Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -re958dae7249cf82b7f4cd8e49e8da8ba4eaaad40 -rf418281ce2ca40cb306f74c1c83bf692de4f5f5e --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision e958dae7249cf82b7f4cd8e49e8da8ba4eaaad40) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision f418281ce2ca40cb306f74c1c83bf692de4f5f5e) @@ -747,6 +747,58 @@ ListTitle = FP conductivity sensor (P18) located in inlet water path reporting error or not being read successfully [187] +Title = Invalid Blood Tubing Set +ListTitle = TD detected invalid tubing set during authentication + +[188] +Title = Install New Blood Tubing Set +ListTitle = TD detected used#@% wet#@% or missing blood tubing set + +[189] +Title = No Blood Tubing Set Detected +ListTitle = TD did not detect tubing set during verification + +[190] +Title = Tubing Set Authentication Timeout +ListTitle = TD did not receive authentication response from UI + +[191] +Title = Pressure Too High to Zero +ListTitle = TD detected high pressure during zeroing + +[192] +Title = Venous Pressurization Timeout +ListTitle = TD failed to reach venous pressure target + +[193] +Title = Venous Pressure Decay +ListTitle = TD detected excessive venous pressure decay + +[194] +Title = Venous Pressure Leak +ListTitle = TD detected venous pressure leak + +[195] +Title = Venous Pressure Relief Failure +ListTitle = TD failed venous pressure relief check + +[196] +Title = Arterial Pressurization Timeout +ListTitle = TD failed to reach arterial pressure target + +[197] +Title = Arterial Pressure Decay +ListTitle = TD detected excessive arterial pressure decay + +[198] +Title = Arterial Pressure Leak +ListTitle = TD detected arterial pressure leak + +[199] +Title = Arterial Pressure Relief Failure +ListTitle = TD failed arterial pressure relief check + +[200] Title = Air Pump Stall Detected ListTitle = Air pump stall detected during treatment