Index: build/AlarmMapping.py =================================================================== diff -u -r6b111c12d352c46d57cf9a991e70d6541f248534 -rd957a403d7e7928de8701fd5bb5535e26539cabd --- build/AlarmMapping.py (.../AlarmMapping.py) (revision 6b111c12d352c46d57cf9a991e70d6541f248534) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision d957a403d7e7928de8701fd5bb5535e26539cabd) @@ -17,7 +17,7 @@ 15 : "HD invalid usage record CRC." 16 : "HD alarm audio failed POST." 17 : "HD UI POST failed." - 18 : "HD didn't get ACK on message to DG that required acknowledgment." + 18 : "DG load cells invalid calibration record." 19 : "HD dialysate temperature above target temperature." 20 : "DG conductivity sensors invalid calibration record." 21 : "DG drain line volume invalid calibration record." @@ -45,7 +45,7 @@ 43 : "DG main primary heater FPGA fault." 44 : "User interface communication timeout." 45 : "HD too many bad communications CRC." - 46 : "HD didn't get ACK on message to UI that required acknowledgment." + 46 : "HD System message that required acknowledgment was not acknowledged." 47 : "Ultrafiltration rate is too high error during treatment." 48 : "Ultrafiltration volume accuracy error during treatment." 49 : "HD FPGA communication down for too long." @@ -91,7 +91,7 @@ 89 : "DG outlet UV reactor on with no flow." 90 : "DG load cells A1/B1 FPGA fault." 91 : "HD No dialysate flow data receive in the last 3 seconds." - 92 : "DG conductivity sensor bad status." + 92 : "Alarm ID available for use." 93 : "DG outlet primary conductivity out of range." 94 : "DG pressure out of range." 95 : "DG watchdog expired." @@ -100,7 +100,7 @@ 98 : "HD battery communication fault." 99 : "HD syringe pump stall alarm." 100 : "HD no cartridge self-test timeout." - 101 : "UI POST HD communication failure." + 101 : "HD dry self-test timeout." 102 : "DG heat disinfect temperature gradient out of range." 103 : "HD invalid calibration CRC." 104 : "Air trap level sensors reporting illegal combination of air/fluid." @@ -115,7 +115,7 @@ 113 : "DG Drain pump current out of range." 114 : "HD venous air bubble detector self-test failure." 115 : "DG temperature sensor out of range." - 116 : "Available for use." + 116 : "DG temperature sensor ADC fault." 117 : "DG CAN message not acked." 118 : "DG RTC config error." 119 : "DG RTC battery low error." @@ -165,7 +165,7 @@ 163 : "HD sees primary load cell for reservoir 1 change too much too fast." 164 : "HD sees primary load cell for reservoir 2 change too much too fast." 165 : "HD in treatment rinseback operation timeout." - 166 : "Available for use." + 166 : "HD in treatment re-circ sub-mode for too long." 167 : "HD cartridge door opened alarm." 168 : "HD active reservoir depletion timeout." 169 : "DG Dialysate flow rate out of maximum range." @@ -175,7 +175,7 @@ 173 : "HD RTC configuration error." 174 : "HD RTC or timer accuracy failure." 175 : "HD pump direction status error." - 176 : "HD dialysate temperature above high safety range." + 176 : "HD dialysate temperature out of high safety range." 177 : "DG software fault.\nSoftware found itself in an unexpected state." 178 : "HD communication timeout." 179 : "DG FPGA communication down for too long." @@ -192,7 +192,7 @@ 190 : "DG drain pump off fault." 191 : "DG flow rate out of upper range." 192 : "DG flow rate out of lower range." - 193 : "HD dialysate temperature above low safety range." + 193 : "HD dialysate temperature out of low safety range." 194 : "DG RO pump duty cycle out of range." 195 : "DG RO pump pressure out of range." 196 : "DG CPi/CPo sensors FPGA fault." @@ -202,7 +202,7 @@ 200 : "DG thermistors/sensors temperature out of range." 201 : "HD pre-treatment mode wet self-test lc vs FMD failure." 202 : "HD pre-treatment mode dry pressure normal self-test failure." - 203 : "Available for use." + 203 : "Pressure sensor fault." 204 : "RO rejection ratio out of range." 205 : "Conductivity sensor fault." 206 : "DG dialysate fill runs out of time." @@ -256,8 +256,8 @@ 254 : "DG invalid service record." 255 : "HD and UI software builds are not compatible." 256 : "HD processor is in Disinfect Chemical Flush mode." - 257 : "DG FPGA power out timeout." - 258 : "***Available for use***." + 257 : "Alarm ID available for use." + 258 : "HD FPGA power out timeout." 259 : "HD temperatures out of range." 260 : "UI POST Application Integrity (Sha256Sum) failure." 261 : "UI POST CANBus failure." @@ -284,8 +284,8 @@ 282 : "Bicarbonate concentration bottle low volume alarm." 283 : "DG load cells weight out of range." 284 : "DG load cells primary/back up drift out of range." - 285 : "Available for use." - 286 : "Available for use." + 285 : "HD treatment recirculate timeout warning." + 286 : "HD treatment rinseback complete timeout warning." 287 : "HD processor clock speed checks against FPGA clock failure." 288 : "HD load cells primary/back up drift out of range." 289 : "DG dialysate or concentrate caps not closed." @@ -305,7 +305,7 @@ 303 : "DG drain pump direction FPGA fault." 304 : "HD invalid usage record." 305 : "HD processor RAM error." - 306 : "***Available for use***." + 306 : "HD air trap over-fill alarm." 307 : "HD AC power lost second alarm." 308 : "DG CPi conductivity sensor invalid character received." 309 : "DG CPo conductivity sensor invalid character received." @@ -321,8 +321,7 @@ 319 : "DG chemical disinfect flush flush sample." 320 : "DG chemical disinfect flush sample timeout." 321 : "DG outlet UV reactor not healthy." - 322 : "RM_ID_HD_DIALYSATE_TEMP_BELOW_LOW_SAFETY_RANGE = 322,." - 323 : "Total number of alarms." + 322 : "Total number of alarms." } # dictionary of rejection reason from application source code.