min = +0.0 # LEAHI-PRS-127, LEAHI-PRS-238 -Ultrafiltration_Volume_Max = +0.8 # LEAHI-PRS-127, LEAHI-PRS-238 +Ultrafiltration_Volume_Max = +8.0 # LEAHI-PRS-127, LEAHI-PRS-238 Ultrafiltration_Volume_Res = +0.1 # LEAHI-PRS-127 [Substitution Volume Ranges] # L @@ -221,3 +221,9 @@ [Water Sample Test Result Required] Disabled Enabled + +[Isolated UF Duration Ranges] +Isolated_UF_Duration_Def = +30 # LEAHI-PRS-125 +Isolated_UF_Duration_Min = +30 # LEAHI-PRS-125 +Isolated_UF_Duration_Max = +480 # LEAHI-PRS-125 +Isolated_UF_Duration_Res = +15 # LEAHI-PRS-125