Index: configurations/Instructions/Hydraulics_1.png =================================================================== diff -u Binary files differ Index: configurations/Instructions/Instructions.conf =================================================================== diff -u -rccb80ea6edea093a50dcc0787d28d1ac2efb6e05 -rf9d140bee7d3473aa38ee4aac0e56f30e827a683 --- configurations/Instructions/Instructions.conf (.../Instructions.conf) (revision ccb80ea6edea093a50dcc0787d28d1ac2efb6e05) +++ configurations/Instructions/Instructions.conf (.../Instructions.conf) (revision f9d140bee7d3473aa38ee4aac0e56f30e827a683) @@ -145,3 +145,10 @@ Press 'Auto-Eject' button to remove blood tubing set. Discard blood tubing set per policy. = Remove_4.png +[Deaeration Pressure] +Title = Calibrate Deaeration Pressure +Connect a gauge in line at the inlet side of the deaeration pump. = Hydraulics_1.png +Connect a gauge to the <$B53333>red Rinse Port. = Hydraulics_1.png +Press 'Start Calibration' to start the Deaeration Pump. = Hydraulics_1.png + +