Index: build/AlarmMapping.py =================================================================== diff -u -rd957a403d7e7928de8701fd5bb5535e26539cabd -r4ad0d5250f340335a7e903cacc6e87f8fe23d934 --- build/AlarmMapping.py (.../AlarmMapping.py) (revision d957a403d7e7928de8701fd5bb5535e26539cabd) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision 4ad0d5250f340335a7e903cacc6e87f8fe23d934) @@ -14,10 +14,10 @@ 12 : "HD firmware image integrity POST test failed." 13 : "DG firmware image integrity POST test failed." 14 : "DG invalid usage record CRC." - 15 : "HD invalid usage record CRC." + 15 : "alarm ID available." 16 : "HD alarm audio failed POST." 17 : "HD UI POST failed." - 18 : "DG load cells invalid calibration record." + 18 : "HD didn't get ACK on message to DG that required acknowledgment." 19 : "HD dialysate temperature above target temperature." 20 : "DG conductivity sensors invalid calibration record." 21 : "DG drain line volume invalid calibration record." @@ -45,7 +45,7 @@ 43 : "DG main primary heater FPGA fault." 44 : "User interface communication timeout." 45 : "HD too many bad communications CRC." - 46 : "HD System message that required acknowledgment was not acknowledged." + 46 : "HD didn't get ACK on message to UI that required acknowledgment." 47 : "Ultrafiltration rate is too high error during treatment." 48 : "Ultrafiltration volume accuracy error during treatment." 49 : "HD FPGA communication down for too long." @@ -100,7 +100,7 @@ 98 : "HD battery communication fault." 99 : "HD syringe pump stall alarm." 100 : "HD no cartridge self-test timeout." - 101 : "HD dry self-test timeout." + 101 : "UI POST HD communication failure." 102 : "DG heat disinfect temperature gradient out of range." 103 : "HD invalid calibration CRC." 104 : "Air trap level sensors reporting illegal combination of air/fluid." @@ -115,7 +115,7 @@ 113 : "DG Drain pump current out of range." 114 : "HD venous air bubble detector self-test failure." 115 : "DG temperature sensor out of range." - 116 : "DG temperature sensor ADC fault." + 116 : "Available for use." 117 : "DG CAN message not acked." 118 : "DG RTC config error." 119 : "DG RTC battery low error." @@ -165,7 +165,7 @@ 163 : "HD sees primary load cell for reservoir 1 change too much too fast." 164 : "HD sees primary load cell for reservoir 2 change too much too fast." 165 : "HD in treatment rinseback operation timeout." - 166 : "HD in treatment re-circ sub-mode for too long." + 166 : "Available for use." 167 : "HD cartridge door opened alarm." 168 : "HD active reservoir depletion timeout." 169 : "DG Dialysate flow rate out of maximum range." @@ -202,7 +202,7 @@ 200 : "DG thermistors/sensors temperature out of range." 201 : "HD pre-treatment mode wet self-test lc vs FMD failure." 202 : "HD pre-treatment mode dry pressure normal self-test failure." - 203 : "Pressure sensor fault." + 203 : "Available for use." 204 : "RO rejection ratio out of range." 205 : "Conductivity sensor fault." 206 : "DG dialysate fill runs out of time." @@ -246,7 +246,7 @@ 244 : "DG cleaning mode temperature sensors difference out of range." 245 : "DG heat disinfect target temperature time out (could not reach to temperature)." 246 : "DG cleaning mode conductivity sensors out of range." - 247 : "DG heat disinfect active cool time tour." + 247 : "Alarm Id available." 248 : "DG chemical disinfect could not reach to target temperature out of range." 249 : "DG chemical disinfect target conductivity out of range." 250 : "DG chemical disinfect insert acid and remove the concentrate cap." @@ -257,7 +257,7 @@ 255 : "HD and UI software builds are not compatible." 256 : "HD processor is in Disinfect Chemical Flush mode." 257 : "Alarm ID available for use." - 258 : "HD FPGA power out timeout." + 258 : "***Available for use***." 259 : "HD temperatures out of range." 260 : "UI POST Application Integrity (Sha256Sum) failure." 261 : "UI POST CANBus failure." @@ -272,8 +272,8 @@ 270 : "HD POST Safety Shutdown failure." 271 : "DG POST Safety Shutdown failure." 272 : "HD Fan RPM out of range." - 273 : "HD measured blood pump flow rate is out of range." - 274 : "HD measured dialysate inlet pump flow rate is out of range." + 273 : "Available for use." + 274 : "Available for use." 275 : "HD arterial pressure sensor is reading out of range." 276 : "HD venous pressure sensor is reading out of range." 277 : "HD BP occlusion sensor is reading out of range." @@ -284,28 +284,28 @@ 282 : "Bicarbonate concentration bottle low volume alarm." 283 : "DG load cells weight out of range." 284 : "DG load cells primary/back up drift out of range." - 285 : "HD treatment recirculate timeout warning." - 286 : "HD treatment rinseback complete timeout warning." + 285 : "Available for use." + 286 : "DG concentrate cap not in proper position." 287 : "HD processor clock speed checks against FPGA clock failure." 288 : "HD load cells primary/back up drift out of range." - 289 : "DG dialysate or concentrate caps not closed." + 289 : "DG dialysate cap not closed." 290 : "HD venous pressure sensor FPGA fault." 291 : "HD processor is in Disinfect Flush mode." 292 : "HD processor is in Disinfect Heat mode." 293 : "HD processor is in Disinfect Chemical mode." - 294 : "HD processor Disinfect Chemical mode instructions to user." + 294 : "Available for use." 295 : "UI POST CloudSync failure." 296 : "DG TDi sensors FPGA fault." 297 : "DG TRo sensors FPGA fault." 298 : "DG baro sensor FPGA fault." 299 : "DG invalid serial number." 300 : "HD invalid serial number." - 301 : "DG primary heater power voltage out of range." + 301 : "Alarm ID available." 302 : "DG processor RAM error." 303 : "DG drain pump direction FPGA fault." 304 : "HD invalid usage record." 305 : "HD processor RAM error." - 306 : "HD air trap over-fill alarm." + 306 : "DG CD1 CD2 check fault." 307 : "HD AC power lost second alarm." 308 : "DG CPi conductivity sensor invalid character received." 309 : "DG CPo conductivity sensor invalid character received."