Index: configurations/Messages/Unhandled.conf =================================================================== diff -u -r92dd2af2869b1c802342e54a6e1fbd79d5d9b1f1 -rce7a3bda3e7e5fcb13c959927e044569d68defdd --- configurations/Messages/Unhandled.conf (.../Unhandled.conf) (revision 92dd2af2869b1c802342e54a6e1fbd79d5d9b1f1) +++ configurations/Messages/Unhandled.conf (.../Unhandled.conf) (revision ce7a3bda3e7e5fcb13c959927e044569d68defdd) @@ -277,6 +277,11 @@ F32=d29RTDResist F32=d43RTDResist F32=d74RTDResist +F32=d17CondUncomp +F32=d27CondUncomp +F32=d29CondUncomp +F32=d43CondUncomp +F32=d74CondUncomp [0x2000] MSG_ID_TD_AIR_PUMP_DATA @@ -546,6 +551,8 @@ F32=p18CondResist F32=p9RTDResist F32=p18RTDResist +F32=p9CondUncomp +F32=p18CondUncomp [0x3800] MSG_ID_AVAILABLE_6 @@ -765,12 +772,10 @@ MSG_ID_TD_BATTERY_DATA [0x5500] -MSG_ID_FP_FILTER_FLUSH_DATA -U32=filterFlushExecState +MSG_ID_UI_PATIENT_DISCONNECT_CONFIRM_REQUEST [0x5600] -MSG_ID_FP_PERMEATE_FLUSH_DATA -U32=permeateFlushExecState +MSG_ID_TD_PATIENT_DISCONNECT_CONFIRM_RESPONSE [0x5700] MSG_ID_FP_CONCENTRATE_FLUSH_DATA @@ -1157,16 +1162,16 @@ MSG_ID_TD_ADJUST_TREATMENT_LOGS_RESPONSE [0xAB00] -MSG_ID_UI_ADJUST_PATIENT_DISCONNECT_CONFIRM_REQUEST +MSG_ID_TD_WATER_SAMPLE_RESULT_RESPONSE [0xAC00] -MSG_ID_TD_ADJUST_PATIENT_DISCONNECT_CONFIRM_RESPONSE +MSG_ID_TD_WATER_SAMPLE_DATA [0xAD00] -MSG_ID_TD_DRY_SELF_TEST_PROGRESS_DATA +MSG_ID_TD_TREATMENT_LOG_AVERAGE_DATA [0xAE00] -MSG_ID_TD_TUBING_SET_BARCODE_SCAN_REQUEST +MSG_ID_TD_DRY_SELF_TEST_PROGRESS_DATA [0xAF00] MSG_ID_TD_TUBE_SET_AUTHENTICATION_REQUEST @@ -1199,15 +1204,14 @@ MSG_ID_FFU_SIGNAL_DD_UPDATE_AVAILABLE [0xB500] -MSG_ID_DD_SUBSTITUTION_PUMP_DATA -F32=d92_PumpTargetRate -F32=d92_PumpCurrentSetSpeed -U32=d92_PumpState -F32=d92_PumpMeasuredSpeed +MSG_ID_TD_UI_GENERIC_CONFIRMATION_REQUEST [0xB600] -MSG_ID_DD_CONDUCTIVITY_SENSOR_RESISTANCE_DATA +MSG_ID_UI_GENERIC_CONFIRMATION_RESULT_RESPONSE +[0xB700] +MSG_ID_AVAILABLE_B7 + [0xB800] MSG_ID_UI_VITALS_ADJUSTMENT_REQUEST U32=systolicBpLowAlarmLimit @@ -1225,6 +1229,24 @@ U32=heartRateHighAlarmLimitRejectionReason U32=vitalsMeasurementIntervalRejectionReason +[0xDB00] +MSG_ID_DD_SUBSTITUTION_PUMP_DATA +F32=d92_PumpTargetRate +F32=d92_PumpCurrentSetSpeed +U32=d92_PumpState +F32=d92_PumpMeasuredSpeed + +[0xDC00] +MSG_ID_DD_CONDUCTIVITY_SENSOR_RESISTANCE_DATA + +[0xDD00] +MSG_ID_FP_FILTER_FLUSH_DATA +U32=filterFlushExecState + +[0xDE00] +MSG_ID_FP_PERMEATE_FLUSH_DATA +U32=permeateFlushExecState + [0x0080] MSG_ID_FIRST_TD_TESTER_MESSAGE @@ -1685,6 +1707,12 @@ [0x6080] MSG_ID_TD_SYRINGE_PUMP_FORCE_SENSOR_CALIBRATION_REQUEST +[0x6180] +MSG_ID_TD_GET_ALARM_PROPERTIES_REQUEST + +[0x6280] +MSG_ID_TD_ALARM_PROPERTIES_RESPONSE + [0x00A0] MSG_ID_FIRST_DD_TESTER_MESSAGE @@ -1728,12 +1756,10 @@ U32=index [0x08A0] -MSG_ID_DD_TREATMENT_PARAMS_OVERRIDE_REQUEST -U32=reset -F32=state +MSG_ID_DD_PRESSURE_SENSOR_READ_COUNTER_OVERRIDE_REQUEST [0x09A0] -MSG_ID_AVAILABLE_3 +MSG_ID_DD_PRESSURE_SENSOR_ERROR_COUNTER_OVERRIDE_REQUEST [0x0AA0] MSG_ID_DD_PRESSURE_SENSOR_PUBLISH_INTERVAL_OVERRIDE_REQUEST @@ -1909,7 +1935,7 @@ MSG_ID_DD_SET_OPERATION_MODE_OVERRIDE_REQUEST U32=data -[0x3EA0] +[0x2AA0] MSG_ID_DD_UF_DATA_PUBLISH_OVERRIDE_REQUEST U32=reset U32=state @@ -2381,6 +2407,11 @@ U32=reset U32=state +[0x75A0] +MSG_ID_DD_TREATMENT_PARAMS_OVERRIDE_REQUEST +U32=reset +F32=state + [0x00B0] MSG_ID_FIRST_FP_TESTER_MESSAGE