Index: configurations/Alarms/Alarms.conf =================================================================== diff -u -r05e5dc72479dd945df0e98df9a3e9f2ade153c7e -rf45e77d1da8294f866eeb7674c9ee8cbaed97bbc --- configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision 05e5dc72479dd945df0e98df9a3e9f2ade153c7e) +++ configurations/Alarms/Alarms.conf (.../Alarms.conf) (revision f45e77d1da8294f866eeb7674c9ee8cbaed97bbc) @@ -1 +1,660 @@ +[ALARM_ID_NO_ALARM] +Title = Indication for when no alarms is a possible situation +ListTitle = +Message = + +[ALARM_ID_TD_SOFTWARE_FAULT] +Title = Service Required: Hemodialysis Device +ListTitle = TD Software fault. Software found itself in an unexpected state +Message = TD Fault: SW Invalid State + +[ALARM_ID_DD_SOFTWARE_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD software fault. Software found itself in an unexpected state +Message = DD Fault: Software Invalid State + +[ALARM_ID_RO_SOFTWARE_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD software fault. Software found itself in an unexpected state +Message = DD Fault: Software Invalid State + +[ALARM_ID_TD_FPGA_POST_TEST_FAILED] +Title = Service Required: Hemodialysis Device +ListTitle = TD FPGA POST failure +Message = TD POST: FPGA Compatibility + +[ALARM_ID_TD_WATCHDOG_POST_TEST_FAILED] +Title = Service Required: Hemodialysis Device +ListTitle = TD Watchdog POST failure +Message = TD POST: WD Safestate + +[ALARM_ID_TD_UI_COMM_POST_FAILED] +Title = Service Required: Hemodialysis Device +ListTitle = UI communication POST failure. UI failed to communicate within a reasonable time after power up +Message = TD POST: Proc- UI Comm + +[ALARM_ID_TD_FPGA_CLOCK_SPEED_CHECK_FAILURE] +Title = Service Required: Hemodialysis Device +ListTitle = TD processor clock speed checks against FPGA clock failure +Message = TD Fault: Proc- FPGA Clock + +[ALARM_ID_TD_FPGA_COMM_TIMEOUT] +Title = Service Required: Hemodialysis Device +ListTitle = TD FPGA communication down for too long +Message = TD Fault: FPGA Com + +[ALARM_ID_TD_UI_COMM_TIMEOUT] +Title = Service Required: Hemodialysis Device +ListTitle = User interface communication timeout +Message = TD Fault: UI Comm + +[ALARM_ID_TD_DD_COMM_TIMEOUT] +Title = Communication Error +ListTitle = Dialysate generator communication timeout +Message = Loss of DD Comm (Timeout) + +[ALARM_ID_TD_COMM_TOO_MANY_BAD_CRCS] +Title = Service Required: Hemodialysis Device +ListTitle = Too many bad CRCs detected on received system messages +Message = TD Fault: CAN CRC Com + +[ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_UI] +Title = Service Required: Hemodialysis Device +ListTitle = System message that required acknowledgment was not acknowledged +Message = TD Fault: CAN ACK UI Com + +[ALARM_ID_TD_CAN_MESSAGE_NOT_ACKED_BY_DD] +Title = Communication Error +ListTitle = TD to DD comm error. +Message = Loss of DD Comm (DD ACK) + +[ALARM_ID_TD_CRITICAL_DATA_ERROR] +Title = Service Required: Hemodialysis Device +ListTitle = TD critical data integrity check failed +Message = TD Fault: Critical Data Err + +[ALARM_ID_DD_CRITICAL_DATA_ERROR] +Title = Service Required: Dialysate Device +ListTitle = DD critical data integrity check failed +Message = DD Fault: Critical Data Err + +[ALARM_ID_TD_WATCHDOG_EXPIRED] +Title = Service Required: Hemodialysis Device +ListTitle = TD watchdog expired +Message = TD Fault: WD Error + +[ALARM_ID_DD_WATCHDOG_EXPIRED] +Title = Service Required: Dialysate Device +ListTitle = DD watchdog expired +Message = DD Fault: WD Error + +[ALARM_ID_DD_COMM_TOO_MANY_BAD_CRCS] +Title = Service Required: Dialysate Device +ListTitle = Too many bad CRCs detected on received system messages +Message = DD Fault: CAN CRC Com + +[ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_RO] +Title = Service Required: Dialysate Device +ListTitle = DD to RO comm error +Message = Loss of RO Comm (RO ACK) + +[ALARM_ID_DD_CAN_MESSAGE_NOT_ACKED_BY_TD] +Title = Service Required: Dialysate Device +ListTitle = DD to TD comm error +Message = Loss of TD Comm (TD ACK) + +[ALARM_ID_DD_FPGA_POST_TEST_FAILED] +Title = Service Required: Dialysate Device +ListTitle = DD FPGA POST failure +Message = DD POST: FPGA Compatibility + +[ALARM_ID_DD_FPGA_CLOCK_SPEED_CHECK_FAILURE] +Title = Service Required: Dialysate Device +ListTitle = FPGA clock speed failure +Message = DD Fault: Proc- FPGA Clock + +[ALARM_ID_DD_FPGA_COMM_TIMEOUT] +Title = Service Required: Dialysate Device +ListTitle = DD FPGA communication down for too long +Message = DD Fault: FPGA Com + +[ALARM_ID_TD_ALARM_AUDIO_SELF_TEST_FAILURE] +Title = Service Required: Hemodialysis Device +ListTitle = TD alarm audio failed POST +Message = TD POST: Primary Alarm Current + +[ALARM_ID_DD_VALVE_CONTROL_FAILURE] +Title = Service Required: Dialysate Device +ListTitle = DD FPGA not accepting commanded valve states +Message = DD Fault: Valve Control + +[ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT] +Title = Service Required: Hemodialysis Device +ListTitle = TD arterial pressure sensor reporting error or not being read successfully +Message = TD Arterial Pressure Sensor Error + +[ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT] +Title = Service Required: Hemodialysis Device +ListTitle = TD venous pressure sensor reporting error or not being read successfully +Message = TD Venous Pressure Sensor Error + +[ALARM_ID_DD_HYD_OUTLET_PRES_TIMEOUT_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD hydraulics outlet pressure sensor reporting error or not being read successfully +Message = DD Hyd outlet Pressure Sensor Error + +[ALARM_ID_DD_BIBAG_PRES_TIMEOUT_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD bibag pressure sensor reporting error or not being read successfully +Message = DD Bibag Pressure Sensor Error + +[ALARM_ID_DD_SPENT_DIALYSATE_PRES_TIMEOUT_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD spent dialysate pressure sensor reporting error or not being read successfully +Message = DD Spent Dial Pressure Sensor Error + +[ALARM_ID_DD_FRESH_DIALYSATE_PRES_TIMEOUT_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD fresh dialysate pressure sensor reporting error or not being read successfully +Message = DD Fresh Dial Pressure Sensor Error + +[ALARM_ID_DD_TRANSMEMB_PRES_TIMEOUT_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD transmembrane sensor reporting error or not being read successfully +Message = DD Transmemb Pressure Sensor Error + +[ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE] +Title = Service Required: Hemodialysis Device +ListTitle = TD monitored voltage is out of range +Message = TD Fault: Voltage Monitor + +[ALARM_ID_TD_AC_POWER_LOST] +Title = Power Loss +ListTitle = TD AC power lost alarm +Message = Loss of AC Power + +[ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT] +Title = Power Loss +ListTitle = TD AC power lost alarm +Message = Blood Stop: Loss of AC Power + +[ALARM_ID_DD_D17_COND_SENSOR_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully +Message = DD Conductivity Sensor Error + +[ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully +Message = DD Conductivity Sensor Error + +[ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully +Message = DD Conductivity Sensor Error + +[ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully +Message = DD Conductivity Sensor Error + +[ALARM_ID_TD_VENOUS_BUBBLE_DETECTED] +Title = Venous Air Detected +ListTitle = TD venous air bubble detected alarm +Message = Blood Stop: Venous Air Bubble + +[ALARM_ID_TD_CARTRIDGE_DOOR_OPENED] +Title = Blood Pump Stopped: Door Open +ListTitle = TD cartridge door opened alarm +Message = Blood Stop: Door Open + +[ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED] +Title = Stuck Button +ListTitle = TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up +Message = TD POST: Stuck Button + +[ALARM_ID_TD_ARTERIAL_PRESSURE_LOW] +Title = Low Arterial Pressure +ListTitle = Arterial pressure too low during treatment +Message = Blood Stop: Arterial Pres. Low + +[ALARM_ID_TD_ARTERIAL_PRESSURE_HIGH] +Title = High Arterial Pressure +ListTitle = Arterial pressure too high during treatment +Message = Blood Stop: Arterial Pres. High + +[ALARM_ID_TD_ARTERIAL_PRESSURE_OUT_OF_RANGE] +Title = Service Required: Hemodialysis Device +ListTitle = TD arterial pressure sensor is reading out of range +Message = TD Fault: Arterial Press Range + +[ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE] +Title = Self