Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -rf45e77d1da8294f866eeb7674c9ee8cbaed97bbc -r7ad71671889230f3e57736f2c057b8d69a6d353c --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision f45e77d1da8294f866eeb7674c9ee8cbaed97bbc) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 7ad71671889230f3e57736f2c057b8d69a6d353c) @@ -1,659 +1,659 @@ -[ALARM_ID_NO_ALARM] +[0] Title = Indication for when no alarms is a possible situation ListTitle = Message = -[ALARM_ID_TD_SOFTWARE_FAULT] +[1] Title = Service Required: Hemodialysis Device ListTitle = TD Software fault. Software found itself in an unexpected state Message = TD Fault: SW Invalid State -[ALARM_ID_DD_SOFTWARE_FAULT] +[2] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state Message = DD Fault: Software Invalid State -[ALARM_ID_RO_SOFTWARE_FAULT] +[3] Title = Service Required: Dialysate Device ListTitle = DD software fault. Software found itself in an unexpected state Message = DD Fault: Software Invalid State -[ALARM_ID_TD_FPGA_POST_TEST_FAILED] +[4] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA POST failure Message = TD POST: FPGA Compatibility -[ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED] +[5] Title = Service Required: Hemodialysis Device ListTitle = TD Watchdog POST failure Message = TD POST: WD Safestate -[ALARM_ID_TD_UI_COMM_POST_FAILED] +[6] Title = Service Required: Hemodialysis Device ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up Message = TD POST: Proc- UI Comm -[ALARM_ID_TD_FPGA_CLOCK_SPEED_CHECK_FAILURE] +[7] Title = Service Required: Hemodialysis Device ListTitle = TD processor clock speed checks against FPGA clock failure Message = TD Fault: Proc- FPGA Clock -[ALARM_ID_TD_FPGA_COMM_TIMEOUT] +[8] Title = Service Required: Hemodialysis Device ListTitle = TD FPGA communication down for too long Message = TD Fault: FPGA Com -[ALARM_ID_TD_UI_COMM_TIMEOUT] +[9] Title = Service Required: Hemodialysis Device ListTitle = User interface communication timeout Message = TD Fault: UI Comm -[ALARM_ID_TD_DD_COMM_TIMEOUT] +[10] Title = Communication Error ListTitle = Dialysate generator communication timeout Message = Loss of DD Comm (Timeout) -[ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS] +[11] Title = Service Required: Hemodialysis Device ListTitle = Too many bad CRCs detected on received system messages Message = TD Fault: CAN CRC Com -[ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_UI] +[12] Title = Service Required: Hemodialysis Device ListTitle = System message that required acknowledgment was not acknowledged Message = TD Fault: CAN ACK UI Com -[ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD] +[13] Title = Communication Error ListTitle = TD to DD comm error. Message = Loss of DD Comm (DD ACK) -[ALARM_ID_TD_CRITICAL_DATA_ERROR] +[14] Title = Service Required: Hemodialysis Device ListTitle = TD critical data integrity check failed Message = TD Fault: Critical Data Err -[ALARM_ID_DD_CRITICAL_DATA_ERROR] +[15] Title = Service Required: Dialysate Device ListTitle = DD critical data integrity check failed Message = DD Fault: Critical Data Err -[ALARM_ID_TD_WATCHDOG_EXPIRED] +[16] Title = Service Required: Hemodialysis Device ListTitle = TD watchdog expired Message = TD Fault: WD Error -[ALARM_ID_DD_WATCHDOG_EXPIRED] +[17] Title = Service Required: Dialysate Device ListTitle = DD watchdog expired Message = DD Fault: WD Error -[ALARM_ID_DD_COMM_TOO_MANY_BAD_CRCS] +[18] Title = Service Required: Dialysate Device ListTitle = Too many bad CRCs detected on received system messages Message = DD Fault: CAN CRC Com -[ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_RO] +[19] Title = Service Required: Dialysate Device ListTitle = DD to RO comm error Message = Loss of RO Comm (RO ACK) -[ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_TD] +[20] Title = Service Required: Dialysate Device ListTitle = DD to TD comm error Message = Loss of TD Comm (TD ACK) -[ALARM_ID_DD_FPGA_POST_TEST_FAILED] +[21] Title = Service Required: Dialysate Device ListTitle = DD FPGA POST failure Message = DD POST: FPGA Compatibility -[ALARM_ID_DD_FPGA_CLOCK_SPEED_CHECK_FAILURE] +[22] Title = Service Required: Dialysate Device ListTitle = FPGA clock speed failure Message = DD Fault: Proc- FPGA Clock -[ALARM_ID_DD_FPGA_COMM_TIMEOUT] +[23] Title = Service Required: Dialysate Device ListTitle = DD FPGA communication down