Index: build/AlarmMapping.py =================================================================== diff -u -r524f60790b9910320bcaa8ba84ee60f5b8266b17 -r3a7cda682fba9d589d7b84635bb7af8d05036fa0 --- build/AlarmMapping.py (.../AlarmMapping.py) (revision 524f60790b9910320bcaa8ba84ee60f5b8266b17) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision 3a7cda682fba9d589d7b84635bb7af8d05036fa0) @@ -91,7 +91,7 @@ 89 : "DG outlet UV reactor on with no flow." 90 : "DG load cells A1/B1 FPGA fault." 91 : "HD No dialysate flow data receive in the last 3 seconds." - 92 : "DG temperature sensors invalid calibration record." + 92 : "Available for use." 93 : "DG outlet primary conductivity out of range." 94 : "DG pressure out of range." 95 : "DG watchdog expired." @@ -256,7 +256,7 @@ 254 : "DG invalid service record." 255 : "HD and UI software builds are not compatible." 256 : "HD processor is in Disinfect Chemical Flush mode." - 257 : "HD invalid institutional record." + 257 : "Alarm ID available for use." 258 : "HD UI POST OS version compatibility failure." 259 : "HD temperatures out of range." 260 : "HD UI POST Application Integrity (Sha256Sum) failure." Index: build/Alarms.xlsx =================================================================== diff -u -r524f60790b9910320bcaa8ba84ee60f5b8266b17 -r3a7cda682fba9d589d7b84635bb7af8d05036fa0 Binary files differ