Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -rbc73f92d529c2a8d507340d158b57685048a1a96 -re958dae7249cf82b7f4cd8e49e8da8ba4eaaad40 --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision bc73f92d529c2a8d507340d158b57685048a1a96) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision e958dae7249cf82b7f4cd8e49e8da8ba4eaaad40) @@ -123,12 +123,12 @@ ListTitle = TD venous air bubble detector self-test failure [31] -Title = Service Required: Dialysate Device -ListTitle = DD fresh dialysate pressure sensor reporting error or not being read successfully +Title = MAX Blood Prime Pause Time +ListTitle = Blood Prime pause approaching timeout (> 4 min). [32] -Title = Service Required: Dialysate Device -ListTitle = DD transmembrane sensor reporting error or not being read successfully +Title = MAX Blood Prime Pause Time Reached +ListTitle = Blood Prime pause timeout reached (> 5 min). [33] Title = Service Required: Hemodialysis Device @@ -662,3 +662,91 @@ Title = Blood Pump Stopped: Venous Air Detector ListTitle = Air trap lowering timeout during treatment +[166] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump self-test failure alarm + +[167] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump is running while the BP is off + +[168] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump ADC error + +[169] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump fault reported by FPGA + +[170] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump DAC write failure + +[171] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump speed check error + +[172] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump volume check error + +[173] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump over travel error + +[174] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump not stopped in off state error + +[175] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump direction (from encoder) error + +[176] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump direction (from controller) error + +[177] +Title = Heparin Syringe Empty +ListTitle = TD syringe empty alarm + +[178] +Title = Blood Stopped: Syringe Not Detected +ListTitle = TD syringe pump syringe removed alarm + +[179] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump stall alarm + +[180] +Title = Service Required: Hemodialysis Device +ListTitle = TD heparin force sensor invalid calibration record + +[181] +Title = Syringe Detected +ListTitle = TD syringe detected alarm + +[182] +Title = Not Enough Heparin +ListTitle = TD syringe pump not enough Heparin alarm + +[183] +Title = Heparin Pump Occlusion +ListTitle = TD syringe pump occlusion alarm + +[184] +Title = Service Required: Hemodialysis Device +ListTitle = TD syringe pump timeout alarm + +[185] +Title = Service Required: FP Device +ListTitle = FP conductivity sensor (P9) located in inlet water path reporting error or not being read successfully + +[186] +Title = Service Required: FP Device +ListTitle = FP conductivity sensor (P18) located in inlet water path reporting error or not being read successfully + +[187] +Title = Air Pump Stall Detected +ListTitle = Air pump stall detected during treatment +