Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -r3062967d4d109258d9ff6d26833e687eb6354d4f -rc913b39e469a884672f4a500c851af99a7035fba --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 3062967d4d109258d9ff6d26833e687eb6354d4f) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision c913b39e469a884672f4a500c851af99a7035fba) @@ -849,4 +849,15 @@ Machine is in bypass and treatment will automatically resume once conductivity is within range. Review pH and conductivity based on facility policies once treatment is resumed. +[251] +Title = Dialysis Treatment Complete +ListTitle = Dialysis Treatment Complete +Press ‘OK’ to continue to end of treatment options +[252] +Title = End of Treatment Timeout +ListTitle = End of Treatment Timeout +The maximum time allowed on this screen has been reached. +Press 'Return Blood & End Treatment' to begin the blood return process + + Index: configurations/Instructions/Instructions.conf =================================================================== diff -u -rf9ec72965da0a260f4b63454e06633d7539b17ce -rc913b39e469a884672f4a500c851af99a7035fba --- configurations/Instructions/Instructions.conf (.../Instructions.conf) (revision f9ec72965da0a260f4b63454e06633d7539b17ce) +++ configurations/Instructions/Instructions.conf (.../Instructions.conf) (revision c913b39e469a884672f4a500c851af99a7035fba) @@ -25,114 +25,123 @@ # ###################################################################################################### [Concentrates] -Title = Connect Acid and Bicarbonate Concentrates -Remove acid connector and connect to acid concentrate. = Concentrates_1.png -Connect dry bicarbonate. = Concentrates_2.png +Title = Connect Acid Concentrate and Dry Bicarbonate Cartridge +Disconnect <$B53333>red acid straw and connect to acid concentrate. = Concentrates_1.png +Pull the <$#2268FF>blue bicarbonate knob upwards and insert the dry bicarbonate cartridge. = Concentrates_2.png [Installation^Blood Tubing Set] Title = Install Blood Tubing Set Open the blood pump door. = Installation_BloodTube_1.png -Insert blood tubing set and ensure it is secure. = Installation_BloodTube_2.png -Push the blood tube into the blood pump. = Installation_BloodTube_3.png -Close the blood pump door. = Installation_BloodTube_4.png +Insert blood tubing bracket into the clip. Push inwards and ensure it is secure. = Installation_BloodTube_2.png +Close the blood pump door. = Installation_BloodTube_3.png +Press 'Auto-Load' button. = Installation_BloodTube_4.png [Installation^Bloodlines] -Title = Install Bloodlines -Insert saline line into the first slot. = Installation_Bloodlines_1.png -Insert <$B53333>arterial line into the second slot. = Installation_Bloodlines_2.png -Ensure the <$2268FF>venous line is placed in the bubble detector. = Installation_Bloodlines_3.png -Insert the <$2268FF>venous line into the third slot. = Installation_Bloodlines_4.png -Connect the <$B53333>arterial and <$2268FF>venous bloodlines using the recirculation connector. = Installation_Bloodlines_5.png +Title = Insert Bloodlines +Insert saline line into the leftmost slot. = Installation_Bloodlines_1.png +Insert <$B53333>arterial line into the middle slot. = Installation_Bloodlines_2.png +Ensure <$2268FF>venous line is placed in the air detector. = Installation_Bloodlines_3.png +Insert <$2268FF>venous line into the rightmost slot. = Installation_Bloodlines_4.png +Ensure <$B53333>arterial and <$2268FF>venous bloodlines are tightened via the recirculation connector and placed behind the bar. = Installation_Bloodlines_5.png [Installation^Pressure Lines and Dialyzer] -Title = Install Pressure Lines and Dialyzer -Connect the <$B53333>arterial pressure line to the <$B53333>arterial pressure transducer port. = Installation_CPLD_1.png -Connect the <$2268FF>venous pressure line to the <$2268FF>venous pressure transducer port. = Installation_CPLD_2.png -Insert dialyzer into the holder and connect the dialysate lines. = Installation_CPLD_3.png -Connect the bloodlines to the dialyzer. = Installation_CPLD_4.png -Clamp the <$2268FF>venous