Test: Cartridge +ListTitle = TD arterial pressure self-test failure alarm +Message = Process: Pre- Tx Remove Cartridge + +[ALARM_ID_TD_VENOUS_PRESSURE_LOW] +Title = Low Venous Pressure +ListTitle = Venous pressure too low during treatment +Message = Blood Stop: Venous Pres. Low + +[ALARM_ID_TD_VENOUS_PRESSURE_HIGH] +Title = High Venous Pressure +ListTitle = Venous pressure too high during treatment +Message = Blood Stop: Venous Pres. High + +[ALARM_ID_TD_VENOUS_PRESSURE_OUT_OF_RANGE] +Title = Service Required: Hemodialysis Device +ListTitle = TD venous pressure sensor is reading out of range +Message = TD Fault: Venous Press Range + +[ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE] +Title = Self Test: Cartridge +ListTitle = TD venous pressure self-test failure alarm +Message = Process: Pre- Tx Remove Cartridge + +[ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE] +Title = Self Test: Cartridge Check +ListTitle = TD pre-treatment mode dry self-tests failure +Message = Process: Pre- Tx Dry Self- Test + +[ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR] +Title = Service Required: Dialysate Device +ListTitle = DD concentrate pump D11_PUMP speed control error +Message = DD Fault: Pump Speed Fault + +[ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR] +Title = Service Required: Dialysate Device +ListTitle = DD concentrate pump D10_PUMP speed control error +Message = DD Fault: Pump Speed Fault + +[ALARM_ID_DD_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD concentrate pump hall sensor out of range. +Message = DD Fault: Pump Speed Fault + +[ALARM_ID_DD_CONCENTRATE_PUMP_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD concentrate pump fault +Message = DD Fault: Pump Fault + +[ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD temperature sensor out of range +Message = DD Fault: Temp Sensor Range + +[ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD temperature sensors not being read successfully +Message = DD Fault: Temperature RTD Comm + +[ALARM_ID_DD_BARO_SENSOR_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD barometric sensor not being read successfully +Message = DD Fault: Baro temp Comm + +[ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC] +Title = Service Required: Dialysate Device +ListTitle = DD barometric pressure sensor bad coefficients CRC +Message = DD Fault: Baro temp Cal CRC + +[ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD fresh dialysate pump RPM out of range +Message = DD Fault: Fresh Dial Pump Speed Fault + +[ALARM_ID_DD_D12_PUMP_OFF_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD fresh dialysate pump off fault +Message = DD Fault: Fresh Dial Pump Speed Not Off + +[ALARM_ID_DD_D12_PUMP_CURRENT_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD fresh dialysate pump current out of range +Message = DD Fault: Fresh Dial Pump Current + +[ALARM_ID_DD_D12_PUMP_DIRECTION_INVALID] +Title = Service Required: Dialysate Device +ListTitle = DD fresh dialysate pump invalid direction +Message = DD Fault: Fresh Dial Pump Dir Fault + +[ALARM_ID_DD_D12_PUMP_DIRECTION_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD fresh dialysate pump direction FPGA fault +Message = DD Fault: Fresh Dial pump Dir Err Cnt + +[ALARM_ID_DD_D48_PUMP_RPM_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD spent dialysate pump RPM out of range +Message = DD Fault: Spent Dial Pump Speed Fault + +[ALARM_ID_DD_D48_PUMP_OFF_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD spent dialysate pump off fault +Message = DD Fault: Spent Dial Pump Speed Not Off + +[ALARM_ID_DD_D48_PUMP_CURRENT_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD spent dialysate pump current out of range +Message = DD Fault: Spent Dial Pump Current + +[ALARM_ID_DD_D48_PUMP_DIRECTION_INVALID] +Title = Service Required: Dialysate Device +ListTitle = DD spent dialysate pump invalid direction +Message = DD Fault: Spent Dial Pump Dir Fault + +[ALARM_ID_DD_D48_PUMP_DIRECTION_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD spent dialysate pump direction FPGA fault +Message = DD Fault: Spent Dial pump Dir Err Cnt + +[ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON] +Title = Service Required: Dialysate Device +ListTitle = DD D5 heater is On with low fluid level +Message = DD Fault: Heater On With No Fluid + +[ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON] +Title = Service Required: Dialysate Device +ListTitle = DD D45 heater is On with low fluid level +Message = DD Fault: Heater On With No Fluid + +[ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD water inlet input pressure sensor reporting error or not being read successfully +Message = DD Water Inlet Input Press Sensor Error + +[ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD water inlet output pressure sensor reporting error or not being read successfully +Message = DD Water Inlet Output Press Sensor Error + +[ALARM_ID_DD_INLET_WATER_PRESSURE_IN_LOW_RANGE] +Title = Incoming Water Pressure Input Low +ListTitle = DD Inlet water pressure input too low +Message = DD: Inlet Water Pres Input Low + +[ALARM_ID_DD_INLET_WATER_PRESSURE_IN_HIGH_RANGE] +Title = Incoming Water Pressure Input High +ListTitle = DD inlet water pressure input is too high +Message = DD: Inlet Water Pres Input High + +[ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_LOW_RANGE] +Title = Incoming Water Pressure Output Low +ListTitle = DD Inlet water pressure output too low +Message = DD: Inlet Water Pres Output Low + +[ALARM_ID_DD_INLET_WATER_PRESSURE_OUT_HIGH_RANGE] +Title = Incoming Water Pressure Output High +ListTitle = DD inlet water pressure output is too high +Message = DD: Inlet Water Pres Output High + +[ALARM_ID_TD_AIR_TRAP_FILL_DURING_TREATMENT] +Title = Blood Pump Stopped: Venous Air Detector +ListTitle = Air trap fill timeout during treatment +Message = Blood Stop: Air Trap Fill + +[ALARM_ID_TD_AIR_TRAP_ILLEGAL_LEVELS] +Title = Service Required: Hemodialysis Device +ListTitle = Air trap level sensors reporting illegal combination of air/fluid +Message = TD Fault: Air Trap Sensor + +[ALARM_ID_TD_PINCH_VALVE_FAULT] +Title = Service Required: Hemodialysis Device +ListTitle = Pinch valve reporting a fault +Message = TD Fault: Pinch Valve + +[ALARM_ID_TD_VALVE_HOMING_FAILED] +Title = Service Required: Hemodialysis Device +ListTitle = TD valve homing failed +Message = TD Fault: Valve Home + +[ALARM_ID_TD_VALVE_TRANSITION_TIMEOUT] +Title = Service Required: Hemodialysis Device +ListTitle = TD valve transition time out +Message = TD Fault: Valve Transition Err + +[ALARM_ID_RO_CAN_MESSAGE_NOT_ACKED_BY_DD] +Title = Communication Error +ListTitle = DD to RO comm error. +Message = RO Loss of DD Comm (DD ACK) + +[ALARM_ID_RO_FPGA_COMM_TIMEOUT] +Title = Service Required: Dialysate Device +ListTitle = RO FPGA communication down for too long +Message = RO Fault: FPGA Comm + +[ALARM_ID_RO_COMM_TOO_MANY_BAD_CRCS] +Title = Service Required: Hemodialysis Device +ListTitle = RO Too many bad CRCs detected on received system messages +Message = RO Fault: CAN CRC Comm + +[ALARM_ID_RO_VALVE_CONTROL_FAILURE] +Title = Service Required: Filtration Device +ListTitle = RO FPGA not accepting commanded valve states +Message = RO Fault: Valve Control + +[ALARM_ID_DD_BC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD balancing chamber state 1 fill start pressure is not in range +Message = DD Fault: BC Pressure not in range + +[ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD balancing chamber state 1 fill end pressure is not in range +Message = DD Fault: BC Pressure not in range + +[ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD balancing chamber state 2 fill start pressure is not in range +Message = DD Fault: BC Pressure not in range + +[ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD balancing chamber state 2 fill end pressure is not in range +Message = DD Fault: BC Pressure not in range + +[ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD hydraulics chamber degassing pressure is not in range +Message = DD Fault: Hyd Cham Neg Pres not in range + +[ALARM_ID_AVAILABLE_92] +Title = +ListTitle = +Message = + +[ALARM_ID_AVAILABLE_93] +Title = +ListTitle = +Message = + +[ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD balancing chamber state 1 fill initiated and pressure drop is not in range +Message = DD Fault: BC Pressure not in range + +[ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE] +Title = Service Required: Dialysate Device +ListTitle = DD balancing chamber state 2 fill initiated and pressure drop is not in range +Message = DD Fault: BC Pressure not in range + +[ALARM_ID_TD_TREATMENT_STOPPED_BY_USER] +Title = Blood Pump Stopped: Treatment Paused +ListTitle = TD treatment stopped by user action - pressed stop button +Message = Blood Stopped: Treatment Pause + +[ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT] +Title = Service Required: Dialysate Device +ListTitle = DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully +Message = DD Conductivity Sensor Error + +[ALARM_ID_AVAILABLE_98] +Title = +ListTitle = +Message = + +[ALARM_ID_AVAILABLE_99] +Title = +ListTitle = +Message = + +[ALARM_ID_TD_COMM_TIMEOUT] +Title = Service Required: Hemodialysis Device +ListTitle = TD communication timeout +Message = TD Fault: UI- Proc Comm + +[ALARM_ID_UI_POST_FAILURE_OS_VERSION] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST OS Version compatibility failure +Message = TD POST: OS Version + +[ALARM_ID_UI_POST_FAILURE_SHASUM] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST Application Integrity (Sha256Sum) failure +Message = TD POST: UI Image + +[ALARM_ID_UI_POST_FAILURE_CANBUS] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST CANBus failure +Message = TD POST: UI CAN Driver + +[ALARM_ID_UI_POST_FAILURE_DISPLAY] +Title = Service Required: Hemodialysis Device +ListTitle = If the UI POST reports display failure. +Message = UI POST Display failure + +[ALARM_ID_UI_POST_FAILURE_TOUCH] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST Touch failure +Message = TD POST: UI Touch Driver + +[ALARM_ID_UI_POST_FAILURE_SDCARD] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST SD-Card failure +Message = TD POST: UI SD- Card + +[ALARM_ID_UI_POST_FAILURE_RTC] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST RTC failure +Message = TD POST: UI RTC + +[ALARM_ID_UI_POST_FAILURE_WIFI] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST WiFi failure +Message = TD POST: UI WiFi + +[ALARM_ID_UI_POST_FAILURE_BLUETOOTH] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST Bluetooth failure +Message = TD POST: UI Bluetooth + +[ALARM_ID_UI_POST_FAILURE_ETHERNET] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST Ethernet failure +Message = TD POST: UI Ethernet + +[ALARM_ID_UI_POST_FAILURE_SOUND] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST Sound failure +Message = TD POST: UI Sound + +[ALARM_ID_UI_POST_FAILURE_CLOUDSYNC] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST cloud sync POST failure +Message = TD POST: UI CloudSync + +[ALARM_ID_UI_POST_FAILURE_INVALID_YEAR] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST Invalid Year +Message = TD POST Alarm: UI Invalid Year + +[ALARM_ID_UI_POST_FAILURE_SETTINGS_BAD] +Title = Service Required: Hemodialysis Device +ListTitle = UI POST Bad Settings +Message = TD POST Alarm: UI Bad Settings + +[ALARM_ID_UI_EVNT_FAILURE_SDCARD] +Title = Service Required: Hemodialysis Device +ListTitle = SD card is non-operational +Message = TD Fault: UI- SD- Card Fault + +[ALARM_ID_UI_RESERVED_116] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_117] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_118] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_119] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_120] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_121] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_122] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_123] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_124] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_125] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_126] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_127] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_128] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_129] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_UI_RESERVED_130] +Title = Service Required: Hemodialysis Device +ListTitle = RESERVED +Message = TD Fault: UI- RESERVED + +[ALARM_ID_BATTERY_LOW] +Title = Battery Capacity Low +ListTitle = Insufficient battery charge to support 20 min without A/C power +Message = TD Alarm (Low): Battery Low +