for too long Message = DD Fault: FPGA Com -[ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE] +[24] Title = Service Required: Hemodialysis Device ListTitle = TD alarm audio failed POST Message = TD POST: Primary Alarm Current -[ALARM_ID_DD_VALVE_CONTROL_FAILURE] +[25] Title = Service Required: Dialysate Device ListTitle = DD FPGA not accepting commanded valve states Message = DD Fault: Valve Control -[ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT] +[26] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor reporting error or not being read successfully Message = TD Arterial Pressure Sensor Error -[ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT] +[27] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor reporting error or not being read successfully Message = TD Venous Pressure Sensor Error -[ALARM_ID_DD_HYD_OUTLET_PRES_TIMEOUT_FAULT] +[28] Title = Service Required: Dialysate Device ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully Message = DD Hyd outlet Pressure Sensor Error -[ALARM_ID_DD_BIBAG_PRES_TIMEOUT_FAULT] +[29] Title = Service Required: Dialysate Device ListTitle = DD bibag pressure sensor reporting error or not being read successfully Message = DD Bibag Pressure Sensor Error -[ALARM_ID_DD_SPENT_DIALYSATE_PRES_TIMEOUT_FAULT] +[30] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pressure sensor reporting error or not being read successfully Message = DD Spent Dial Pressure Sensor Error -[ALARM_ID_DD_FRESH_DIALYSATE_PRES_TIMEOUT_FAULT] +[31] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pressure sensor reporting error or not being read successfully Message = DD Fresh Dial Pressure Sensor Error -[ALARM_ID_DD_TRANSMEMB_PRES_TIMEOUT_FAULT] +[32] Title = Service Required: Dialysate Device ListTitle = DD transmembrane sensor reporting error or not being read successfully Message = DD Transmemb Pressure Sensor Error -[ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE] +[33] Title = Service Required: Hemodialysis Device ListTitle = TD monitored voltage is out of range Message = TD Fault: Voltage Monitor -[ALARM_ID_TD_AC_POWER_LOST] +[34] Title = Power Loss ListTitle = TD AC power lost alarm Message = Loss of AC Power -[ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT] +[35] Title = Power Loss ListTitle = TD AC power lost alarm Message = Blood Stop: Loss of AC Power -[ALARM_ID_DD_D17_COND_SENSOR_FPGA_FAULT] +[36] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully Message = DD Conductivity Sensor Error -[ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT] +[37] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully Message = DD Conductivity Sensor Error -[ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT] +[38] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully Message = DD Conductivity Sensor Error -[ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT] +[39] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully Message = DD Conductivity Sensor Error -[ALARM_ID_TD_VENOUS_BUBBLE_DETECTED] +[40] Title = Venous Air Detected ListTitle = TD venous air bubble detected alarm Message = Blood Stop: Venous Air Bubble -[ALARM_ID_TD_CARTRIDGE_DOOR_OPENED] +[41] Title = Blood Pump Stopped: Door Open ListTitle = TD cartridge door opened alarm Message = Blood Stop: Door Open -[ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED] +[42] Title = Stuck Button ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up Message = TD POST: Stuck Button -[ALARM_ID_TD_ARTERIAL_PRESSURE_LOW] +[43] Title = Low Arterial Pressure ListTitle = Arterial pressure too low during treatment Message = Blood Stop: Arterial Pres. Low -[ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH] +[44] Title = High Arterial Pressure ListTitle = Arterial pressure too high during treatment Message = Blood Stop: Arterial Pres. High -[ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE] +[45] Title = Service Required: Hemodialysis Device ListTitle = TD arterial pressure sensor is reading out of range Message = TD Fault: Arterial Press Range -[ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE] +[46] Title = Self Test: Cartridge ListTitle = TD arterial pressure self-test failure alarm Message = Process: Pre- Tx Remove Cartridge -[ALARM_ID_TD_VENOUS_PRESSURE_LOW] +[47] Title = Low Venous Pressure ListTitle = Venous pressure too low during treatment Message = Blood Stop: