Index: firmware/App/Services/StateServices/DrySelfTests.c =================================================================== diff -u -redc719dbcce02d17d2693ab81712615c617527a0 -r3afb9f74631d80a7c8ef99e000d33c3f0552b7f8 --- firmware/App/Services/StateServices/DrySelfTests.c (.../DrySelfTests.c) (revision edc719dbcce02d17d2693ab81712615c617527a0) +++ firmware/App/Services/StateServices/DrySelfTests.c (.../DrySelfTests.c) (revision 3afb9f74631d80a7c8ef99e000d33c3f0552b7f8) @@ -395,7 +395,7 @@ if ( STATE_CLOSED == door ) { doorClosedRequired( TRUE ); - state = DRY_SELF_TESTS_USED_TUBING_SET_CHECK_STATE; + state = DRY_SELF_TESTS_PRESSURE_SENSOR_NORMAL_SETUP_STATE; } return state; @@ -506,7 +506,7 @@ { //tubingSetChecksCompleted = TRUE; pressureSelfTestPreNormalStartTime = getMSTimerCount(); - state = DRY_SELF_TESTS_PRESSURE_SENSOR_NORMAL_SETUP_STATE; +// state = DRY_SELF_TESTS_PRESSURE_SENSOR_NORMAL_SETUP_STATE; } } @@ -873,6 +873,7 @@ authResponseModalityAccepted = (BOOL)payload.modalityMatch; authResponseReceived = TRUE; result = TRUE; + barcodeScanResponseReceived = authResponseValidTubingSet && authResponseModalityAccepted; } return result;