Index: build/AlarmMapping.py =================================================================== diff -u -r3a7cda682fba9d589d7b84635bb7af8d05036fa0 -rb025700b82571054b74d13e4d2f8cd05c8770dac --- build/AlarmMapping.py (.../AlarmMapping.py) (revision 3a7cda682fba9d589d7b84635bb7af8d05036fa0) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision b025700b82571054b74d13e4d2f8cd05c8770dac) @@ -91,15 +91,15 @@ 89 : "DG outlet UV reactor on with no flow." 90 : "DG load cells A1/B1 FPGA fault." 91 : "HD No dialysate flow data receive in the last 3 seconds." - 92 : "Available for use." + 92 : "DG temperature sensors invalid calibration record." 93 : "DG outlet primary conductivity out of range." 94 : "DG pressure out of range." 95 : "DG watchdog expired." 96 : "DG inlet water temperature in high range." 97 : "DG fill conductivity out of range." 98 : "HD battery communication fault." 99 : "HD syringe pump stall alarm." - 100 : "Available for use." + 100 : "DG conductivity sensors invalid temperature compensation calibration record." 101 : "UI POST HD communication failure." 102 : "DG heat disinfect temperature gradient out of range." 103 : "HD invalid calibration CRC."