Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -r6d6249aaa4fcf4409b18c87d1a3f461f472e2c9e -r351632c8f5849be7ff648fc9ec3360049963a21c --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 6d6249aaa4fcf4409b18c87d1a3f461f472e2c9e) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 351632c8f5849be7ff648fc9ec3360049963a21c) @@ -658,3 +658,108 @@ ListTitle = Insufficient battery charge to support 20 min without A/C power Message = TD Alarm (Low): Battery Low +[132] +Title = High Dialysate Temperature +ListTitle = Dialysate temperature is out of high safety range +Message = Dial. Gen: Dialysate Temp High + +[133] +Title = High Dialysate Temperature +ListTitle = Dialysate temperature is above the target temperature +Message = Dial. Gen: Dialysate Temp High + +[134] +Title = Low Dialysate Temperature +ListTitle = Dialysate temperature is below the target temperature +Message = Dial. Gen: Dialysate Temp Low + +[135] +Title = Service Required: FP Device +ListTitle = FP pressure relief is too low +Message = FP Fault: Pressure Relief Valve Low + +[136] +Title = Service Required: FP Device +ListTitle = FP pressure relief is too High +Message = FP Fault: Pressure Relief Valve High + +[137] +Title = Service Required: FP Device +ListTitle = FP device is using too much power +Message = FP Fault: Power usage has exceeded budget + +[138] +Title = Service Required: FP Device +ListTitle = FP RO pump flow out of range +Message = FP Fault: RO Pump Flow Low + +[139] +Title = Low Inlet Water Temperature +ListTitle = Inlet Water temperature is too low +Message = FP Water Gen: Inlet Water Temp Low + +[140] +Title = High Inlet Water Temperature +ListTitle = Inlet Water temperature is too high +Message = FP Water Gen: Inlet Water Temp High + +[141] +Title = Low Inlet Water Pressure +ListTitle = Inlet Water pressure is too low +Message = FP Water Gen: Inlet Pressure Low + +[142] +Title = High Inlet Water Pressure +ListTitle = Inlet Water pressure is too high +Message = FP Water Gen: Inlet Pressure High + +[143] +Title = High Inlet Water Conductivity +ListTitle = Inlet Water conductivity is too high +Message = FP Water Gen: Inlet Conductivity High + +[144] +Title = Low Inlet Water Conductivity +ListTitle = Inlet Water conductivity is too low +Message = FP Water Gen: Inlet Conductivity Low + +[145] +Title = Service Required: FP Device +ListTitle = FP RO pump pressure out of range +Message = FP Fault: RO Pump High Pressure + +[146] +Title = Service Required: FP Device +ListTitle = FP outlet primary conductivity sensor out of range +Message = FP Fault: RO Permeate Cond High + +[147] +Title = Service Required: FP Device +ListTitle = FP inlet primary conductivity sensor out of range +Message = FP Fault: RO Permeate Cond Low + +[148] +Title = Service Required: FP Device +ListTitle = FP RO rejection ratio out of range +Message = FP Fault: RO Rejection + +[149] +Title = Service Required: FP Device +ListTitle = FP pressure regulator is out of bounds +Message = FP Fault: Pressure regulator not in range + +[150] +Title = FP: Filter Health +ListTitle = FP pressure difference between the filters is too high +Message = FP Filters: Pressure difference too high + +[151] +Title = FP: Filter Health +ListTitle = FP conductivity after flushing the system is too high +Message = FP Filters: Permeate conductivity high + +[152] +Title = FP: Filter Health +ListTitle = FP volume difference between integrated flow and expected volume +Message = FP Filters: Volume difference too high +