Index: firmware/App/Controllers/AirTrap.c =================================================================== diff -u -re7525696d18a5a7c7481fc6a920e95d227a6e72f -r74fb4e00de266776cfc5d92163592ee895c23ed0 --- firmware/App/Controllers/AirTrap.c (.../AirTrap.c) (revision e7525696d18a5a7c7481fc6a920e95d227a6e72f) +++ firmware/App/Controllers/AirTrap.c (.../AirTrap.c) (revision 74fb4e00de266776cfc5d92163592ee895c23ed0) @@ -55,7 +55,7 @@ /// LDT-2384 test formula for air trap lower events for air pump duty cycle based on venous pressure /// air pump lowering duty cycle (% PWM) = 0.0005x^2 + 0.2313x + 63.667, where x is long filtered (60 sec) H14 venous pressure in mmHg before air pump lower event #define QUADRATIC_COEFFICIENT 0.0005F ///< X2 quadratic coefficient -#define LINEAR_COEFFICIENT 00.2313 ///< X linear coefficient +#define LINEAR_COEFFICIENT 0.2313F ///< X linear coefficient #define CONSTANT_TERM 63.667F ///< Constant term /// Defined states for the air trap controller state machine.