chamber line. = Installation_CPLD_5.png -Clamp the heparin line. = Installation_CPLD_6.png +Title = Connect Pressure Lines and Dialyzer +Connect <$B53333>arterial pressure line to <$B53333>arterial pressure transducer port. = Installation_CPLD_1.png +Connect <$2268FF>venous pressure line to <$2268FF>venous pressure transducer port. = Installation_CPLD_2.png +Connect <$2268FF>venous bloodline to the top of dialyzer and <$B53333>arterial line to the bottom. = Installation_CPLD_4.png +Connect <$2268FF>venous dialysate line to the top of dialyzer and <$B53333>arterial line to the bottom. = Installation_CPLD_4.png +Insert dialyzer into dialyzer holder. = Installation_CPLD_3.png +Clamp heparin line and <$2268FF>venous infusion line. = Installation_CPLD_6.png [Installation^Saline] Title = Connect Saline Bag +Uncap and insert saline line spike into saline bag. = Saline_3.png +Clamp saline line and Y-connector. = Saline_2.png Hang saline bag on the hook. = Saline_1.png -Clamp saline line. = Saline_2.png -Use the saline line to spike saline bag. = Saline_3.png [Installation^Substitution] Title = Connect Substitution Line +Connect substitution line to the machine. Connect substitution line to the appropriate Pre or Post dilution port. = Saline_1.png [@Installation] Title = Install Concentrates and Blood Tubing Set -Install acid and bicarbonate concentrates. = Install_Advanced_1.png +Install acid concentrate and dry bicarbonate cartridge. = Install_Advanced_1.png Install blood tubing set, dialyzer, and saline bag. = Install_Advanced_2.png +[@Installation^Substitution] +Title = Install Concentrates and Blood Tubing Set +Install acid concentrate and dry bicarbonate cartridge. = Install_Advanced_1.png +Install blood tubing set, dialyzer, and connect substitution line. + [Connect] -Title = Connect Bloodlines to the Patient +Title = Connect Patient Bloodlines +Wash hands and apply personal protective equipment, per policy. Clamp <$B53333>arterial and <$2268FF>venous bloodlines (2 clamps). = Connect_1.png -Remove recirculation connector from the bloodlines. = Connect_2.png -Connect bloodlines to patient access lines. = Connect_3.png -Unclamp <$B53333>arterial and <$2268FF>venous bloodlines (2 clamps). = Connect_4.png +Remove <$B53333>arterial and <$2268FF>venous bloodlines from recirculation connector and attach to patient access lines. = Connect_2.png +Unclamp <$B53333>arterial and <$2268FF>venous bloodlines and patient access lines (4 clamps). = Connect_4.png [@Connect] Title = Connect Bloodlines to the Patient -Connect bloodlines to patient access lines. = Connect_Advanced_1.png +Connect <$B53333>arterial and <$2268FF>venous bloodlines to patient access lines. = Connect_Advanced_1.png -[Disconnection] -Title = Remove Concentrates and Blood Tubing Set -Remove blood tubing set and dialyzer. = Install_Advanced_2.png -Remove acid and bicarbonate concentrates. = Install_Advanced_1.png [Blood Return Setup] Title = Blood Return Connection -Clamp <$B53333>arterial bloodlines and disconnect. = Return_Blood_1.png -Connect <$B53333>arterial bloodline to the saline y-connector. = Return_Blood_2.png -Unclamp <$B53333>arterial bloodline and saline y-connector. = Return_Blood_3.png +Clamp <$B53333>arterial bloodline and <$B53333>arterial access line (2 clamps). = Return_Blood_1.png +Disconnect <$B53333>arterial bloodline from <$B53333>arterial access line. +Connect <$B53333>arterial bloodline to the saline Y-connector. = Return_Blood_2.png +Unclamp <$B53333>arterial bloodline and saline Y-connector (2 clamps). = Return_Blood_3.png +Press 'Start Blood Return' button to begin blood return. [$Blood Return Setup] Title = Blood Return Connection -Clamp <$B53333>arterial bloodlines and disconnect. -Connect <$B53333>arterial bloodline to the substitution line using a sterile connector. -Unclamp <$B53333>arterial bloodline and substitution line. +Disconnect substitution line from dilution port and connect to a new, sterile recirculation connector. +Clamp <$B53333>arterial bloodline and <$B53333>arterial access line (2 