Venous Pres. Low -[ALARM_ID_TD_VENOUS_PRESSURE_HIGH] +[48] Title = High Venous Pressure ListTitle = Venous pressure too high during treatment Message = Blood Stop: Venous Pres. High -[ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE] +[49] Title = Service Required: Hemodialysis Device ListTitle = TD venous pressure sensor is reading out of range Message = TD Fault: Venous Press Range -[ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE] +[50] Title = Self Test: Cartridge ListTitle = TD venous pressure self-test failure alarm Message = Process: Pre- Tx Remove Cartridge -[ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE] +[51] Title = Self Test: Cartridge Check ListTitle = TD pre-treatment mode dry self-tests failure Message = Process: Pre- Tx Dry Self- Test -[ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR] +[52] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D11_PUMP speed control error Message = DD Fault: Pump Speed Fault -[ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR] +[53] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump D10_PUMP speed control error Message = DD Fault: Pump Speed Fault -[ALARM_ID_DD_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE] +[54] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump hall sensor out of range. Message = DD Fault: Pump Speed Fault -[ALARM_ID_DD_CONCENTRATE_PUMP_FAULT] +[55] Title = Service Required: Dialysate Device ListTitle = DD concentrate pump fault Message = DD Fault: Pump Fault -[ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE] +[56] Title = Service Required: Dialysate Device ListTitle = DD temperature sensor out of range Message = DD Fault: Temp Sensor Range -[ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT] +[57] Title = Service Required: Dialysate Device ListTitle = DD temperature sensors not being read successfully Message = DD Fault: Temperature RTD Comm -[ALARM_ID_DD_BARO_SENSOR_FPGA_FAULT] +[58] Title = Service Required: Dialysate Device ListTitle = DD barometric sensor not being read successfully Message = DD Fault: Baro temp Comm -[ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC] +[59] Title = Service Required: Dialysate Device ListTitle = DD barometric pressure sensor bad coefficients CRC Message = DD Fault: Baro temp Cal CRC -[ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE] +[60] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump RPM out of range Message = DD Fault: Fresh Dial Pump Speed Fault -[ALARM_ID_DD_D12_PUMP_OFF_FAULT] +[61] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump off fault Message = DD Fault: Fresh Dial Pump Speed Not Off -[ALARM_ID_DD_D12_PUMP_CURRENT_OUT_OF_RANGE] +[62] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump current out of range Message = DD Fault: Fresh Dial Pump Current -[ALARM_ID_DD_D12_PUMP_DIRECTION_INVALID] +[63] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump invalid direction Message = DD Fault: Fresh Dial Pump Dir Fault -[ALARM_ID_DD_D12_PUMP_DIRECTION_FPGA_FAULT] +[64] Title = Service Required: Dialysate Device ListTitle = DD fresh dialysate pump direction FPGA fault Message = DD Fault: Fresh Dial pump Dir Err Cnt -[ALARM_ID_DD_D48_PUMP_RPM_OUT_OF_RANGE] +[65] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump RPM out of range Message = DD Fault: Spent Dial Pump Speed Fault -[ALARM_ID_DD_D48_PUMP_OFF_FAULT] +[66] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump off fault Message = DD Fault: Spent Dial Pump Speed Not Off -[ALARM_ID_DD_D48_PUMP_CURRENT_OUT_OF_RANGE] +[67] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump current out of range Message = DD Fault: Spent Dial Pump Current -[ALARM_ID_DD_D48_PUMP_DIRECTION_INVALID] +[68] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump invalid direction Message = DD Fault: Spent Dial Pump Dir Fault -[ALARM_ID_DD_D48_PUMP_DIRECTION_FPGA_FAULT] +[69] Title = Service Required: Dialysate Device ListTitle = DD spent dialysate pump direction FPGA fault Message = DD Fault: Spent Dial pump Dir Err Cnt -[ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON] +[70] Title = Service Required: Dialysate Device ListTitle = DD D5 heater is On with low fluid level Message = DD Fault: Heater On With No Fluid -[ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON] +[71] Title = Service Required: Dialysate Device ListTitle = DD D45 heater is On with low