Index: build/AlarmMapping.py =================================================================== diff -u -r01d0bf9e30c83a568dd7d98fe66dad1d5ec09cc2 -rad26c85077eb2df24a6241a8027879029db5732c --- build/AlarmMapping.py (.../AlarmMapping.py) (revision 01d0bf9e30c83a568dd7d98fe66dad1d5ec09cc2) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision ad26c85077eb2df24a6241a8027879029db5732c) @@ -284,7 +284,7 @@ 282 : "Bicarbonate concentration bottle low volume alarm." 283 : "DG load cells weight out of range." 284 : "DG load cells primary/back up drift out of range." - 285 : "Available for use." + 285 : "HD processor is in RO permeate sample mode." 286 : "DG concentrate cap not in proper position." 287 : "HD processor clock speed checks against FPGA clock failure." 288 : "HD load cells primary/back up drift out of range." @@ -293,14 +293,14 @@ 291 : "HD processor is in Disinfect Flush mode." 292 : "HD processor is in Disinfect Heat mode." 293 : "HD processor is in Disinfect Chemical mode." - 294 : "Available for use." + 294 : "HD processor is in Disinfect active cool mode." 295 : "HD UI POST CloudSync failure." 296 : "DG TDi sensors FPGA fault." 297 : "DG TRo sensors FPGA fault." 298 : "DG baro sensor FPGA fault." 299 : "DG invalid serial number." 300 : "HD invalid serial number." - 301 : "Alarm ID available." + 301 : "DG RO permeate sample remove the dialysate cap." 302 : "DG processor RAM error." 303 : "DG drain pump direction FPGA fault." 304 : "HD invalid usage record." Index: build/Alarms.xlsx =================================================================== diff -u -r688c81190091be3b06d7c8659b91cd7aebdc3688 -rad26c85077eb2df24a6241a8027879029db5732c Binary files differ