clamps). +Disconnect <$B53333>arterial bloodline and connect to recirculation connector. +Unclamp <$B53333>arterial bloodline and substitution line (2 clamps). +Press 'Start Blood Return' button to begin blood return. [Recirculate Set Up] -Title = Connect Recirculation Lines -Clamp <$2268FF>venous bloodline and disconnect. -Clamp <$B53333>arterial bloodlines and disconnect from saline y-connector. -Connect <$B53333>arterial and <$2268FF>venous bloodlines using the recirculation connector. -Unclamp <$B53333>arterial and <$2268FF>venous bloodlines. +Title = Connect Lines for Temporary Break +Clamp <$2268FF>venous bloodline, <$2268FF>venous access line, and disconnect (2 clamps). +Clamp <$B53333>arterial bloodline, saline Y-connector, and disconnect (2 clamps). +Connect <$B53333>arterial and <$2268FF>venous bloodlines using a new, sterile recirculation connector. +Unclamp <$B53333>arterial and <$2268FF>venous bloodlines (2 clamps). +Press 'Start Temporary Break' button to begin the break. [$Recirculate Set Up] -Title = Connect Recirculation Lines -Clamp <$2268FF>venous bloodline and disconnect. -Clamp <$B53333>arterial and disconnect from the substitution line. -Connect substitution line to the appropriate dilution port. -Connect <$B53333>arterial and <$2268FF>venous bloodlines using the recirculation connector. -Unclamp <$B53333>arterial and <$2268FF>venous bloodlines. +Title = Connect Lines for Temporary Break +Clamp <$2268FF>venous bloodline, <$2268FF>venous access line, and disconnect (2 clamps). +Clamp <$B53333>arterial bloodline and disconnect from substitution line. +Cap the end of the substitution port. +Connect <$B53333>arterial and <$2268FF>venous bloodlines using a new, recirculation connector. +Unclamp <$B53333>arterial and <$2268FF>venous bloodlines (2 clamps). +Press 'Start Temporary Break' button to begin the break. [Reconnect To Patient] -Title = Connect Patient Bloodlines -Clamp <$B53333>arterial and <$2268FF>venous bloodlines. -Remove <$B53333>arterial and <$2268FF>venous line recirculation connector. -Connect <$B53333>arterial and <$2268FF>venous bloodlines to patient access lines. -Unclamp <$B53333>arterial and <$2268FF>venous bloodlines. +Title = Reconnect Patient Bloodlines +Clamp <$B53333>arterial and <$2268FF>venous bloodlines (2 clamps). +Remove <$B53333>arterial and <$2268FF>venous bloodlines from recirculation connector and connect to patient access lines. +Unclamp <$B53333>arterial and <$2268FF>venous bloodlines (4 clamps). [$Reconnect To Patient] Title = Connect Patient Bloodlines -Clamp <$B53333>arterial and <$2268FF>venous bloodlines. -Remove <$B53333>arterial and <$2268FF>venous line recirculation connector. -Connect <$B53333>arterial and <$2268FF>venous bloodlines to patient access lines. -Unclamp <$B53333>arterial and <$2268FF>venous bloodlines. +Clamp <$B53333>arterial and <$2268FF>venous bloodlines (2 clamps). +Remove <$B53333>arterial and <$2268FF>venous bloodlines from recirculation connector and connect to patient access lines. +Unclamp <$B53333>arterial and <$2268FF>venous bloodlines (4 clamps). +Connect substitution line to the appropriate Pre or Post dilution port. [Disconnect^Disconnect Patient BloodLines] Title = Disconnect Patient BloodLines -Clamp and disconnect patient bloodlines. = Install_Advanced_2.png +Clamp and disconnect <$2268FF>venous bloodline. = Install_Advanced_2.png +Connect <$2268FF>venous bloodline into <$2268FF>venous needleless port. +Connect <$B53333>arterial bloodline into <$B53333>arterial needleless port. [Remove] Title = Remove Consumables and Disposables -Reconnect the Hanson connectors from the dialyzer to the machine. = Install_Advanced_2.png -Disconnect and remove acid concentrate and dry bicarbonate as needed. = Install_Advanced_1.png -Tap Auto Eject button and remove the blood tube set. = Install_Advanced_2.png +Remove acid concentrate and dry bicarbonate cartridge as needed. +Remove dialysate quick connectors and connect to the machine. +Press 'Auto-Eject' button to remove blood tubing set. +Discard blood tubing set per policy. -