Index: build/AlarmMapping.py =================================================================== diff -u -rac80d5567503d93c26c947f1e819a43906cd1bd5 -r7e326f8547a1c9f76e717f2a53e0c29861843553 --- build/AlarmMapping.py (.../AlarmMapping.py) (revision ac80d5567503d93c26c947f1e819a43906cd1bd5) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision 7e326f8547a1c9f76e717f2a53e0c29861843553) @@ -165,17 +165,17 @@ 163 : "HD sees primary load cell for reservoir 1 change too much too fast." 164 : "HD sees primary load cell for reservoir 2 change too much too fast." 165 : "HD in treatment rinseback operation timeout." - 166 : "Test alarm, clear top only." + 166 : "Available for use." 167 : "HD cartridge door opened alarm." - 168 : "Test alarm, clear top only." + 168 : "Available for use." 169 : "DG Dialysate flow rate out of maximum range." 170 : "HD syringe empty alarm." 171 : "HD syringe pump occlusion alarm." 172 : "HD syringe pump not enough Heparin alarm." 173 : "HD RTC configuration error." 174 : "HD RTC or timer accuracy failure." 175 : "HD pump direction status error." - 176 : "Test alarm, clear top only." + 176 : "HD dialysate temperature above high safety range." 177 : "DG software fault.\nSoftware found itself in an unexpected state." 178 : "HD communication timeout." 179 : "DG FPGA communication down for too long."