Index: firmware/App/Monitors/Bubbles.c =================================================================== diff -u -r17d7a287b62351847759e24e5be390db19c4a531 -r539f3837b28f2c3998b559aae085eb8fd872e334 --- firmware/App/Monitors/Bubbles.c (.../Bubbles.c) (revision 17d7a287b62351847759e24e5be390db19c4a531) +++ firmware/App/Monitors/Bubbles.c (.../Bubbles.c) (revision 539f3837b28f2c3998b559aae085eb8fd872e334) @@ -225,7 +225,7 @@ if ( BUBBLE_DETECTED == getBubbleDetectedState( bubble ) ) { //Skip venous bubble alarm when no tubing set is installed - if ( FALSE == getTestConfigStatus(TEST_CONFIG_TESTING_WITHOUT_TUBING_SET) ) + if ( TRUE != getTestConfigStatus(TEST_CONFIG_TESTING_WITHOUT_TUBING_SET) ) { activateAlarmNoData( ALARM_ID_TD_VENOUS_BUBBLE_DETECTED ); }