Index: firmware/App/Monitors/Pressures.c =================================================================== diff -u -r8740324a3f7730bdf614df73dcab916f35d03bce -r2a1c42890a75e4c4b8b7bec4c05efbfddc21247e --- firmware/App/Monitors/Pressures.c (.../Pressures.c) (revision 8740324a3f7730bdf614df73dcab916f35d03bce) +++ firmware/App/Monitors/Pressures.c (.../Pressures.c) (revision 2a1c42890a75e4c4b8b7bec4c05efbfddc21247e) @@ -64,21 +64,18 @@ #define ARTERIAL_PRESSURE_SELF_TEST_MAX ( 100.0F ) ///< Maximum self-test value for arterial pressure sensor reading (in mmHg). #define ARTERIAL_PRESSURE_MAX_MMHG ( 2000.0F ) ///< Maximum arterial pressure reading (in mmHg) for range check. #define ARTERIAL_PRESSURE_MIN_MMHG ( -1500.0F ) ///< Minimum arterial pressure reading (in mmHg) for range check. -#define ARTERIAL_PRESSURE_LIMIT_WINDOW_DEFAULT_MMHG ( 200.0F ) ///< Default Arterial pressure limit window (mmHg). Used as Window/2 for min/max limits #define ARTERIAL_PRESSURE_LIMIT_MAX_MMHG ( 100.0F ) ///< Maximum arterial pressure limit (in mmHg). #define ARTERIAL_PRESSURE_LIMIT_MIN_MMHG ( -300.0F ) ///< Minimum arterial pressure limit (in mmHg). #define VENOUS_PRESSURE_SELF_TEST_MIN ( -100.0F ) ///< Minimum self-test value for venous pressure sensor reading (in mmHg). #define VENOUS_PRESSURE_SELF_TEST_MAX ( 100.0F ) ///< Maximum self-test value for venous pressure sensor reading (in mmHg). #define VENOUS_PRESSURE_MAX_MMHG ( 2000.0F ) ///< Maximum venous pressure reading (in mmHg) for range check. #define VENOUS_PRESSURE_MIN_MMHG ( -1500.0F ) ///< Minimum venous pressure reading (in mmHg) for range check. -#define VENOUS_PRESSURE_LIMIT_WINDOW_DEFAULT_MMHG ( 200.0F ) ///< Default Venous pressure limit window (mmHg). Used as Window/2 for min/max limits. #define VENOUS_PRESSURE_LIMIT_MAX_MMHG ( 400.0F ) ///< Maximum venous pressure limit (in mmHg). #define VENOUS_PRESSURE_LIMIT_MIN_MMHG ( 20.0F ) ///< Minimum venous pressure limit (in mmHg). #define VENOUS_PRESSURE_EXEMPTION_PERIOD ( ( 2 * MS_PER_SECOND ) / \ TASK_GENERAL_INTERVAL ) ///< Venous pressure low exemption period (in task interval) after fill for all blood flow rate -#define TMP_PRESSURE_LIMIT_WINDOW_DEFAULT_MMHG ( 200.0F ) ///< Default TMP pressure limit window (mmHg). Used as Window/2 for min/max limits. #define TMP_PRESSURE_LIMIT_MAX_MMHG ( 500.0F ) ///< Maximum transmembrane pressure limit (in mmHg) #define TMP_PRESSURE_LIMIT_MIN_MMHG ( -500.0F ) ///< Minimum transmembrane pressure limit (in mmHg) #define TMP_PRESSURE_HIGH_ABSOLUTE_MMHG ( 500.0F ) ///< Transmembrane pressure high absolute (in mmHg)