Index: build/AlarmMapping.py =================================================================== diff -u -r0ccc06cd10ffae3cf76d1abebe0ab7f93d506337 -r0ca14c2752248717d9befc48f72cd89f74ea5a24 --- build/AlarmMapping.py (.../AlarmMapping.py) (revision 0ccc06cd10ffae3cf76d1abebe0ab7f93d506337) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision 0ca14c2752248717d9befc48f72cd89f74ea5a24) @@ -202,7 +202,7 @@ 200 : "DG thermistors/sensors temperature out of range." 201 : "HD pre-treatment mode wet self-test lc vs FMD failure." 202 : "HD pre-treatment mode dry pressure normal self-test failure." - 203 : "Available for use." + 203 : "DG fill conductivities acid/bicarb invalid calibration record." 204 : "DG RO rejection ratio out of range." 205 : "DG conductivity sensor fault." 206 : "DG dialysate fill runs out of time." @@ -256,7 +256,7 @@ 254 : "DG invalid service record." 255 : "HD and UI software builds are not compatible." 256 : "HD processor is in Disinfect Chemical Flush mode." - 257 : "Alarm ID available for use." + 257 : "HD invalid institutional record." 258 : "HD UI POST OS version compatibility failure." 259 : "HD temperatures out of range." 260 : "HD UI POST Application Integrity (Sha256Sum) failure." @@ -305,7 +305,7 @@ 303 : "DG drain pump direction FPGA fault." 304 : "HD invalid usage record." 305 : "HD processor RAM error." - 306 : "DG CD1 CD2 check fault." + 306 : "Alarm ID available." 307 : "HD AC power lost second alarm." 308 : "DG CPi conductivity sensor invalid character received." 309 : "DG CPo conductivity sensor invalid character received." Index: build/Alarms.xlsx =================================================================== diff -u -rd2d5ca82148368f221c3beaf607b7234fc123673 -r0ca14c2752248717d9befc48f72cd89f74ea5a24 Binary files differ