Index: build/AlarmMapping.py =================================================================== diff -u -rd6b6841488e8bc33ba727ffcdba2c37bea8fb983 -rad9b54d94c0e1d0513e562e6d2ef4baa8a738e30 --- build/AlarmMapping.py (.../AlarmMapping.py) (revision d6b6841488e8bc33ba727ffcdba2c37bea8fb983) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision ad9b54d94c0e1d0513e562e6d2ef4baa8a738e30) @@ -167,7 +167,7 @@ 165 : "HD in treatment rinseback operation timeout." 166 : "Available for use." 167 : "HD cartridge door opened alarm." - 168 : "HD active reservoir depletion timeout." + 168 : "Available for use." 169 : "DG Dialysate flow rate out of maximum range." 170 : "HD syringe empty alarm." 171 : "HD syringe pump occlusion alarm." @@ -213,10 +213,10 @@ 211 : "Blood sitting too long warning (>4 min)." 212 : "HD end of treatment alarm (user not acting to end treatment)." 213 : "HD prime completed medium priority alarm." - 214 : "HD Saline bolus volume check failure." - 215 : "HD rinseback volume check failure." + 214 : "Available for use." + 215 : "Available for use." 216 : "HD end treatment sub-mode timeout alarm." - 217 : "HD blood prime volume check failure." + 217 : "Available for use." 218 : "HD syringe detected alarm." 219 : "HD syringe pump syringe removed alarm." 220 : "DG THd sensors FPGA fault." @@ -385,6 +385,7 @@ 54 : "REQUEST_REJECT_REASON_DG_RO_ONLY_MODE_DG_BUSY" # DG RO only mode DG is busy. 55 : "REQUEST_REJECT_REASON_DG_RO_ONLY_MODE_INVALID_PARAMETER" # DG RO only mode invalid parameter. 56 : "REQUEST_REJECT_REASON_DG_RO_ONLY_MODE_INVALID_PAYLOAD_LENGTH" # DG RO only mode invalid payload length. - 57 : "NUM_OF_REQUEST_REJECT_REASONS" # Number of settings change reject codes. + 57 : "REQUEST_REJECT_REASON_TREATMENT_CANNOT_BE_RESUMED" # Cannot resume treatment per alarm. + 58 : "NUM_OF_REQUEST_REJECT_REASONS" # Number of settings change reject codes. }