Index: firmware/App/Monitors/Pressures.c =================================================================== diff -u -refea0f2103b95889d77f98daaa4175a6efb777c4 -rfe9b77e60850fe37da225d4348f0a6a8defc28db --- firmware/App/Monitors/Pressures.c (.../Pressures.c) (revision efea0f2103b95889d77f98daaa4175a6efb777c4) +++ firmware/App/Monitors/Pressures.c (.../Pressures.c) (revision fe9b77e60850fe37da225d4348f0a6a8defc28db) @@ -542,8 +542,10 @@ determineArtVenPressureLimits(); // Check in-line pressures are in range +#ifndef TEST_NO_PRESSURE_CHECKS checkArterialPressureInRange(); checkVenousPressureInRange(); +#endif return result; } @@ -806,7 +808,9 @@ { if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_TD_VENOUS_PRESSURE_LOW, venPresLow ) ) { +#ifndef TEST_UI_ONLY SET_ALARM_WITH_2_F32_DATA( ALARM_ID_TD_VENOUS_PRESSURE_LOW, venPres, (F32)currentVenousMinLimit ); +#endif } } // during reset period, if airTrapValve open, still alarm needs to be raised , hence persistence clearance should happen before or after reset period.