Index: build/AlarmMapping.py =================================================================== diff -u -r92e3c08040ba5a628b99f06675423b6011affed2 -r0ccc06cd10ffae3cf76d1abebe0ab7f93d506337 --- build/AlarmMapping.py (.../AlarmMapping.py) (revision 92e3c08040ba5a628b99f06675423b6011affed2) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision 0ccc06cd10ffae3cf76d1abebe0ab7f93d506337) @@ -202,7 +202,7 @@ 200 : "DG thermistors/sensors temperature out of range." 201 : "HD pre-treatment mode wet self-test lc vs FMD failure." 202 : "HD pre-treatment mode dry pressure normal self-test failure." - 203 : "DG fill conductivities acid/bicarb invalid calibration record." + 203 : "Available for use." 204 : "DG RO rejection ratio out of range." 205 : "DG conductivity sensor fault." 206 : "DG dialysate fill runs out of time." @@ -305,7 +305,7 @@ 303 : "DG drain pump direction FPGA fault." 304 : "HD invalid usage record." 305 : "HD processor RAM error." - 306 : "Alarm ID available." + 306 : "DG CD1 CD2 check fault." 307 : "HD AC power lost second alarm." 308 : "DG CPi conductivity sensor invalid character received." 309 : "DG CPo conductivity sensor invalid character received."