fluid level Message = DD Fault: Heater On With No Fluid -[ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT] +[72] Title = Service Required: Dialysate Device ListTitle = DD water inlet input pressure sensor reporting error or not being read successfully Message = DD Water Inlet Input Press Sensor Error -[ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT] +[73] Title = Service Required: Dialysate Device ListTitle = DD water inlet output pressure sensor reporting error or not being read successfully Message = DD Water Inlet Output Press Sensor Error -[ALARM_ID_DD_INLET_WATER_PRESSURE_IN_LOW_RANGE] +[74] Title = Incoming Water Pressure Input Low ListTitle = DD Inlet water pressure input too low Message = DD: Inlet Water Pres Input Low -[ALARM_ID_DD_INLET_WATER_PRESSURE_IN_HIGH_RANGE] +[75] Title = Incoming Water Pressure Input High ListTitle = DD inlet water pressure input is too high Message = DD: Inlet Water Pres Input High -[ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_LOW_RANGE] +[76] Title = Incoming Water Pressure Output Low ListTitle = DD Inlet water pressure output too low Message = DD: Inlet Water Pres Output Low -[ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_HIGH_RANGE] +[77] Title = Incoming Water Pressure Output High ListTitle = DD inlet water pressure output is too high Message = DD: Inlet Water Pres Output High -[ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT] +[78] Title = Blood Pump Stopped: Venous Air Detector ListTitle = Air trap fill timeout during treatment Message = Blood Stop: Air Trap Fill -[ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS] +[79] Title = Service Required: Hemodialysis Device ListTitle = Air trap level sensors reporting illegal combination of air/fluid Message = TD Fault: Air Trap Sensor -[ALARM_ID_TD_PINCH_VALVE_FAULT] +[80] Title = Service Required: Hemodialysis Device ListTitle = Pinch valve reporting a fault Message = TD Fault: Pinch Valve -[ALARM_ID_TD_VALVE_HOMING_FAILED] +[81] Title = Service Required: Hemodialysis Device ListTitle = TD valve homing failed Message = TD Fault: Valve Home -[ALARM_ID_TD_VALVE_TRANSITION_TIMEOUT] +[82] Title = Service Required: Hemodialysis Device ListTitle = TD valve transition time out Message = TD Fault: Valve Transition Err -[ALARM_ID_RO_CAN_MESSAGE_NOT_ACKED_BY_DD] +[83] Title = Communication Error ListTitle = DD to RO comm error. Message = RO Loss of DD Comm (DD ACK) -[ALARM_ID_RO_FPGA_COMM_TIMEOUT] +[84] Title = Service Required: Dialysate Device ListTitle = RO FPGA communication down for too long Message = RO Fault: FPGA Comm -[ALARM_ID_RO_COMM_TOO_MANY_BAD_CRCS] +[85] Title = Service Required: Hemodialysis Device ListTitle = RO Too many bad CRCs detected on received system messages Message = RO Fault: CAN CRC Comm -[ALARM_ID_RO_VALVE_CONTROL_FAILURE] +[86] Title = Service Required: Filtration Device ListTitle = RO FPGA not accepting commanded valve states Message = RO Fault: Valve Control -[ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE] +[87] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill start pressure is not in range Message = DD Fault: BC Pressure not in range -[ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE] +[88] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill end pressure is not in range Message = DD Fault: BC Pressure not in range -[ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE] +[89] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill start pressure is not in range Message = DD Fault: BC Pressure not in range -[ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE] +[90] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill end pressure is not in range Message = DD Fault: BC Pressure not in range -[ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE] +[91] Title = Service Required: Dialysate Device ListTitle = DD hydraulics chamber degassing pressure is not in range Message = DD Fault: Hyd Cham Neg Pres not in range -[ALARM_ID_AVAILABLE_92] +[92] Title = ListTitle = Message = -[ALARM_ID_AVAILABLE_93] +[93] Title = ListTitle = Message = -[ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE] +[94] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range Message = DD Fault: BC Pressure not in range -[ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE] +[95] Title = Service Required: Dialysate Device ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range Message = DD Fault: BC Pressure not in range -[ALARM_ID_TD_TREATMENT_STOPPED_BY_USER] +[96] Title = Blood Pump Stopped: Treatment Paused ListTitle = TD treatment stopped by user action - pressed stop button Message = Blood Stopped: Treatment Pause -[ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT] +[97] Title = Service Required: Dialysate Device ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully Message = DD Conductivity Sensor Error -[ALARM_ID_AVAILABLE_98] +[98] Title = ListTitle = Message = -[ALARM_ID_AVAILABLE_99] +[99] Title = ListTitle = Message = -[ALARM_ID_TD_COMM_TIMEOUT] +[100] Title = Service Required: Hemodialysis Device ListTitle = TD communication timeout Message = TD Fault: UI- Proc Comm -[ALARM_ID_UI_POST_FAILURE_OS_VERSION] +[101] Title = Service Required: Hemodialysis Device ListTitle = UI POST OS Version compatibility failure Message = TD POST: OS Version -[ALARM_ID_UI_POST_FAILURE_SHASUM] +[102] Title = Service Required: Hemodialysis Device ListTitle = UI POST Application Integrity (Sha256Sum) failure Message = TD POST: UI Image -[ALARM_ID_UI_POST_FAILURE_CANBUS] +[103] Title = Service Required: Hemodialysis Device ListTitle = UI POST CANBus failure Message = TD POST: UI CAN Driver -[ALARM_ID_UI_POST_FAILURE_DISPLAY] +[104] Title = Service Required: Hemodialysis Device ListTitle = If the UI POST reports display failure. Message = UI POST Display failure -[ALARM_ID_UI_POST_FAILURE_TOUCH] +[105] Title = Service Required: Hemodialysis Device ListTitle = UI POST Touch failure Message = TD POST: UI Touch Driver -[ALARM_ID_UI_POST_FAILURE_SDCARD] +[106] Title = Service Required: Hemodialysis Device ListTitle = UI POST SD-Card failure Message = TD POST: UI SD- Card -[ALARM_ID_UI_POST_FAILURE_RTC] +[107] Title = Service Required: Hemodialysis Device ListTitle = UI POST RTC failure Message = TD POST: UI RTC -[ALARM_ID_UI_POST_FAILURE_WIFI] +[108] Title = Service Required: Hemodialysis Device ListTitle = UI POST WiFi failure Message = TD POST: UI WiFi -[ALARM_ID_UI_POST_FAILURE_BLUETOOTH] +[109] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bluetooth failure Message = TD POST: UI Bluetooth -[ALARM_ID_UI_POST_FAILURE_ETHERNET] +[110] Title = Service Required: Hemodialysis Device ListTitle = UI POST Ethernet failure Message = TD POST: UI Ethernet -[ALARM_ID_UI_POST_FAILURE_SOUND] +[111] Title = Service Required: Hemodialysis Device ListTitle = UI POST Sound failure Message = TD POST: UI Sound -[ALARM_ID_UI_POST_FAILURE_CLOUDSYNC] +[112] Title = Service Required: Hemodialysis Device ListTitle = UI POST cloud sync POST failure Message = TD POST: UI CloudSync -[ALARM_ID_UI_POST_FAILURE_INVALID_YEAR] +[113] Title = Service Required: Hemodialysis Device ListTitle = UI POST Invalid Year Message = TD POST Alarm: UI Invalid Year -[ALARM_ID_UI_POST_FAILURE_SETTINGS_BAD] +[114] Title = Service Required: Hemodialysis Device ListTitle = UI POST Bad Settings Message = TD POST Alarm: UI Bad Settings -[ALARM_ID_UI_EVNT_FAILURE_SDCARD] +[115] Title = Service Required: Hemodialysis Device ListTitle = SD card is non-operational Message = TD Fault: UI- SD- Card Fault -[ALARM_ID_UI_RESERVED_116] +[116] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_117] +[117] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_118] +[118] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_119] +[119] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_120] +[120] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_121] +[121] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_122] +[122] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_123] +[123] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_124] +[124] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_125] +[125] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_126] +[126] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_127] +[127] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_128] +[128] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_129] +[129] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_UI_RESERVED_130] +[130] Title = Service Required: Hemodialysis Device ListTitle = RESERVED Message = TD Fault: UI- RESERVED -[ALARM_ID_BATTERY_LOW] +[131] Title = Battery Capacity Low ListTitle = Insufficient battery charge to support 20 min without A/C power Message = TD Alarm (Low): Battery Low