Index: build/AlarmMapping.py =================================================================== diff -u -rbb566bcebdefe7e22e429eb8bd0e36ae27c345ba -r0e748d1af8c6854f89b9183270a533f9398088fb --- build/AlarmMapping.py (.../AlarmMapping.py) (revision bb566bcebdefe7e22e429eb8bd0e36ae27c345ba) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision 0e748d1af8c6854f89b9183270a533f9398088fb) @@ -9,25 +9,25 @@ 7 : "HD RTC battery low." 8 : "HD accelerometer failed POST." 9 : "DG accelerometer failed POST." - 10 : "Alarm ID available for use." - 11 : "Alarm ID available for use." + 10 : "HD blood leak sensor setting embedded mode failure." + 11 : "HD dialysate temperature below target temperature." 12 : "HD firmware image integrity POST test failed." 13 : "DG firmware image integrity POST test failed." - 14 : "HD blood flow invalid calibration." - 15 : "HD dialysate flow invalid calibration." + 14 : "DG invalid usage record CRC." + 15 : "HD invalid usage record CRC." 16 : "HD alarm audio failed POST." 17 : "HD UI POST failed." 18 : "DG load cells invalid calibration record." - 19 : "DG flow sensors invalid calibration record." + 19 : "HD dialysate temperature above target temperature." 20 : "DG conductivity sensors invalid calibration record." 21 : "DG drain line volume invalid calibration record." 22 : "DG reservoirs invalid calibration record." 23 : "DG acid concentrate calibration record." 24 : "DG bicarb concentrate calibration record." 25 : "DG accelerometer invalid calibration record." 26 : "HD accelerometer invalid calibration record." - 27 : "HD blood flow invalid calibration record." - 28 : "HD dialyzer flow invalid calibration record." + 27 : "HD blood leak sensor zero and self test sequence failed." + 28 : "Not used." 29 : "HD heparin force sensor invalid calibration record." 30 : "HD Software fault.\nSoftware found itself in an unexpected state." 31 : "Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running." @@ -71,8 +71,8 @@ 69 : "HD blood leak sensor invalid calibration record." 70 : "HD arterial pressure self-test failure alarm." 71 : "HD venous pressure self-test failure alarm." - 72 : "HD blood flow meter status check self-test failure alarm." - 73 : "HD dialysate flow meter status check self-test failure alarm." + 72 : "HD No load cell data message receive at least once every 2 seconds." + 73 : "HD No dialysate temperature data message receive at least once every 2 seconds." 74 : "HD blood leak detector self-test transmit timeout." 75 : "HD syringe pump self-test failure alarm." 76 : "HD monitored voltage is out of range." @@ -102,9 +102,9 @@ 100 : "HD no cartridge self-test timeout." 101 : "HD dry self-test timeout." 102 : "DG heat disinfect temperature gradient out of range." - 103 : "HD non-volatile calibration group invalid record CRC." + 103 : "HD invalid calibration CRC." 104 : "Air trap level sensors reporting illegal combination of air/fluid." - 105 : "DG non-volatile system group invalid record CRC." + 105 : "DG invalid calibration CRC." 106 : "DG dialysate flow sensor invalid calibration record." 107 : "HD reports DG restarted fault." 108 : "HD syringe pump ADC error." @@ -154,8 +154,8 @@ 152 : "Inlet water pressure in low range." 153 : "HD prime completed high priority alarm." 154 : "EEPROM operations (read, write, erase) failure." - 155 : "Alarm ID available for use." - 156 : "Alarm ID available for use." + 155 : "HD No reservoirs data message receive at least once every 2 seconds." + 156 : "HD No DG operation mode message receive at least once every 2 seconds." 157 : "DG chemical disinfect prime acid line timeout." 158 : "Inlet water temperature is in the low range." 159 : "Dialysate inlet pump failed flow vs motor speed check." @@ -175,7 +175,7 @@ 173 : "HD RTC configuration error." 174 : "HD RTC or timer accuracy failure." 175 : "HD pump direction status error." - 176 : "HD active reservoir is full during treatment and DG not ready to switch reservoirs." + 176 : "HD dialysate temperature out of high safety range." 177 : "DG software fault.\nSoftware found itself in an unexpected state." 178 : "HD communication timeout." 179 : "DG FPGA communication down for too long." @@ -192,7 +192,7 @@ 190 : "DG drain pump off fault." 191 : "DG flow rate out of upper range." 192 : "DG flow rate out of lower range." - 193 : "Alarm available for use." + 193 : "HD dialysate temperature out of low safety range." 194 : "DG RO pump duty cycle out of range." 195 : "DG RO pump pressure out of range." 196 : "DG temperature sensor error bit fault." @@ -227,7 +227,7 @@ 225 : "HD end of treatment warning." 226 : "HD prime completed low priority alarm." 227 : "HD prime out of time alarm." - 228 : "HD time out on prime sub-mode purge air state." + 228 : "HD time out on prime saline purge air state." 229 : "HD prime dialysate dialyzer time out alarm." 230 : "HD prime dialysate bypass time out alarm." 231 : "HD pre-treatment mode dry pressure self-test failure." @@ -290,7 +290,12 @@ 288 : "HD load cells primary/back up drift out of range." 289 : "DG dialysate or concentrate caps not closed." 290 : "DG flow too low while heater is on." - 291 : "Total number of alarms." + 291 : "HD processor is in Disinfect Flush mode." + 292 : "HD processor is in Disinfect Heat mode." + 293 : "HD processor is in Disinfect Chemical mode." + 294 : "HD processor Disinfect Chemical mode instructions to user." + 295 : "UI POST CloudSync failure." + 296 : "Total number of alarms." } # dictionary of rejection reason from application source code.