Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -r351632c8f5849be7ff648fc9ec3360049963a21c -r29df90b720dfc6f3aea1e24efd9943c82203fdb5 --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 351632c8f5849be7ff648fc9ec3360049963a21c) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 29df90b720dfc6f3aea1e24efd9943c82203fdb5) @@ -489,14 +489,14 @@ Message = DD Conductivity Sensor Error [98] -Title = -ListTitle = -Message = +Title = Approaching MAX Blood Pump Stop Time +ListTitle = Blood sitting too long warning (> 4 min) +Message = Blood Stopped: Warning [99] -Title = -ListTitle = -Message = +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood sitting too long after treatment stopped by user alarm (> 5 min) +Message = Blood Stop: Blood Clot Risk [100] Title = Service Required: Hemodialysis Device @@ -763,3 +763,8 @@ ListTitle = FP volume difference between integrated flow and expected volume Message = FP Filters: Volume difference too high +[153] +Title = FP: Permeate Overfill +ListTitle = FP tank level is too high +Message = FP Tank: Tank is too full +