Index: firmware/App/Monitors/Pressures.c =================================================================== diff -u -r468cf0b9c944111231359948c833c20ad92219d0 -r694e3225e2e359fcb014a75bfaa8a93ef30f354c --- firmware/App/Monitors/Pressures.c (.../Pressures.c) (revision 468cf0b9c944111231359948c833c20ad92219d0) +++ firmware/App/Monitors/Pressures.c (.../Pressures.c) (revision 694e3225e2e359fcb014a75bfaa8a93ef30f354c) @@ -423,16 +423,15 @@ { filtArt = getFilteredArterialPressure(); filtVen = getFilteredVenousPressure(); + filtTmp = getTMPPressure(); } else { filtArt = getLongFilteredArterialPressure(); filtVen = getLongFilteredVenousPressure(); + filtTmp = getLongFilteredTMPPressure(); } - // Using final TMP value for override. - filtTmp = getLongFilteredTMPPressure(); - curArtPres = FLOAT_TO_INT_WITH_ROUND( filtArt ); curVenPres = FLOAT_TO_INT_WITH_ROUND( filtVen ); curTmpPres = FLOAT_TO_INT_WITH_ROUND( filtTmp ); @@ -497,8 +496,8 @@ { // apply pressure windows when stable S32 artOffset = (S32)getSysConfigTreatmentParameterU32( TREATMENT_PARAM_ART_PRES_LIMIT_WINDOW ); - S32 venMinOffset = (S32)getSysConfigTreatmentParameterU32( TREATMENT_PARAM_VEN_PRES_HIGH_LIMIT_WINDOW ); - S32 venMaxOffset = (S32)getSysConfigTreatmentParameterU32( TREATMENT_PARAM_VEN_PRES_LOW_LIMIT_WINDOW ); + S32 venMinOffset = (S32)getSysConfigTreatmentParameterU32( TREATMENT_PARAM_VEN_PRES_LOW_LIMIT_WINDOW ); + S32 venMaxOffset = (S32)getSysConfigTreatmentParameterU32( TREATMENT_PARAM_VEN_PRES_HIGH_LIMIT_WINDOW ); S32 tmpOffset = (S32)getSysConfigTreatmentParameterU32( TREATMENT_PARAM_TMP_PRES_LIMIT_WINDOW ); currentArterialMinLimit = stableArterialPressure - artOffset;