Index: firmware/App/Services/StateServices/TubeSetAutoEject.c =================================================================== diff -u -r8dc6572eca96163f165be33a4c58d77bb4a53b44 -r05e43e76b6ea947bf51326c43aae77582ad70943 --- firmware/App/Services/StateServices/TubeSetAutoEject.c (.../TubeSetAutoEject.c) (revision 8dc6572eca96163f165be33a4c58d77bb4a53b44) +++ firmware/App/Services/StateServices/TubeSetAutoEject.c (.../TubeSetAutoEject.c) (revision 05e43e76b6ea947bf51326c43aae77582ad70943) @@ -239,7 +239,7 @@ else if ( bpTorque > BP_TORQUE_LIMIT_MNM ) { signalBloodPumpHardStop(); - SET_ALARM_WITH_1_F32_DATA( ALARM_ID_TD_AUTO_LOAD_EJECT_BLOOD_TUBE_SET_FAILURE, bpTorque ); + SET_ALARM_WITH_1_F32_DATA( ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE, bpTorque ); retractEjector(); autoEjectTimerCounter = 0; state = TUBE_SET_AUTO_EJECT_STATE_RETRACTING_EJECTOR; @@ -286,7 +286,7 @@ { autoEjectTimerCounter = 0; - if ( TRUE == isAlarmActive( ALARM_ID_TD_AUTO_LOAD_EJECT_BLOOD_TUBE_SET_FAILURE ) ) + if ( TRUE == isAlarmActive( ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE ) ) { setBloodPumpTargetFlowRate( AUTO_EJECT_BLOOD_FLOW_RATE_ML_MIN, MOTOR_DIR_REVERSE, PUMP_CONTROL_MODE_OPEN_LOOP ); state = TUBE_SET_AUTO_EJECT_STATE_BACK_OFF; @@ -327,9 +327,9 @@ else if ( TRUE == isPeristalticPumpHome() ) { signalBloodPumpHardStop(); - clearAlarmTD( ALARM_ID_TD_AUTO_LOAD_EJECT_BLOOD_TUBE_SET_FAILURE ); + clearAlarmTD( ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE ); - if ( FALSE == isAlarmActive( ALARM_ID_TD_AUTO_LOAD_EJECT_BLOOD_TUBE_SET_FAILURE ) ) + if ( FALSE == isAlarmActive( ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE ) ) { autoEjectTimerCounter = 0; state = TUBE_SET_AUTO_EJECT_STATE_AWAIT_CONFIRMATION;