Index: configurations/Parameters/DataList.conf =================================================================== diff -u -r1deb0e75df62918fea0efe74027b9f7a775dd562 -r9281a3f35bc0a4260d881353b3648698f74dbb5d --- configurations/Parameters/DataList.conf (.../DataList.conf) (revision 1deb0e75df62918fea0efe74027b9f7a775dd562) +++ configurations/Parameters/DataList.conf (.../DataList.conf) (revision 9281a3f35bc0a4260d881353b3648698f74dbb5d) @@ -78,6 +78,15 @@ [Heparin Type Options] # U32 Unfractionated 1,000 IU/mL # LEAHI-SRS-291 +[Vitals Measurement Interval Options] # min (0 = OFF) +0 # LEAHI-PRS-131 +5 # LEAHI-PRS-131 +10 # LEAHI-PRS-131 +15 # LEAHI-PRS-131 +20 # LEAHI-PRS-131 +30 # LEAHI-PRS-131 +60 # LEAHI-PRS-131 + [Trancememberance Pressure Ranges] # Adjustment, Arterial Window Trancememberance_Adjust_Window_Def = +40 # LEAHI-PRS-215 ? @@ -116,12 +125,6 @@ Venous_Monitor_Min = -100 # LEAHI-PRS-253 LEAHI-SRS-220 Venous_Monitor_Max = +500 # LEAHI-PRS-253 LEAHI-SRS-220 -[Blood Pressure Measure Interval Ranges] # min (0 = OFF) -Blood_Pressure_Measure_Interval_Def = +0 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Min = +5 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Max = +60 # LEAHI-SRS-304 -Blood_Pressure_Measure_Interval_Res = +5 # LEAHI-SRS-304 - [Rinseback Volume Ranges] # mL/min Rinseback_Volume_Def = 300 # LEAHI-PRS-559 Rinseback_Volume_Min = 0 # LEAHI-?