Index: firmware/App/Modes/ModeTreatment.h =================================================================== diff -u -r11a668a2f4a0401eed544be570d9102d143b6f00 -r316cb40192e5d6be69682d5c9df01d890e2c213f --- firmware/App/Modes/ModeTreatment.h (.../ModeTreatment.h) (revision 11a668a2f4a0401eed544be570d9102d143b6f00) +++ firmware/App/Modes/ModeTreatment.h (.../ModeTreatment.h) (revision 316cb40192e5d6be69682d5c9df01d890e2c213f) @@ -95,14 +95,13 @@ U32 venPresLimitAsymmetricmmHg; ///< Venous pressure limit asymmetric in mmHg. U32 tmpPresLimitWindowmmHg; ///< TMP pressure limit window in mmHg. } PRESSURE_LIMIT_CHANGE_RESPONSE_T; +#pragma pack(pop) /// Maximum time in this mode before blood sitting alarm given (in general task intervals). #define MAX_TIME_BLOOD_SITTING ( ( 5 * SEC_PER_MIN * MS_PER_SECOND ) / TASK_GENERAL_INTERVAL ) /// Maximum time in this mode before blood sitting warning given (in general task intervals). #define WARN_TIME_BLOOD_SITTING ( ( 4 * SEC_PER_MIN * MS_PER_SECOND ) / TASK_GENERAL_INTERVAL ) -#pragma pack(pop) - // ********** public function prototypes ********** void initTreatmentMode( void ); // Initialize this unit @@ -126,6 +125,9 @@ U32 getTreatmentStartTimeStamp( void ); // Get the treatment start time stamp U32 getTreatmentEndTimeStamp( void ); // Get the treatment end time stamp +// MSG_ID_TD_PRESSURE_LIMITS_CHANGE_RESPONSE +BOOL sendPressureLimitsChangeResponse( PRESSURE_LIMIT_CHANGE_RESPONSE_T *data ); + /**@}*/ #endif