Index: suite_leahi/shared/scripts/configuration/config.py =================================================================== diff -u -r4b929ef6eef32479f10e84325b0ad12173474a4d -ra6d893b28af0dfab579a0c4ee00055af0a6d152d --- suite_leahi/shared/scripts/configuration/config.py (.../config.py) (revision 4b929ef6eef32479f10e84325b0ad12173474a4d) +++ suite_leahi/shared/scripts/configuration/config.py (.../config.py) (revision a6d893b28af0dfab579a0c4ee00055af0a6d152d) @@ -11,6 +11,36 @@ EXPORT_LOG_PROGRESS_STATUS = "Service log export to USB in progress ... " SERVICE_SCREEN_TITLE_TEXT = "Service" +#Main Treatement +UF_VOLUME_GOAL = "UF Volume Goal" +UF_RATE = "UF Rate" +BLOOD_FLOW = "Blood Flow" +DIALYSATE_FLOW = "Dialysate Flow" +DIALYSATE_TEMP = "Dialysate Temp." +DIALYSATE_COND = "Dialysate Cond." + +#ultrafiltration + +MAX_UF_VOLUME = 10.00 +MAX_UF_RATE = 10.00 +MAX_VOLUME_DELIVERED = 8.00 + +MIN_UF_VOLUME = 0.35 +MIN_UF_RATE = 0.22 +MIN_VOLUME_DELIVERED = 0.09 + +MID_UF_VOLUME = 5.00 +MID_UF_RATE = 5.00 +MID_VOLUME_DELIVERED = 4.00 + +ULTRAFILTRATION_VOLUME = "Ultrafiltration Volume (L)" +UF_VOLUME_REMOVED = "UF Volume Removed" +UF_VOLUME_GOAL = "UF Volume Goal" + + +MAX_NEW_UF_VOLUME = 8.0 +MAX_NEW_UF_RATE = 2.0 + # Headerbar_information_popup INFORMATION_PARAMETERS = [ "OS Version", "UI Version", "TD Version", "TD FPGA Version", @@ -63,10 +93,10 @@ DISABLED = False #Create Rx -ACID_CONCENTRATE = ["1.0 K, 2.50 Ca, 1 Mg","2.0 K, 2.50 Ca, 1 Mg","3.0 K, 2.50 Ca, 1 Mg"] +ACID_CONCENTRATE = ["1.0 K, 2.50 Ca","2.0 K, 2.50 Ca","3.0 K, 2.50 Ca"] DIALYZER_TYPE = ["Diacap Pro 13H","Diacap Pro 16H", "Diacap Pro 19H", "Optiflux F160NRe", "Optiflux F180NRe", "Optiflux F200NRe","Optiflux F250NRe"] VITALS = ["OFF","5","10","15","20","30","60"] -BICARBONATE = ["Sodium Bicarbonate"] +BICARBONATE = ["650 g","720 g","760 g", "900 g", "1100 g","1250 g"] #Heparin heparin_off = { "Heparin Type" :"OFF", @@ -75,8 +105,8 @@ "Stop Time" :"OFF"} heparin_values = { "Heparin Type" :"Unfractionated 1,000", - "Dispensing Rate" :"0.5", - "Bolus Volume" :"0.4", + "Dispensing Rate" :"0.3", + "Bolus Volume" :"0.2", "Stop Time" :"60"} PAUSE = 'Pause' @@ -85,8 +115,8 @@ COMPLETE = 'Complete' HEPARIN = 'Heparin' CREATERX = 'Rx' -VALIDATE = 'VALIDATE' -CONFIRM = 'CONFIRM' +VALIDATE = 'Validate' +CONFIRM = 'Confirm' NEXT = 'Next' AUTO_LOAD = 'Auto Load' WATER_SAMPLE = 'Water Sample' @@ -210,3 +240,20 @@ "Install blood tubing set and dialyzer.", ] START_TREATMENT_REJECTION_NOTIFICATION = "[1] Request is not allowed in the current operating mode" + +COLOR_CODES = "#ffa500" +HDF ="HDF" +SYSTOLIC = "Systolic" +DIASTOLIC = "Diastolic" +HEARTRATE = "Heart Rate" + +MIN_SYSTOLIC = "60" +MIN_DIASTOLIC = "40" +MIN_HEARTRATE = "40" + +DISCONNECT_PATIENT_TITLE ="Disconnect Patient BloodLines" +DISCONNECT_PATIENT_TEXT ="Clamp and disconnect patient bloodlines." + + + +