Index: firmware/App/Modes/InletPressureCheck.c =================================================================== diff -u -rbdf1e3ac2eef7ec37034127e18aa6308148e894a -r44b3345d1d7e550011b93e32a5055eb5fb424b36 --- firmware/App/Modes/InletPressureCheck.c (.../InletPressureCheck.c) (revision bdf1e3ac2eef7ec37034127e18aa6308148e894a) +++ firmware/App/Modes/InletPressureCheck.c (.../InletPressureCheck.c) (revision 44b3345d1d7e550011b93e32a5055eb5fb424b36) @@ -15,16 +15,13 @@ * ***************************************************************************/ -#include "Conductivity.h" #include "BoostPump.h" #include "FlowSensor.h" #include "InletPressureCheck.h" #include "FPOperationModes.h" #include "MessageSupport.h" #include "Messaging.h" -#include "PermeateTank.h" #include "Pressure.h" -#include "ROPump.h" #include "TaskGeneral.h" #include "Timers.h" #include "Valves.h" @@ -212,7 +209,7 @@ setValveState( P39_VALV, VALVE_STATE_CLOSED ); setValveState( P20_VALV, VALVE_STATE_CLOSED ); setValveState( P43_VALV, VALVE_STATE_CLOSED ); - startPermeateTankControl(); + if ( TRUE == isBoostPumpInstalled() ) { setBoostPumpTargetDutyCycle( INLET_PRES_CHECK_BOOST_PUMP_TGT_PWM ); Index: firmware/App/Modes/InletPressureCheck.h =================================================================== diff -u -rbdf1e3ac2eef7ec37034127e18aa6308148e894a -r44b3345d1d7e550011b93e32a5055eb5fb424b36 --- firmware/App/Modes/InletPressureCheck.h (.../InletPressureCheck.h) (revision bdf1e3ac2eef7ec37034127e18aa6308148e894a) +++ firmware/App/Modes/InletPressureCheck.h (.../InletPressureCheck.h) (revision 44b3345d1d7e550011b93e32a5055eb5fb424b36) @@ -40,7 +40,6 @@ // ********** public function prototypes ********** void initInletPressureCheck( void ); // Initialize this Concentrate Flush unit -//U32 transitionToConcentrateFlush( void ); // Transition to Concentrate Flush void execInletPressureCheck( void ); // Execute the Concentrate Flush state machine BOOL isInletPressureCheckCompleted( void ); // returns if flush is complete Index: firmware/App/Modes/ModePreGenPermeate.c =================================================================== diff -u -rbdf1e3ac2eef7ec37034127e18aa6308148e894a -r44b3345d1d7e550011b93e32a5055eb5fb424b36 --- firmware/App/Modes/ModePreGenPermeate.c (.../ModePreGenPermeate.c) (revision bdf1e3ac2eef7ec37034127e18aa6308148e894a) +++ firmware/App/Modes/ModePreGenPermeate.c (.../ModePreGenPermeate.c) (revision 44b3345d1d7e550011b93e32a5055eb5fb424b36) @@ -241,6 +241,7 @@ if ( TRUE == isInletPressureCheckCompleted() ) { signalStartFilterFlush(); + state = FP_PRE_GENP_FILTER_FLUSH; } } return state;