Index: firmware/App/Monitors/Pressures.c =================================================================== diff -u -r036a75d76ab01912646a480b935d97187a231a19 -rb88a8fd2c5fca8f055d6df1437635692388fcb0d --- firmware/App/Monitors/Pressures.c (.../Pressures.c) (revision 036a75d76ab01912646a480b935d97187a231a19) +++ firmware/App/Monitors/Pressures.c (.../Pressures.c) (revision b88a8fd2c5fca8f055d6df1437635692388fcb0d) @@ -542,8 +542,10 @@ determineArtVenPressureLimits(); // Check in-line pressures are in range +#ifndef TEST_NO_PRESSURE_CHECKS checkArterialPressureInRange(); checkVenousPressureInRange(); +#endif return result; } @@ -806,7 +808,9 @@ { if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_TD_VENOUS_PRESSURE_LOW, venPresLow ) ) { +#ifndef TEST_UI_ONLY SET_ALARM_WITH_2_F32_DATA( ALARM_ID_TD_VENOUS_PRESSURE_LOW, venPres, (F32)currentVenousMinLimit ); +#endif } } // during reset period, if airTrapValve open, still alarm needs to be raised , hence persistence clearance should happen before or after reset period.