Index: build/AlarmMapping.py =================================================================== diff -u -rb769fb894601322b4266f20544c08fe4e32e5687 -rc57dc49b40988daa35c7e823e42e83c21a7d6009 --- build/AlarmMapping.py (.../AlarmMapping.py) (revision b769fb894601322b4266f20544c08fe4e32e5687) +++ build/AlarmMapping.py (.../AlarmMapping.py) (revision c57dc49b40988daa35c7e823e42e83c21a7d6009) @@ -17,7 +17,7 @@ 15 : "HD invalid usage record CRC." 16 : "HD alarm audio failed POST." 17 : "HD UI POST failed." - 18 : "DG load cells invalid calibration record." + 18 : "HD didn't get ACK on message to DG that required acknowledgment." 19 : "HD dialysate temperature above target temperature." 20 : "DG conductivity sensors invalid calibration record." 21 : "DG drain line volume invalid calibration record." @@ -45,7 +45,7 @@ 43 : "DG main primary heater FPGA fault." 44 : "User interface communication timeout." 45 : "HD too many bad communications CRC." - 46 : "HD System message that required acknowledgment was not acknowledged." + 46 : "HD didn't get ACK on message to UI that required acknowledgment." 47 : "Ultrafiltration rate is too high error during treatment." 48 : "Ultrafiltration volume accuracy error during treatment." 49 : "HD FPGA communication down for too long." @@ -100,7 +100,7 @@ 98 : "HD battery communication fault." 99 : "HD syringe pump stall alarm." 100 : "HD no cartridge self-test timeout." - 101 : "HD dry self-test timeout." + 101 : "UI POST HD communication failure." 102 : "DG heat disinfect temperature gradient out of range." 103 : "HD invalid calibration CRC." 104 : "Air trap level sensors reporting illegal combination of air/fluid." @@ -175,7 +175,7 @@ 173 : "HD RTC configuration error." 174 : "HD RTC or timer accuracy failure." 175 : "HD pump direction status error." - 176 : "HD dialysate temperature out of high safety range." + 176 : "HD dialysate temperature above high safety range." 177 : "DG software fault.\nSoftware found itself in an unexpected state." 178 : "HD communication timeout." 179 : "DG FPGA communication down for too long." @@ -192,7 +192,7 @@ 190 : "DG drain pump off fault." 191 : "DG flow rate out of upper range." 192 : "DG flow rate out of lower range." - 193 : "HD dialysate temperature out of low safety range." + 193 : "HD dialysate temperature above low safety range." 194 : "DG RO pump duty cycle out of range." 195 : "DG RO pump pressure out of range." 196 : "DG CPi/CPo sensors FPGA fault." @@ -257,7 +257,7 @@ 255 : "HD and UI software builds are not compatible." 256 : "HD processor is in Disinfect Chemical Flush mode." 257 : "DG FPGA power out timeout." - 258 : "HD FPGA power out timeout." + 258 : "***Available for use***." 259 : "HD temperatures out of range." 260 : "UI POST Application Integrity (Sha256Sum) failure." 261 : "UI POST CANBus failure." @@ -305,7 +305,7 @@ 303 : "DG drain pump direction FPGA fault." 304 : "HD invalid usage record." 305 : "HD processor RAM error." - 306 : "HD air trap over-fill alarm." + 306 : "***Available for use***." 307 : "HD AC power lost second alarm." 308 : "DG CPi conductivity sensor invalid character received." 309 : "DG CPo conductivity sensor invalid character received." @@ -321,7 +321,8 @@ 319 : "DG chemical disinfect flush flush sample." 320 : "DG chemical disinfect flush sample timeout." 321 : "DG outlet UV reactor not healthy." - 322 : "Total number of alarms." + 322 : "RM_ID_HD_DIALYSATE_TEMP_BELOW_LOW_SAFETY_RANGE = 322,." + 323 : "Total number of alarms." } # dictionary of rejection reason from application source code.